Skip to content

GORE VIABAHN Endoprosthesis with Heparin VBHR070702ARecalls & reported problems

W. L. Gore & Associates, Inc. · VIABAHN SX ENDO HEPARIN 35 RO 7mmX7.5cm 8Fr120cmCath · Class 3 · Implant

All models of GORE VIABAHN Endoprosthesis with Heparin (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00733132624058), received between Nov 15, 2024 and Jun 4, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2025Activation Failure; Activation Failure

    THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2025, THE PATIENT UNDERWENT A DIALYSIS GRAFT STENT PROCEDURE WHERE A GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN DEVICE) WAS INTENDED TO BE USED. THE PHYSICIAN WAS ABLE TO REACH THE TARGET TREATMENT AREA BY USINGS AN UNKNOWN INTRODUCER SHEATH AND AN UNKNOWN GUIDEWIRE. DEPLOYMENT WAS INITIATED. WHEN PULLING THE DEPLOYMENT LINE, THE PHYSICIAN EXPERIENCED RESISTANCE. THE DEPLOYMENT LINE WAS DIFFICULT TO PULL SO THEY WITHDREW THE VIABAHN DEVICE AND DISCARDED IT. THE PROCEDURE WAS COMPLETED USING A NEW VIABAHN DEVICE. IT WAS REPORTED THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22142810 ↗
  • MalfunctionNov 15, 2024

    PLACED INTO PATIENT UPON ATTEMPTED DEPLOYMENT OF STENT. PHYSICIAN HAD DIFFICULTLY AND ELECTED TO REMOVE. UPON INSPECTION, IT WAS NOTED THE STENT WAS NOT IN ITS PROPER POSITION BY THE SURGEON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20694041 ↗

Other System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices

All System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices →

Sources and updates