Skip to content

GORE VIABAHN Endoprosthesis with Heparin VBJR070202ARecalls & reported problems

W. L. Gore & Associates, Inc. · VIABAHN SX ENDO HEPARIN 18 RO 7mmX2.5cm 7Fr 120cm Cath · Class 3 · Implant

All models of GORE VIABAHN Endoprosthesis with Heparin (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00733132623754), received between May 26, 2022 and Jul 22, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure2
  • Break1
  • Difficult to Open or Close1
  • Malposition of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 22, 2026Activation Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25906642 ↗
  • MalfunctionNov 18, 2025Activation Failure; Activation Failure

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2025, THE PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF AN UNKNOWN ETIOLOGY WITH A GORE® VIABAHN® ENDOPROSTHESIS (VSX DEVICE). REPORTEDLY, AFTER THE DEVICE WAS ADVANCED ACROSS THE LESION, THE PHYSICIAN ATTEMPTED TO DEPLOY THE DEVICE. HOWEVER, AFTER DEPLOYMENT WAS INITIATED, THE DEPLOYMENT LINE WAS STUCK AND THE VSX DEVICE WAS REMOVED. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23586953 ↗
  • MalfunctionJul 13, 2023Malposition of Device; Difficult to Open or Close; Malposition of Device; Difficult to Open or Close

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2023, A GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN DEVICE) WAS USED IN THE LEFT SPLENIC ARTERY TO TREAT A SPLENIC ARTERY ANEURYSM. ACCESS WAS GAINED THROUGH THE LEFT RADIAL ARTERY WITH A BOSTON SCIENTIFIC V-18 GUIDEWIRE THROUGH A COOK 8 FR INTRODUCER SHEATH. IT WAS REPORTED WHEN THE PHYSICIAN PULLED THE VIABAHN DEVICE DELIVERY SYSTEM DEPLOYMENT LINE THE VIABAHN DEVICE STARTED TO DEPLOY AND EXPAND, HOWEVER, DID NOT REACH FULL EXPANSION BEFORE ENTERING THE SPLENIC ARTERY ANEURYSMAL SAC. THE VIABAHN DEVICE WAS REMOVED FROM THE PATIENT WITHOUT DIFFICULTY. THE PROCEDURE WAS COMPLETED WITH COIL EMBOLIZATION OF THE LEFT SPLENIC ARTERY. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17322257 ↗
  • InjuryMay 26, 2022Break; Break

    IT WAS REPORTED TO GORE THAT DURING A PROCEDURE TO TREAT THE CELIAC ARTERY WHILE USING A 7 FR OSCAR STEERABLE SHEATH WITH A 0.18" V18 BOSTON SCIENTIFIC GUIDEWIRE THE GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE WAS UNABLE TO CROSS THE LESION. REPORTEDLY THE PHYSICIAN DID NOT PRE-DILATE THE LESION. IT WAS REPORTED THAT WHEN THE PHYSICIAN WAS ATTEMPTING TO PULL THE DEVICE BACK THROUGH THE SHEATH, THE DEVICE TIP GOT HUNG UP ON THE OSCAR SHEATH AND STARTED TO PARTIALLY DEPLOY. REPORTEDLY THE PHYSICIAN ATTEMPTED TO PULL THE SHEATH AND DEVICE DELIVERY SYSTEM OUT OF THE RIGHT COMMON FEMORAL ARTERY AND THE DEVICE DEPLOYED IN THE RIGHT EXTERNAL ILIAC ARTERY. IT WAS REPORTED THAT THE BLUE OLIVE TIP OF THE DELIVERY SYSTEM CAME OFF AND LODGED IN THE INTERNAL ILIAC ARTERY. REPORTEDLY THE PHYSICIAN CLOSED AND STATED THAT HE MAY ATTEMPT TO BRING HER BACK FOR ANOTHER PROCEDURE AT A LATER DATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14518520 ↗

Other System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices

All System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page