GORE VIABAHN Endoprosthesis with Heparin VBJR061002ARecalls & reported problems
W. L. Gore & Associates, Inc. · VIABAHN SX ENDO HEPARIN 18 RO 6mmX10cm 6Fr120cm Cath · Class 3 · Implant
All models of GORE VIABAHN Endoprosthesis with Heparin (62)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00733132623969), received between Oct 9, 2018 and Feb 26, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure4
- Activation Problem2
- Deformation Due to Compressive Stress1
- Detachment of Device or Device Component1
- Failure to Unfold or Unwrap1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Hemorrhage/Blood Loss/Bleeding1
- No Consequences Or Impact To Patient1
- Occlusion1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 26, 2026Activation Failure
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2026, A GORE® VIABAHN® ENDOPROSTHESIS (VSX DEVICE) WAS INTENDED FOR USE IN THE SUPERFICIAL FEMORAL ARTERY DURING TREATMENT OF A FISTULA. REPORTEDLY, AFTER BALLOONING, A QUESTIONABLE AV FISTULA FORMED THAT THE PHYSICIAN WANTED TO COVER AND SEAL. THE PHYSICIAN ADVANCED THE VSX DEVICE OVER A 0.018" GLIDE WIRE ADVANTAGE INTO POSITION AND ATTEMPTED TO DEPLOY THE VSX DEVICE. HOWEVER, AFTER DEPLOYMENT WAS INITIATED, THE DEPLOYMENT LINE HUNG UP AFTER A FEW INCHES. THE PHYSICIAN DID NOT WANT TO PULL HARDER AND BREAK THE DEPLOYMENT LINE AND REMOVED THE VSX DEVICE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NEW VSX DEVICE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24457244 ↗MalfunctionAug 12, 2025Activation Failure; Activation Failure
ON (B)(6) 2025, DURING AN ENDOVASCULAR TREATMENT, A GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN DEVICE) WAS INTENDED TO BE IMPLANTED IN THE RIGHT LOWER EXTREMITY, POPLITEAL ARTERY. THE TARGET TREATMENT AREA WAS REACHED WITH NO KNOWN RESISTANCE BY USING AN UNKNOWN SHEATH WITH A .018" GUIDEWIRE (MANUFACTURE UNKNOWN). THE PHYSICIAN INITIATED DEPLOYMENT BY PULLING ON THE DEPLOYMENT LINE. ITS BELIEVED THE DEPLOYMENT LINE WAS PULLED ALL THE WAY, NOTICING NOTHING OUT OF THE ORDINARY, WHEN THE VIABAHN DEVICE DID NOT DEPLOY. THERE WERE NO REPORTS OF DEVICE EXPANSION. THE VIABAHN DEVICE WAS WITHDRAWN THROUGH THE SHEATH (FOR RETURN) AND THE PROCEDURE WAS COMPLETED USING A NEW SHEATH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22787311 ↗MalfunctionMar 4, 2025Activation Problem; Activation Problem
ON (B)(6) 2025, A PATIENT RECEIVED TREATMENT FOR A PERFORATION IN THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA), UTILIZING A GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN® DEVICE). IT WAS REPORTED THAT DURING THE PROCEDURE, THE VIABAHN® DEVICE FAILED TO DEPLOY. A 6FR TERUMO DESTINATION BRAND SHEATH AND A TERUMO GLIDEWIRE ADVANTAGE BRAND GUIDEWIRE WERE UTILIZED. THE TARGET LESION HAD REPORTEDLY BEEN PREDILATED. IT WAS REPORTED THAT THERE WAS NO DISTAL EXPANSION OR OPENING OF THE STENT. ANOTHER VIABAHN® DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21516407 ↗MalfunctionSep 18, 2024
