GORE VIABAHN Endoprosthesis VBCR081501ARecalls & reported problems
W. L. Gore & Associates, Inc. · VIABAHN SX ENDO 35 RO 8mmX15cm 8Fr 75cmCath · Class 3 · Implant
All models of GORE VIABAHN Endoprosthesis (26)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00733132623549), received between Sep 28, 2023 and May 23, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Activation Failure” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure1
- Detachment of Device or Device Component1
- Device Damaged by Another Device1
- Patient-Device Incompatibility1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 23, 2025Detachment of Device or Device Component; Activation Failure; Detachment of Device or Device Component; Activation Failure
ON (B)(6) 2025, A PATIENT HAD A PROCEDURE TO TREAT VENOUS OUTFLOW WHERE AN 8 MM X 15 CM GORE® VIABAHN® ENDOPROSTHESIS WAS INTENDED TO BE USED IN AN UNKNOWN ANATOMICAL LOCATION. IT WAS REPORTED THE PHYSICIAN ACCESSED THE PATIENT USING AN 8F TERUMO INTRODUCER SHEATH OVER AN .035" GLIDEWIRE. IT IS UNKNOWN IF THE TARGET AREA WAS PRE-DILATED. REPORTEDLY, THE DELIVERY CATHETER ADVANCED TO THE TARGET TREATMENT ZONE WITHOUT INCIDENCE. DURING DEPLOYMENT, THE DEPLOYMENT LINE WAS PULLED AND BROKE. REPORTEDLY, 3-4 CENTIMETERS OF THE ENDOPROSTHESIS REMAINED CONSTRAINED AND ON THE CATHETER. THE DELIVERY CATHETER COULD NOT BE REMOVED WITH THE ENDOPROSTHESIS STILL ATTACHED. IT WAS REPORTED A BALLOON WAS ADVANCED AND INFLATED TO HIGH PRESSURE TO GET THE DEPLOYMENT LINE TO RUPTURE AND THE STENT TO EXPAND. ANOTHER ATTEMPT TO REMOVE THE DELIVERY CATHETER WAS MADE, AND IT APPEARED THE STENT STARTED TO BUCKLE. REPORTEDLY, MORE BALLOONING WAS DONE, AND THE AGGRESSIVE BALLOONING CAUSED THE STENT TO MOVE. A SECOND GORE® VIABAHN® DEVICE WAS USED TO FULLY TREAT THE TARGET AREA. IT WAS REPORTED THERE WAS NO IMPACT TO THE PATIENT OTHER THAN THE PROCEDURE TOOK LONGER THAN NORMAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22081810 ↗InjurySep 28, 2023Patient-Device Incompatibility; Device Damaged by Another Device; Patient-Device Incompatibility; Device Damaged by Another Device
THE FOLLOWING WAS REPORTED TO GORE: DURING A DECLOTTING PROCEDURE IN A PATIENT'S UPPER ARM, A SUBOPTIMAL VEIN WAS RUPTURED. THE TREATMENT AREA HAD BEEN PREDILATED. AS REPORTED, THE INTENTION WAS NOT TO COVER ANY PART OF THE DIALYSIS ACCESS GRAFT (UNKNOWN MANUFACTURER). IT WAS TO COVER THE DRAINING VEIN TO INCLUDE ONLY THE VEIN GRAFT ANASTOMOSIS. AN 8FR MERIT MEDICAL SHEATH WAS USED TO ADVANCE THE GORE® VIABAHN® ENDOPROSTHESIS (VSX DEVICE/STENT) TO BUT WAS UNABLE TO ADVANCE BEYOND THE LESION. DECISION WAS MADE TO WITHDRAW THE VSX DEVICE TO DILATE THE LESION AGAIN. WHILE TRYING TO REMOVE THE VSX DEVICE FROM THE PATIENT THROUGH THE 8FR SHEATH, THE PROXIMAL END OF THE VSX STENT WOULD NOT ENTER THE DISTAL END OF THE SHEATH. REPEATED ATTEMPTS WERE MADE TO MANIPULATE THE VSX STENT INTO THE SHEATH. AT THIS TIME, IT WAS VISUALIZED THE PROXIMAL PORTION OF THE STENT APPEARED FLARED AND THE DEPLOYMENT LINE WAS SEPARATED. OUTSIDE THE PATIENT, THE HUB ASSEMBLY WAS CUT TO ACCOMMODATE THE REMOVAL OF THE 8FR SHEATH. AFTER THE INITIAL SHEATH WAS REMOVED, A 9FR TERUMO SHEATH WAS ADVANCED INTO PLACE. ATTEMPTS TO REMOVE THE VSX STENT THROUGH THE 9FR SHEATH WERE UNSUCCESSFUL. AS REPORTED, THE PHYSICIAN
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 17836793 ↗
Other Stent, Superficial Femoral Artery devices
- EverFlex™ EVD35-06-150-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 death110 reports
- SUPERA S-60-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY108 reports
- SUPERA S-55-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY88 reports
- LifeStent® Solo Vascular Stent EX062003CLBard Peripheral Vascular, Inc. · STENT, SUPERFICIAL FEMORAL ARTERY80 reports
- BioMimics 3D Vascular Stent System 144700-12Veryan Medical Limited · Stent, Superficial Femoral Artery74 reports
- SUPERA S-50-120-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY55 reports
- EverFlex™ EVD35-07-060-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
- EverFlex™ EVD35-07-120-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page