W. L. Gore & Associates, Inc. · GORE VIABAHN Endoprosthesis with Heparin 10mmx15cm 8Fr 120cmCath · Class 3 · Implant

All models of GORE VIABAHN Endoprosthesis with Heparin (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00733132646067), received between Jul 17, 2023 and Jan 15, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Activation Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure1
  • Fracture1
  • Malposition of Device1
  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Obstruction/Occlusion1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 15, 2025Activation Failure; Activation Failure

    THE FOLLOWING WAS REPORTED TO GORE: DURING TREATMENT OF A PATIENT'S LEFT POPLITEAL ARTERY ANEURYSM, ACCESS WAS MADE FROM THE RIGHT LEG, UP AND OVER THE BIFURCATION. AN 8FR TERUMO SHEATH WAS USED TO ADVANCE THE GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN DEVICE) OVER A 0.035 BENTSON WIRE. THE VIABAHN DEVICE WAS DELIVERED TO THE TREATMENT AREA. DEPLOYMENT WAS STARTED AND THE DEPLOYMENT LINE UNRAVELED TO THE DISTAL TIP, THEN THE DEVICE STARTED BOWSTRINGING DUE TO LINE TENSION. IT WAS REPORTED THE PHYSICIAN PULLED HARD ON THE DEPLOYMENT LINE, BUT THERE WAS NO FURTHER LINE DEPLOYMENT. THE GUIDEWIRE WAS EXCHANGED FOR AN AMPLATZ WIRE AND DEPLOYMENT LINE WAS PULLED ONCE MORE. THE VIABAHN DEVICE STILL CONTINUED TO BOWSTRING WITH A STUCK DEPLOYMENT LINE. THE VIABAHN DEVICE WAS WITHDRAWN FROM THE PATIENT WITH NO ISSUES. USING THE SAME AMPLATZ WIRE, A NEW VIABAHN DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES. AS REPORTED, THE DEPLOYMENT LINE MAY HAVE BEEN WRAPPED AROUND THE DISTAL TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21160332 ↗
  • InjuryJul 3, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2024, A GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN DEVICE) WAS IMPLANTED IN AN OFF-LABEL PROCEDURE TO REPAIR A POPLITEAL ANEURYSM. INTRAVASCULAR ULTRASOUND (IVUS) WAS USED FOR THE SIZING OF THE VIABAHN DEVICE. ON (B)(6) 2024, A REINTERVENTION OCCURRED IN WHICH AN ADDITIONAL VIABAHN DEVICE WAS ADDED TO EXTEND THE PROXIMAL PORTION OF THE VIABAHN DEVICE STENT GRAFT THAT WAS IMPLANTED ON (B)(6) 2024 DUE TO THROMBOSIS. THE PHYSICIAN REPORTEDLY PUT THE PATIENT ON ASPIRIN AND PLAVIX. THE PHYSICIAN DID NOT SUSPECT HEPARIN-INDUCED THROMBOCYTOPENIA (HIT). THE PHYSICIAN SUSPECTED THE EVENT WAS RELATED TO THE PATIENT IMMEDIATELY DOING PHYSICAL THERAPY AFTER THE (B)(6) 2024 PROCEDURE. THE PATIENT TOLERATED BOTH PROCEDURES AND IS DOING WELL.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 19671035 ↗
  • InjuryJul 17, 2023Fracture; Malposition of Device; Fracture; Malposition of Device

    ON (B)(6) 2023, A PATIENT UNDERWENT TREATMENT FOR AN ANEURYSM IN THE POPLITEAL ARTERY IN THE LEFT LEG UTILIZING A GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN DEVICE). IT WAS REPORTED THE PHYSICIAN PERFORMED USUAL RUNOFFS DETERMINING SUFFICIENT DISTAL RUNOFF AND TO DETERMINE THE LOCATION OF THE ANEURYSM. INTRAVASCULAR ULTRASOUND (IVUS) IMAGING WAS THEN PERFORMED TO SELECT THE VIABAHN DEVICE SIZE. REPORTEDLY, THE PROXIMAL VESSEL DIAMETER WAS 9 MM, THEREFORE A 10 MM X 15 CM VIABAHN DEVICE WAS SELECTED. REPORTEDLY, THE PHYSICIAN ACCESSED THE PATIENT CONTRALATERALLY VIA THE FEMORAL ARTERY. SINCE THE DEVICE REQUIRED AN 8F SHEATH, THE DECISION WAS MADE TO DEPLOY THE STENT WITHOUT A SHEATH OVER A .035" AMPLATZ GUIDE WIRE. THE STENT PASSED COMPLETELY INTO THE VESSEL, HOWEVER RESISTANCE WAS MET JUST PROXIMAL TO THE AORTIC BIFURCATION AND IT WOULD NOT ADVANCE ANY FURTHER. AN ATTEMPT TO WITHDRAW THE DEVICE WAS MADE. THE DELIVERY CATHETER WAS WITHDRAWN, BUT THE STENT-GRAFT WITH THE DEPLOYMENT LINE STILL ATTACHED REMAINED IN THE PATIENT'S COMMON ILIAC ARTERY. THE DECISION WAS MADE TO DEPLOY THE DEVICE. UNDER FLUOROSCOPY, IT WAS CONFIRMED THE STENT-GRAFT DEPLOYED SUCCESS

    Patient problems: Obstruction/Occlusion

    MAUDE report 17333574 ↗