WE WERE ONLY ABLE TO DEPLOY THE DISTAL PORTION OF THE STENT AND THE STENT BECAME STUCK IN THIS LOCATION. MULTIPLE ATTEMPTS TO FURTHER DEPLOY THE VIABAHN STENT WITH THE STRING WERE MET WITH RESISTANCE AND BENDING OF THE CATHETER
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20245481 ↗MalfunctionAug 13, 2024Activation Failure; Activation Failure
ON (B)(6) 2024, A PATIENT UNDERWENT TREATMENT FOR STENOTIC OCCLUSION WHERE A 6 MM X 10 CM GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (GORE® VIABAHN® DEVICE) WAS INTENDED TO BE USED IN THE SUPERFICIAL FEMORAL ARTERY (SFA). IT WAS REPORTED THE PHYSICIAN USED ROTAREX¿ ROTATIONAL EXCISIONAL ATHERECTOMY SYSTEM AND THEN ACCESSED THE PATIENT USING A 6F MERIT MEDICAL ROADSTER GUIDE SHEATH OVER AN .018" ABBOTT COMMAND GUIDE WIRE TO THE TARGET TREATMENT ZONE. REPORTEDLY, THERE WAS NO RESISTANCE FELT DURING ADVANCEMENT OF THE DELIVERY CATHETER. DURING DEPLOYMENT, THE LINE WAS PULLED AND BECAME STUCK. THE ENDOPROSTHESIS STARTED TO BUCKLE ('COBRA HEAD') AND WOULD NOT RELEASE. REPORTEDLY, THERE WAS NO DISTAL EXPANSION INITIATED AND THE DEPLOYMENT LINE DID NOT BREAK. THE PHYSICIAN WITHDREW THE CATHETER, WITH THE ENDOPROSTHESIS FULLY CONSTRAINED. THE PROCEDURE WAS COMPLETED USING ANOTHER 6 MM X 10 CM GORE® VIABAHN® DEVICE. IT WAS REPORTED THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19975330 ↗MalfunctionMay 23, 2024Activation Failure; Activation Failure
ON (B)(6) 2024, A PATIENT UNDERWENT TREATMENT FOR STENOSIS IN THE LEFT DISTAL SUPERFICIAL FEMORAL ARTERY / PROXIMAL POPLITEAL ARTERY USING A 6 MM X 10 CM GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN® DEVICE). IT WAS REPORTED THE PHYSICIAN ACCESSED THE PATIENT VIA THE RIGHT GROIN UP-AND-OVER USING A 7F PINNACLE® DESTINATION® GUIDING SHEATH (90 CM) OVER A 0.018" COMMAND GUIDEWIRE TO THE TARGET TREATMENT ZONE. REPORTEDLY, THE LESION WAS NOT PREDILATED, AND IT WAS TIGHT ADVANCING THE CATHETER TO THE LANDING ZONE DUE TO THE STENOSIS. DURING THE ATTEMPTED DEPLOYMENT, THE DEPLOYMENT LINE WAS PULLED AND ENDOPROSTHESIS INITIATED DISTAL EXPANSION (ABOUT 1 CM) AND THEN BECAME STUCK. THE PHYSICIAN MANIPULATED (MOVED AROUND) THE CATHETER AND ATTEMPTED TO PULL AGAIN. THE DEPLOYMENT LINE WAS STILL STUCK, THE DEVICE STARTED TO BOWSTRING AND DEPLOYMENT WAS ABORTED. DUE TO CONCERN THE LINE MAY BREAK, THE DECISION WAS MADE TO WITHDRAW THE DELIVERY CATHETER. DURING WITHDRAWAL, THE DELIVERY CATHETER WAS PULLED CLOSE TO THE SHEATH, AND BOTH WERE PULLED BACK UP AND OVER THE AORTIC BIFURCATION. THE PHYSICIAN COULD NOT PULL THE DELIVERY CATHETER ALL THE WAY BACK INTO THE SHEATH DUE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19382833 ↗MalfunctionJun 2, 2022Activation Problem; Activation Problem
