Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00733132645923), received between Mar 11, 2024 and Jul 24, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure4
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 24, 2025Activation Failure; Physical Resistance/Sticking; Activation Failure; Physical Resistance/Sticking
ON (B)(6) 2025, THE PATIENT UNDERWENT TREATMENT FOR STENOSIS IN THE CEPHALIC ARCH/SUBCLAVIAN VEIN WHERE A GORE® VIABAHN® ENDOPROSTHESIS (VIABAHN DEVICE) WAS INTENDED TO BE USED. ACCESS TO THE TARGET SITE WAS SUCCESSFULLY ACHIEVED VIA A 10FR BOSTON SCIENTIFIC SUPER SHEATH AND A 180 CM BOSTON SCIENTIFIC BENTSON GUIDEWIRE WITHOUT ANY RESISTANCE. UPON DELIVERY OF THE VIABAHN DEVICE, DEPLOYMENT WAS INITIATED BY PULLING ON THE DEPLOYMENT LINE. DURING THIS PROCESS, THE PHYSICIAN ENCOUNTERED RESISTANCE. THE ENDOPROSTHESIS PARTIALLY EXPANDED AT THE DISTAL END; HOWEVER, EXPANSION CEASED MIDWAY AND COULD NOT BE COMPLETED. MULTIPLE ATTEMPTS WERE MADE TO FULLY DEPLOY THE DEVICE, BUT RESISTANCE ALONG THE DEPLOYMENT LINE PERSISTED. ULTIMATELY, THE DEVICE WAS RETRIEVED IN TANDEM WITH THE SHEATH AND WAS SUCCESSFULLY REMOVED FROM THE PATIENT. THE PROCEDURE WAS THEN COMPLETED USING BALLOON ANGIOPLASTY. IT WAS REPORTED THERE WAS NO IMPACT TO THE PATIENT. NO TORTUOUS ANATOMY NOR CALCIFICATION WAS PRESENT IN THE TARGET TREATMENT AREA. HOWEVER, A PREVIOUSLY PLACED STENT (MANUFACTURER UNKNOWN) WAS PRESENT AT THE TREATMENT SITE. THE PHYSICIAN IS UNSURE WHAT CAUSED THE DEPLOYMENT LINE TO BECOME STUCK SINCE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22592942 ↗MalfunctionJun 10, 2024Activation Failure; Activation Failure
THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2024, THE PATIENT UNDERWENT A REVISION FOR AN ARTERIOVENOUS (AV) FISTULA IN THE UPPER ARM USING A GORE® VIABAHN® ENDOPROSTHESIS (VIABAHN DEVICE). IT WAS REPORTED THAT THE PHYSICIAN WAS ABLE TO REACH THE TARGET TREATMENT AREA WITHOUT ANY RESISTANCE. THE PHYSICIAN WAS READY TO INITIATED DEPLOYMENT WHEN THEY NOTICE SOME RESISTANCE WHILE PULLING THE DEPLOYMENT LINE. THE VIABAHN DEVICE STARTED TO BOWSTRING. THE PHYSICIAN DID NOT TRY TO FORCE IT OPEN SO THEY TOOK OUT THE VIABAHN DEVICE OUT THROUGH THE SHEATH. THERE WAS NO DEVICE EXPANSION. THE PROCEDURE WAS COMPLETED USING A NEW VIABAHN DEVICE. IT WAS REPORTED THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19506044 ↗MalfunctionMar 11, 2024Activation Failure; Activation Failure
ON (B)(6) 2024, A PATIENT UNDERWENT TREATMENT FOR CENTRAL VEIN STENOSIS USING A 13 MM X 10 CM GORE® VIABAHN® ENDOPROSTHESIS DEVICES (VIABAHN® DEVICES) IN THE INNOMINATE VEIN / VENA CAVA. IT WAS REPORTED THE PHYSICIAN ACCESSED THE PATIENT THROUGH A DIALYSIS ACCESS GRAFT IN THE UPPER ARM USING A 10F TERUMO INTRODUCER SHEATH OVER AN 0.035" GUIDEWIRE (UNKNOWN BRAND) TO THE TARGET TREATMENT SITE. REPORTEDLY, THE PHYSICIAN PULLED THE DEPLOYMENT LINE AND IT UNZIPPED HUB TO TIP, BUT GOT STUCK AT THE TURN AROUND AND COULD NOT BE PULLED. THE ENDOPROSTHESIS STARTED TO BOW WHEN FORCE WAS APPLIED TO PULL THE LINE. THE DECISION WAS MADE TO WITHDRAW THE DELIVERY CATHETER FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING A BD DEVICE. IT WAS REPORTED THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18882202 ↗MalfunctionMar 11, 2024Activation Failure; Activation Failure
ON (B)(6) 2024, A PATIENT UNDERWENT TREATMENT FOR CENTRAL VEIN STENOSIS USING A 13 MM X 10 CM GORE® VIABAHN® ENDOPROSTHESIS DEVICES (VIABAHN® DEVICES) IN THE INNOMINATE VEIN / VENA CAVA. IT WAS REPORTED THE PHYSICIAN ACCESSED THE PATIENT THROUGH A DIALYSIS ACCESS GRAFT IN THE UPPER ARM USING A 10F TERUMO INTRODUCER SHEATH OVER AN 0.035" GUIDEWIRE (UNKNOWN BRAND) TO THE TARGET TREATMENT SITE. REPORTEDLY, DURING DEPLOYMENT, THE PHYSICIAN PULLED THE DEPLOYMENT LINE AND THE ENDOPROSTHESIS INITIATED EXPANSION. IT EXPANDED APPROXIMATELY 1 CENTIMETER, BEFORE THE LINE COULD NOT BE PULLED ANY MORE. THE DELIVERY CATHETER AND THE SHEATH WERE WITHDRAWN FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING A BD DEVICE. IT WAS REPORTED THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18882130 ↗
Other System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices
- COVERA AVSM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall33 reports
- COVERA AVSM08060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall21 reports
- COVERA AVSM09060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall18 reports
- COVERA AVSM10080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall16 reports
- COVERA AVSM10040Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall15 reports
- COVERA AVFM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall14 reports
- COVERA AVSM09080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall1 death14 reports
- COVERA AVSM10100Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page