GORE VIABAHN Endoprosthesis VBCR100502ARecalls & reported problems

W. L. Gore & Associates, Inc. · GORE VIABAHN Endoprosthesis 10mmx5cm 8Fr 120cmCath · Class 3 · Implant

All models of GORE VIABAHN Endoprosthesis (26)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00733132645800), received between Sep 30, 2024 and Jan 16, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure3
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 16, 2026Activation Failure

    THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2025, A PATIENT UNDERWENT AN ENDOVASCULAR TREATMENT FOR A LEFT UPPER ARM AV FISTULA ACCESS IN THE CEPHALIC ARCH. IT IS NOT SPECIFIED IF THERE WAS A PRE-EXISTING DEVICE. AS REPORTED, THE GORE® VIABAHN® ENDOPROSTHESIS (VIABAHN® DEVICE) REACHED THE INTENDED TREATMENT SITE AND DEPLOYMENT WAS STARTED. HOWEVER, THE VIABAHN® DEVICE DID NOT EXPAND COMPLETELY AT THE TREATMENT AREA (CEPHALIC ARCH). THE VIABAHN® DEVICE COULD NOT BE RELEASED FROM THE CATHETER UNTIL IT WAS PULLED BACK MORE PERIPHERALLY INTO THE CEPHALIC VEIN. IT WAS REPORTED THE CEPHALIC VEIN AREA WAS CONSIDERED TOO SMALL FOR THE VIABAHN® DEVICE. THE PHYSICIAN CHOSE A LARGER STENT (12X80 PROTEGE) TO TRAP THE VIABAHN® DEVICE TO PATIENT'S VESSEL.

    Patient problems: Foreign Body In Patient

    MAUDE report 24096741 ↗
  • MalfunctionJul 15, 2025Activation Failure; Physical Resistance/Sticking; Activation Failure; Physical Resistance/Sticking

    ON (B)(6) 2025, A PATIENT UNDERWENT A DIALYSIS ACCESS MAINTENANCE PROCEDURE TO ADDRESS A VENOUS OUTFLOW STENOSIS. THE INTERVENTION WAS PERFORMED IN AN UPPER EXTREMITY ARTERIOVENOUS (AV) ACCESS CIRCUIT; HOWEVER, IT IS UNKNOWN WHETHER THE ACCESS WAS VIA A GRAFT OR A NATIVE FISTULA. THE PROCEDURE INVOLVED THE INTENDED USE OF A GORE® VIABAHN® ENDOPROSTHESIS (GORE® VIABAHN® DEVICE). IT WAS REPORTED THE PHYSICIAN ACHIEVED VASCULAR ACCESS USING AN 8F INTRODUCER SHEATH (MANUFACTURER UNSPECIFIED), IN CONJUNCTION WITH A BOSTON SCIENTIFIC 0.035" ZIPWIRE¿ HYDROPHILIC GUIDEWIRE. IT IS NOT KNOWN WHETHER PRE-DILATION OF THE TARGET LESION WAS PERFORMED PRIOR TO ATTEMPTED DEVICE DEPLOYMENT. THE GORE® VIABAHN® DEVICE DELIVERY SYSTEM WAS REPORTEDLY ADVANCED TO THE INTENDED TREATMENT SITE. DURING THE DEPLOYMENT PHASE, THE OPERATOR ATTEMPTED TO ACTUATE THE DEPLOYMENT MECHANISM BY PULLING THE DEPLOYMENT LINE. HOWEVER, THE LINE BECAME IMMOBILE AND COULD NOT BE RETRACTED FURTHER. IT WAS REPORTED THAT THE DEPLOYMENT ATTEMPT OCCURRED AT AN ATYPICAL ANGLE, WHICH MAY HAVE CONTRIBUTED TO THE MALFUNCTION. THERE WAS NO REPORTED BREAKAGE OF THE DEPLOYMENT LINE, NOR WAS THERE ANY REPORT OF INITIATION OF DISTAL END

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22510211 ↗
  • MalfunctionSep 30, 2024Activation Failure; Activation Failure

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2024, A GORE® VIABAHN® ENDOPROSTHESIS (VSX DEVICE) WAS INTENDED FOR USE IN AN UNKNOWN ANATOMICAL LOCATION DURING TREATMENT OF AN UNKNOWN ETIOLOGY. DURING THE DEPLOYMENT OF THE VSX DEVICE, THE DEPLOYMENT LINE STUCK AND STARTED TO BOW STRING. THE VSX DEVICE WAS REMOVED THROUGH THE INTRODUCER SHEATH (SIZE AND MANUFACTURER UNKNOWN) AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN ADDITIONAL VSX DEVICE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20334092 ↗