ON (B)(6) 2022, THE PATIENT WAS TREATED FOR A PSEUDOANEURYSM IN THE FEMORAL ARTERY USING A GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE. THE DEVICE WAS DELIVERED THROUGH A 7FR TERUMO DESTINATION SHEATH OVER A V-18 WIRE. THE DEVICE MET RESISTANCE ADVANCING THROUGH THE SHEATH. HOWEVER, THE DEVICE WAS ABLE TO BE ADVANCED TO THE INTENDED TARGET LESION. THE DEVICE WAS ATTEMPTED TO BE DEPLOYED. THE DEPLOYMENT LINE WAS PULLED AND THE DEVICE PARTIALLY EXPANDED. THE PHYSICIAN WITHDREW THE PARTIALLY EXPANDED DEVICE BACK THROUGH THE SHEATH. THE PHYSICIANS STATED THE SUSPECTED CAUSE OF THE PARTIAL DEPLOYMENT IS UNKNOWN. ANOTHER GORE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14580698 ↗MalfunctionJan 9, 2020Failure to Unfold or Unwrap; Failure to Unfold or Unwrap
THE FOLLOWING WAS REPORTED TO GORE: THE DOCTOR REPORTED THAT DURING AN UP AND OVER CASE, THE DEVICE WAS POSITIONED OVER A .014 WIRE AND DEPLOYMENT INITIATED. HOWEVER, THERE WAS NO TENSION ON THE DEPLOYMENT LINE AND THE DEPLOYMENT LINE WAS PULLED COMPLETELY OUT WITHOUT DEVICE EXPANSION. THE DEVICE WAS REMOVED FROM THE PATIENT.
Patient problems: Occlusion
MAUDE report 9570886 ↗MalfunctionDec 1, 2018Deformation Due to Compressive Stress; Material Split, Cut or Torn; Deformation Due to Compressive Stress; Material Split, Cut or Torn
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2018 A PATIENT WAS BEING TREATED EMERGENTLY FOR A HEMORRHAGE. A GORE® VIABAHN® ENDOPROSTHESIS WAS LOADED OVER A .018" WIRE AND ADVANCED, BUT DURING ADVANCEMENT, THE DELIVERY CATHETER BROKE. THE DEVICE WAS REMOVED ENTIRELY AND A SECOND DEVICE WAS ADVANCED AND DEPLOYED SUCCESSFULLY. THERE WAS NO ADVERSE OUTCOME FOR THE PATIENT. DURING THE COURSE OF THE INVESTIGATION IT WAS LEARNED THAT THE DELIVERY CATHETER DID NOT BREAK APART, BUT RATHER IT KINKED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 8123024 ↗MalfunctionOct 9, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2018 A PATIENT WAS UNDERGOING TREATMENT OF UNKNOWN TYPE AND ANATOMICAL LOCATION WITH A GORE® VIABAHN® ENDOPROSTHESIS. WHEN DEPLOYMENT WAS ATTEMPTED, THE PROXIMAL END OF THE STENT WOULD NOT EXPAND. THE DELIVERY CATHETER WAS THEN PULLED BACK AND THE STENT CAME BACK WITH IT. THE DELIVERY CATHETER THEN SEPARATED FROM THE STENT AND THEN THE STENT DEPLOYED. A SECOND STENT WAS NEEDED TO BRIDGE TO THE FIRST ONE, AND THE PROCEDURE WAS CONCLUDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7947558 ↗
Other Stent, Superficial Femoral Artery devices
- EverFlex™ EVD35-06-150-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 death110 reports
- SUPERA S-60-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY108 reports
- SUPERA S-55-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY88 reports
- LifeStent® Solo Vascular Stent EX062003CLBard Peripheral Vascular, Inc. · STENT, SUPERFICIAL FEMORAL ARTERY80 reports
- BioMimics 3D Vascular Stent System 144700-12Veryan Medical Limited · Stent, Superficial Femoral Artery74 reports
- SUPERA S-50-120-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY55 reports
- EverFlex™ EVD35-07-060-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
- EverFlex™ EVD35-07-120-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page