GORE EXCLUDER Conf AAA Endo - Trunk-Ipsilateral CXT201212Recalls & reported problems

W. L. Gore & Associates, Inc. · EXCLUDER CONFORM AAA TRNK ENDO 20mmX12mmX12cm 15Fr · Class 3 · Implant

All models of GORE EXCLUDER Conf AAA Endo - Trunk-Ipsilateral (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
1
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00733132650989), received between Jul 6, 2021 and Apr 2, 2026 — classified as 1 death, 5 injuries.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deformation Due to Compressive Stress2
  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1
  • Obstruction of Flow1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Aneurysm expansion1
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Insufficient Information1
  • Obstruction/Occlusion1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 2, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT THE DEVICES WERE EXPLANTED DUE TO LIMB OCCLUSION, THE PATIENT PRESENTED WITH ISCHEMIA OF LOWER LIMB.

    Patient problems: Insufficient Information

    MAUDE report 24771888 ↗
  • InjurySep 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2022, THE PATIENT UNDERWENT AN ENDOVASCULAR PROCEDURE TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA) UTILIZING GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS (TRUNK IPSILATERAL LEG CONFORMABLE) AND GORE® EXCLUDER® AAA ENDOPROSTHESIS (CONTRALATERAL LEG AND ILIAC EXTENDER). ON (B)(6) 2024, PATIENT PRESENTED FOR A YEARLY FOLLOW UP AFTER THE INITIAL PROCEDURE WHERE PATIENT UNDERWENT AN ABDOMINAL ULTRASOUND THAT SHOWED THE AAA HAD INCREASED IN SIZE TO 4.6 X 4.1CM FROM 4.3 X 3.8CM IN 2023 (DATE UNKNOWN). THE PATIENT IS REPORTEDLY DOING WELL WITH NO REPORTED COMPLAINTS OF ABDOMINAL OR BACK PAIN AND NO CLAUDICATION. ON (B)(6) 2025, THE PATIENT UNDERWENT AN ENDOVASCULAR REINTERVENTION PROCEDURE TO TREAT AN EXPANDING ANEURYSM (55MM) UTILIZING A GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS (AORTIC CONFORMABLE EXTENDER). REPORTEDLY, DUE TO THE PATIENT'S EXPANDING ANEURYSM (CAUSE IS UNKNOWN) WITH NO CLEAR EVIDENCE OF ENDOLEAK, PHYSICIAN DECIDED TO DO AN ELECTIVE PROCEDURE AS A PRECAUTION BY EXTENDING PROXIMALLY WITH AN AORTIC CONFORMABLE EXTENDER FROM THE TRUNK IPSILATERAL LEG CONFORMABLE. PATIENT TOLERATED THE PROCEDURE. 1001-E: -UNKNOWN IS USED TO

    Patient problems: Aneurysm expansion

    MAUDE report 22992644 ↗
  • InjuryOct 9, 2024Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    ON (B)(6) 2024, PATIENT UNDERWENT AN ENDOVASCULAR PROCEDURE TO TREAT A ABDOMINAL AORTIC ANEURYSM, UTILIZING GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS. ON (B)(6) 2024, DURING A FOLLOW-UP VISIT, AN ULTRASOUND WAS PERFORMED WHICH SHOWED INCREASED VELOCITIES IN LEFT LEG. A CT WAS ALSO PERFORMED, AND IT SHOWED PATIENT'S LEFT ILIAC LIMB WAS COMPRESSED, NEAR THE AORTIC BIFURCATION. PHYSICIAN RECOMMENDED A REINTERVENTION SURGERY TO TREAT THIS COMPRESSION. ON (B)(6) 2024, PATIENT UNDERWENT A REINTERVENTION SURGERY TO TREAT THIS COMPRESSION. PHYSICIAN BALLOONED THE AREA WITH 8MMX40MM BALLOON AND THEN A 10MMX40MM BALLOON. PHYSICIAN WAS SATISFIED WITH RESULTS, AND STATED HE MUST'VE NOT BALLOONED THIS AREA ENOUGH IN THE INITIAL CASE. PATIENT TOLERATED THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20408293 ↗
  • InjuryMay 24, 2022Obstruction of Flow; Obstruction of Flow

