Skip to content

TIGRIS Vascular Stent PHA070802ARecalls & reported problems

W. L. Gore & Associates, Inc. · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00733132636372), received between Dec 27, 2018 and Dec 27, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Mechanical Problem1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 27, 2018Mechanical Problem; Mechanical Problem

    THE FOLLOWING WAS REPORTED TO GORE: PATIENT PRESENTED FOR TREATMENT OF A STENOSIS IN THE POPLITEAL ARTERY. A GORE® TIGRIS® VASCULAR STENT WAS ADVANCED THROUGH A 6FR CORDIS SHEATH OVER A .014 SPIDER GUIDEWIRE. THE PHYSICIAN REPORTED THERE WAS TREMENDOUS RESISTANCE AS THE TIGRIS STENT WAS ADVANCED THROUGH THE SHEATH. THE TIGRIS STENT WAS ABLE TO REACH THE INTENDED TREATMENT SITE AND DEPLOYMENT WAS INITIATED. ABOUT THREE-FOURTH OF THE STENT EXPANDED WHEN DEPLOYMENT WHEEL COULD NO LONGER BE ROTATED. MOST OF THE STENT WAS PULLED BACK INTO THE 6FR CORDIS SHEATH AND REMOVED IN TANDEM. WIRE ACCESS WAS MAINTAINED, WHICH ALLOWED FOR EVERFLEX STENTS TO BE DEPLOYED. AS REPORTED, PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8201753 ↗

Other Stent, Superficial Femoral Artery devices

All Stent, Superficial Femoral Artery devices →

Sources and updates