TIGRIS Vascular Stent PHA051002ARecalls & reported problems
W. L. Gore & Associates, Inc. · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00733132636198), received between Aug 10, 2018 and Sep 25, 2019 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Activation Failure1
- Activation, Positioning or Separation Problem1
- Patient-Device Incompatibility1
- Positioning Failure1
- Positioning Problem1
Patient problems
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2019Patient-Device Incompatibility; Patient-Device Incompatibility
DURING TREATMENT IN THE PATIENT'S SUPERFICIAL FEMORAL ARTERY (SFA) DUE TO OCCLUSIVE DISEASE, A GORE® TIGRIS® VASCULAR STENT (STENT) WAS ADVANCED AFTER THE ARTERY WAS PREDILATED BY A 3MM BALLOON. IT WAS REPORTED THAT DURING ADVANCEMENT, THERE WAS DIFFICULTY PUSHING THE DELIVERY CATHETER PAST THE PROXIMAL SFA. THE PHYSICIAN REALIZED THAT ABOUT 2CM OF THE STENT WAS UNSHEATHED AT THE DISTAL TIP AND THE CONSTRAINT SLEEVE APPEARED TO BE DISPLACED. THE DELIVERY CATHETER WAS PULLED BACK THROUGH THE INTRODUCER SHEATH AND REMOVED FROM THE PATIENT. AS REPORTED, THE CONSTRAINT SLEEVE LIKELY GOT STUCK ON VESSEL PLAQUE AS THE DEVICE WAS PUSHED. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9119530 ↗MalfunctionAug 21, 2018Positioning Failure; Activation, Positioning or Separation Problem; Positioning Problem; Positioning Failure; Activation, Positioning or Separation Problem; Positioning Problem
THE FOLLOWING WAS REPORTED TO GORE: DURING A SFA OCCLUSION CASE, THE FIRST GORE® TIGRIS® VASCULAR STENT WAS IMPLANTED WITHOUT INCIDENT. AS THE SECOND GORE® TIGRIS® VASCULAR STENT WAS BEING ADVANCED THROUGH THE PREVIOUSLY IMPLANTED GORE® TIGRIS® VASCULAR STENT, THE DOCTOR REPORTED THE DEVICE APPEARED TO PARTIALLY DEPLOY AND THEN BEGAN TO 'ACCORDION'. THE DEVICE WAS REMOVED WITHOUT INCIDENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING 2 ADDITIONAL GORE® TIGRIS® VASCULAR STENTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7803265 ↗MalfunctionAug 10, 2018Activation Failure; Activation Failure
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON B)(6) 2018 A PATIENT WAS UNDERGOING TREATMENT OF A HEAVILY CALCIFIED POPLITEAL ARTERY STENOSIS WITH A GORE® TIGRIS® VASCULAR STENT. AN ANTEGRADE APPROACH WAS UTILIZED. THE DEVICE WAS ADVANCED OVER A .035" WOOLY WIRE THROUGH A 6FR TERUMO SHEATH. WHEN THE TARGET LOCATION WAS REACHED DEPLOYMENT WAS INITIATED. APPROXIMATELY 3-4 CM OF THE DEVICE EXPANDED WHEN DEPLOYMENT STOPPED. THE DEPLOYMENT LINE SNAPPED, HOWEVER NO RESISTANCE WAS FELT PRIOR IN THE DEPLOYMENT WHEEL. THEN THE DEPLOYMENT WHEEL WAS TURNING WITHOUT ADDITIONAL DEVICE EXPANSION. THE DELIVERY CATHETER WAS THEN CUT OPEN IN ATTEMPTS TO RETRIEVE THE DEPLOYMENT LINE, BUT THIS WAS NOT SUCCESSFUL. THE SHEATH WAS ADVANCED FURTHER TO TRY AND CAPTURE THE DEVICE, WHICH WAS EVENTUALLY SUCCESSFUL. THE PROCEDURE WAS COMPLETED USING A BARE METAL BARD LIFESTENT. THERE WAS NO INJURY TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7774668 ↗
Other Stent, Superficial Femoral Artery devices
- EverFlex™ EVD35-06-150-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 death110 reports
- SUPERA S-60-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY108 reports
- SUPERA S-55-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY88 reports
- LifeStent® Solo Vascular Stent EX062003CLBard Peripheral Vascular, Inc. · STENT, SUPERFICIAL FEMORAL ARTERY80 reports
- BioMimics 3D Vascular Stent System 144700-12Veryan Medical Limited · Stent, Superficial Femoral Artery74 reports
- SUPERA S-50-120-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY55 reports
- EverFlex™ EVD35-07-060-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
- EverFlex™ EVD35-07-120-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page