    ON (B)(6) 2022, THE PATIENT WAS ADMITTED FOR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM AND WAS IMPLANTED WITH A GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS TRUNK IPSILATERAL COMPONENT AND A GORE® EXCLUDER® AAA ENDOPROSTHESIS CONTRALATERAL LEG COMPONENT. OVER THE FOLLOWING WEEKEND THE PATIENT WENT TO THE EMERGENCY ROOM, AND IT WAS IDENTIFIED THAT THERE WAS AN OCCLUDED LIMB. A REINTERVENTION WAS PERFORMED ON (B)(6) 2022. THE REINTERVENTION INVOLVED A THROMBECTOMY FOLLOWED BY THE IMPLANTATION OF TWO GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESES AT THE AORTIC BIFURCATION VIA KISSING BALLOON TECHNIQUE IN ORDER TO PRESERVE BLOOD FLOW. AS REPORTED, THE OCCLUSION OCCURRED ON THE IPSILATERAL SIDE. THE CAUSE FOR THE NEEDED REINTERVENTION WAS DIRECTLY IDENTIFIED AS THE PATIENT ANATOMY THAT NARROWED GREATLY AROUND THE BIFURCATION. AS REPORTED THE FLOW DIVIDER WAS LATER OBSERVED TO HAVE LANDED DIRECTLY IN THE RESTRICTED AREA OF THE PATIENT¿S ANATOMY. AS REPORTED, THERE WAS NO MORE THAN APPROXIMATELY 1CM TO A 1/2CM OF SPACE IN ONE SPOT. IT WAS INDICATED THAT FOLLOWING THE INITIAL TREATMENT PROCEDURE ON (B)(6) 2022, THAT THE POST-IMPLANTATION IMAGING CONDUCTED SHOWED THAT ALL VESSELS WE

    Patient problems: Obstruction/Occlusion

    MAUDE report 14488357 ↗
  • DeathAug 17, 2021Use of Device Problem; Use of Device Problem

    ON (B)(6) 2021, THIS PATIENT UNDERWENT ENDOVASCULAR TREATMENT FOR AORTIC OCCLUSIVE DISEASE AND WAS IMPLANTED WITH GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESES. THE TRUNK IPSILATERAL LEG COMPONENT DEPLOYED CORRECTLY BUT WAS POSITIONED SLIGHTLY HIGHER THAN INTENDED WITH PARTIAL COVERAGE OF THE OSTIUM OF A RENAL ARTERY. THE DEVICE WAS RECONSTRAINED AND THE PHYSICIAN ATTEMPTED TO REPOSITION BUT STRUGGLED WITH TRYING TO REPOSITION THE CXT AND CHOSE TO LEAVE IT AS IS AND IMPLANT A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX) COVERED STENT IN ONE OF THE RENAL ARTERIES TO ENSURE GOOD FLOW. DURING POST-DILITATION OF THE VBX WITH A CODA BALLOON CATHETER THE RENAL ARTERY WAS RUPTURED DUE TO BELIEVED OVER-INFLATION OF THE BALLOON. THE PATENT EXPIRED INTRA-PROCEDURE, A CAUSE OF DEATH WAS NOT PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12334625 ↗
  • InjuryJul 6, 2021Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    ON (B)(6) 2021 THE PATIENT UNDERWENT EMERGENT ENDOVASCULAR TREATMENT OF A RUPTURED ABDOMINAL AORTIC ANEURYSM USING GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESES. THERE WERE NO REPORTED ISSUES. ON (B)(6) 2021 IT WAS REPORTED BY THE PHYSICIAN THAT, DURING THE ONE MONTH FOLLOW-UP ON AN UNKNOWN DATE, CT IMAGING WAS PERFORMED AND POTENTIAL DEVICE OCCLUSION WAS OBSERVED. THE CAUSE OF THE SUSPECTED DEVICE OCCLUSION WAS UNKNOWN. ON (B)(6) 2021 A REINTERVENTION WAS PERFORMED. IT WAS REPORTED THAT A SLIGHT KINK ON THE IPSILATERAL LEG OF THE TRUNK-IPSILATERAL LEG COMPONENT WAS CONFIRMED. THE IPSILATERAL LIMB WAS LOCATED ON THE PATIENT'S RIGHT SIDE. REPORTEDLY THERE WAS NO DEVICE OCCLUSION, AND DISTAL BLOOD FLOW WAS PRESENT. IT WAS REPORTED THAT THE IPSILATERAL LIMB WAS SLIGHTLY PINCHED/KINKED DUE TO PATIENT ANATOMY. REPORTEDLY THE PATIENT'S AORTA MEASURED 15MM IN DIAMETER AND HAD A LARGE ANGLE. DUE TO THE PATIENT'S ANATOMY, THE IPSILATERAL LEG OF THE TRUNK-IPSILATERAL LEG COMPONENT WAS LANDED ON THE INNER CURVE OF THE PATIENT'S AORTA, THUS CAUSING A SLIGHT KINK. THERE WERE NO PATIENT ADVERSE EVENTS NOTED IN RELATION TO THE KINKED IPSILATERAL LIMB. A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12125098 ↗

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →