GORE VIABAHN Endoprosthesis with Heparin VBHR111002ARecalls & reported problems

W. L. Gore & Associates, Inc. · GORE VIABAHN Endoprosthesis with Heparin 11mmx10cm 10Fr 120cmCath · Class 3 · Implant

All models of GORE VIABAHN Endoprosthesis with Heparin (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00733132646104), received between Apr 15, 2024 and Jun 8, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Activation Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure1
  • Patient-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Thrombocytopenia1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 8, 2024Activation Failure; Activation Failure

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2024, A GORE® VIABAHN® ENDOPROSTHESIS (VSX DEVICE) WAS INTENDED FOR USE IN THE ARM DURING TREATMENT OF AN ARTERIOVENOUS ACCESS REVISION. AFTER THE DEVICE WAS ADVANCED THROUGH THE TERUMO INTRODUCER SHEATH (SIZE UNKNOWN) OVER AN 0.035" BENTSON GUIDE WIRE, THE PHYSICIAN ATTEMPTED TO DEPLOY THE VSX DEVICE IN THE PREVIOUSLY IMPLANTED ARTERIOVENOUS ACCESS GRAFT (MANUFACTURER UNKNOWN). HOWEVER, AFTER DEPLOYMENT WAS INITIATED THE DEPLOYMENT LINE GOT HUNG UP AT THE TURN AROUND AND WOULD NOT DEPLOY. WITH NO DEVICE EXPANSION THE VSX DEVICE WAS REMOVED THOUGH THE INTRODUCER SHEATH. A NEW VSX DEVICE WAS SUCCESSFULLY DELIVERED TO COMPLETE THE CASE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19493564 ↗
  • InjuryApr 15, 2024Patient-Device Incompatibility; Patient-Device Incompatibility

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2024, TWO GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN DEVICES) WERE IMPLANTED TO REPAIR A POPLITEAL ANEURYSMS. IT WAS REPORTED ON AN UNKNOWN DATE AFTER THE IMPLANT PROCEDURE THROMBOSIS OF THE REPAIRED SITE WAS NOTED. REPORTEDLY, THE PHYSICIAN BELIEVES, DUE TO TIME OF ONSET AND AGGRESSIVENESS OF THROMBOSIS THAT THE PATIENT HAD HEPARIN INDUCED THROMBOCYTOPENIA (HIT). ADDITIONALLY, ON AN UNKNOWN DATE A REINTERVENTION OCCURRED (BYPASS REPAIR) IN WHICH BOTH VIABAHN DEVICES WERE EXPLANTED. AT AN UNKNOWN TIME SHORTLY AFTER THE BYPASS REPAIR THE SITE WAS THROMBOSED AS WELL. REPORTEDLY, ON AN UNKNOWN DATE AFTER PROCEDURE A PLATELET FACTOR 4 (PF4) TEST WAS COMPLETED WITH THE RESULT OF POSITIVE ANTIBODIES IN THE PATIENT. IT IS UNKNOWN WHAT TREATMENT THE PATIENT RECEIVED AFTER THE POSITIVE PF4, HOWEVER, THE PATIENT IS DOING WELL AND THE SYMPTOMS OF HIT HAVE RESOLVED.

    Patient problems: Thrombocytopenia; Thrombosis/Thrombus

    MAUDE report 19110963 ↗

Other Stent, Superficial Femoral Artery devices

All Stent, Superficial Femoral Artery devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page