Skip to content

GORE TAG Thoracic Branch Endoprosthesis TAC123420ARecalls & reported problems

W. L. Gore & Associates, Inc. · TAG TBE AORTIC COMPONENT TAG TBE AC 12MMX34MMX20CM · Class 3 · Implant

All models of GORE TAG Thoracic Branch Endoprosthesis (30)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00733132654208), received between Apr 30, 2026 and Apr 30, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Activation Failure” (1 report).

1 other recall cover devices cleared under the same FDA submission (P210032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure1
  • Material Frayed1
  • Off-Label Use1

Patient problems

  • Unspecified Vascular Problem1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 30, 2026Material Frayed; Off-Label Use; Activation Failure

    THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2026, A PATIENT WAS TREATED FOR A ZONE 0 CHRONIC DISSECTION USING A PHYSICIAN MODIFIED GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS (TBE). THE DEVICE WAS MODIFIED TO CREATE FENESTRATIONS FOR THE LEFT COMMON CAROTID AND LEFT SUBCLAVIAN. IN THE PROCESS OF PERFORMING THIS MODIFICATION, THE PRIMARY SLEEVE WAS OPENED FOR A 3 CM SEGMENT ON THE SIDE OF THE DEVICE ADJACENT TO THE PORTAL. AFTER ADVANCING THE DEVICE TO THE TRANSVERSE THORACIC AORTA, THEY REALIZED THE DEVICE HAD WIRE WRAP WITH THE INNOMINATE WIRE. THEY ROTATED 2-3 TIMES COUNTER-CLOCKWISE, REALIZED THAT WAS THE WRONG DIRECTION, AND ROTATED THE OTHER WAY. AFTER 10 MINUTES, THE TBE WAS ABLE TO BE ADVANCED THROUGH THE SHEATH. ONCE IN POSITION, THEY PULLED THE DEPLOYMENT LINE. HOWEVER, THE DEVICE DID NOT DEPLOY AND REMAINED CONSTRAINED. THE PHYSICIAN TRIED TO SNARE THE PARTIALLY OPEN END OF THE TBE DEVICE, AND PULL IT BACK INTO THE DSF, BUT AFTER 45 MINUTES THEY ABANDONED THIS. THEY THEN REMOVED THE DSF AND TBE TOGETHER. THE RIGHT EXTERNAL AND COMMON ILIACS WERE DAMAGED BY THE GRAFT AND REPAIRED USING A PATCH ENDARTERECTOMY. THE PROCEDURE WAS ABORTED, AND NO DEVICES WERE IMPLANTED. THE PA

    Patient problems: Unspecified Vascular Problem

    MAUDE report 25049241 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P210032. One submission often covers many models, so check the product and lot details in each record.

  • Z-2672-2026 · May 19, 2026 · Open, Classified
    GORE TAG Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Region) United States (includes Puerto Rico) and Canada: TSB080806A, TSB081006A, TSB081206A, TSB08

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates