Skip to content

GORE TAG Thoracic Branch Endoprosthesis TE4043ARecalls & reported problems

W. L. Gore & Associates, Inc. · TAG TBE THORACIC AORTIC EXT 40mmX4.3cm 6mm · Class 3 · Implant

All models of GORE TAG Thoracic Branch Endoprosthesis (30)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00733132654338), received between Feb 16, 2024 and Apr 2, 2026 — classified as 5 injuries.

The most frequently reported specific device problem is “Malposition of Device” (1 report).

1 other recall cover devices cleared under the same FDA submission (P210032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Malposition of Device1
  • Migration1
  • Patient-Device Incompatibility1

Patient problems

  • Vascular Dissection2
  • Clinical Endoleak1
  • Obstruction/Occlusion1
  • Reduced Blood Flow1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 2, 2026Adverse Event Without Identified Device or Use Problem

    THE FOLLOWING WAS REPORTED: ON (B)(6) 2026, THE PATIENT WAS IMPLANTED WITH GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS AORTIC EXTENDER AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS) USING A GORE®¿DRYSEAL FLEX INTRODUCER SHEATH AND A GORE® TRI-LOBE BALLOON CATHETER TO TREAT AN ASCENDING THORACIC DISSECTION. REPORTEDLY DURING THE PROCEDURE WHILE RETRACTING THE DELIVERY CATHETER OF THE GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS AORTIC EXTENDER THE DEVICE MOVED DISTALLY AND PARTIALLY COVERED THE INNOMINATE ARTERY. THE PHYSICIAN IMPLANTED A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS IN THE INNOMINATE ARTERY TO MAINTAIN PATENCY. THEN THE PHYSICIAN IMPLANTED A SECOND GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS AORTIC EXTENDER PROXIMALLY. THE PATIENT TOLERATED THE PROCEDURE. POST PROCEDURE IMAGES REVEALED A TYPE B DISSECTION FROM IMPLANTED GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS AORTIC EXTENDER EXTENDING DOWN INTO THE ILIAC ARTERY. THE PHYSICIAN IS NOT TREATING THE DISSECTION AT THIS TIME. THE EVENT IS ONGOING. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Vascular Dissection

    MAUDE report 24770070 ↗
  • InjuryJan 2, 2026Migration

    THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2025, A PATIENT WAS TREATED FOR A DISSECTION USING A GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS. ON (B)(6) 2025, THE PATIENT WAS TREATED FOR A DISSECTION USING A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) SYSTEM, INCLUDING GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS DEVICES. ON FOLLOW-UP CT, THE PHYSICIAN NOTED A PROXIMAL ENDOLEAK ON THE TBE DEVICE. ON (B)(6) 2025, A REINTERVENTION OCCURRED. A TBE EXTENDER WAS PLACED PROXIMALLY OF THE TBE DEVICE. AFTER DEPLOYMENT, THE PHYSICIAN ATTEMPTED TO WITHDRAW THE CATHETER, BUT IT FELT LIKE IT WAS STUCK. THE PHYSICIAN PUSHED THE CATHETER FORWARD, AND WHEN THIS OCCURRED, THE DEPLOYED DEVICE ALSO PUSHED FORWARD, COVERING 80% OF THE INNOMINATE ARTERY. THE PHYSICIAN PERFORMED A LEFT SUBCLAVIAN - LEFT CAROTID BYPASS, AND THEN PLACED A VBX STENT IN THE INNOMINATE ARTERY. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Obstruction/Occlusion

    MAUDE report 23946006 ↗
  • InjuryJan 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2023, THE PATIENT WAS IMPLANTED WITH GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS (TBE AC) TO TREAT A THORACIC DISSECTION. THE PATIENT TOLERATED THE PROCEDURE. IT WAS REPORTED THAT ON (B)(6) 2024, THE PATIENT UNDERWENT A REINTERVENTION TO TREAT A PROXIMAL TYPE I ENDOLEAK. THE PATIENT TOLERATED THE PROCEDURE. THE FSA WAS NOT AT THIS PROCEDURE AND WAS NOT MADE AWARE OF A REINTERVENTION UNTIL (B)(6) 2024. CONSIGNMENT DEVICE WAS USED FOR THE PROCEDURE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Clinical Endoleak

    MAUDE report 21112840 ↗
  • InjuryApr 25, 2024Patient-Device Incompatibility; Patient-Device Incompatibility

    ON (B)(6) 2024, THE PATIENT WAS TREATED FOR A ZONE 2 THORACIC AORTIC DISSECTION. THE PHYSICIAN IMPLANTED A GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS AORTIC COMPONENT TAC084020A, TWO GORE® TAG® THORACIC BRANCH ENDOPROSTHESES SIDE BRANCHES, (TSB081506A), AND A GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS AORTIC EXTENDER TE4043A. THE PATIENT TOLERATED THE PROCEDURE. ON A FOLLOW UP VISIT, IMAGING REVEALED A DISSECTION PROXIMAL TO THE TE4043A. THE PHYSICIAN BELIEVES THE DAMAGE IS TOO EXTENSIVE FOR ENDOVASCULAR REPAIR, AND IS PLANNING AN OPEN CONVERSION. IMAGES WERE RETURNED FOR EVALUATION.

    Patient problems: Vascular Dissection

    MAUDE report 19188321 ↗
  • InjuryFeb 16, 2024Malposition of Device; Malposition of Device

    ON (B)(6) 2024, THE PATIENT WAS TREATED FOR A PAU. THE PHYSICIAN IMPLANTED A GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS AORTIC COMPONENT, A GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS SIDE BRANCH AND A GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS AORTIC BRANCH EXTENDER. WHEN THE EXTENDER WAS PLACED, IT ENDED UP PARTIALLY COVERING THE CAROTID ARTERY. BLOOD FLOW WAS RESTORED WITH A VBX, AND THE PATIENT TOLERATED THE PROCEDURE. THE CAUSE OF THE PLACEMENT ISSUE WAS BEING UNABLE TO SEE THE CAROTID DUE TO A PACEMAKER ON THE IMAGES.

    Patient problems: Reduced Blood Flow

    MAUDE report 18721350 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P210032. One submission often covers many models, so check the product and lot details in each record.

  • Z-2672-2026 · May 19, 2026 · Open, Classified
    GORE TAG Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Region) United States (includes Puerto Rico) and Canada: TSB080806A, TSB081006A, TSB081206A, TSB08

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates