GORE TAG Thoracic Branch Endoprosthesis TSB080806ARecalls & reported problems
W. L. Gore & Associates, Inc. · TAG TBE THORACIC SIDE BRANCH 8mmX6cm 8mm Heparin · Class 3 · Implant
All models of GORE TAG Thoracic Branch Endoprosthesis (30)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00733132654352), received between Jun 8, 2023 and Aug 21, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (P210032); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation, Positioning or Separation Problem1
- Complete Blockage1
- Improper or Incorrect Procedure or Method1
- Material Integrity Problem1
Patient problems
- Aneurysm2
- Reduced Blood Flow1
- Vascular Dissection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 21, 2025Material Integrity Problem; Material Integrity Problem
ON (B)(6) 2025, DURING AN ENDOVASCULAR PROCEDURE IT WAS REPORTED THAT POST DEPLOYMENT OF THE GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS (SIDE BRANCH COMPONENT TSB). WHEN TRYING TO WITHDRAW THE DELIVERY CATHETER, THE NOSE CONE CAME OFF. THE PHYSICIAN SNARED THE NOSE CONE AND IN THE PROCESS OF REMOVING THE ARTIFACT FROM THE PATIENT, THE GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS (TBE) WAS PULLED BACK AND SEAL WAS LOST AND A GORE® TAG® THORACIC BRANCH AORTIC EXTENDER WAS IMPLANTED. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 22866197 ↗InjuryJun 8, 2023Complete Blockage; Improper or Incorrect Procedure or Method; Activation, Positioning or Separation Problem; Complete Blockage; Improper or Incorrect Procedure or Method; Activation, Positioning or Separation Problem
ON (B)(6) 2023, THE PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF A ZONE 2 THORACIC AORTIC ANEURYSM USING GORE® TAG® THORACIC BRANCH ENDOPROSTHESES (TBE). IT WAS REPORTED THAT THE CASE WAS WITH A SIDE BRANCH INTO LEFT SUBCLAVIAN ARTERY (LSA). REPORTEDLY, THE ACCESS AND THE DEPLOYMENT OF THE TBE AORTIC COMPONENT AND TBE SIDE BRANCH TSB080806/26297332 WERE PERFORMED WITH NO REPORTED ISSUES. DURING THE BALLOONING OF THE SIDE BRANCH DEVICE WITH A CODA BALLOON, THE PROXIMAL END OF TSB WAS INFLATED PAST VESSEL PROFILE. A LSA ANGIOGRAPHY WAS PERFORMED, AND A RUPTURE WAS NOTICED. ACCORDING TO THE PHYSICIAN, THE OVER BALLOONING CAUSED THE VESSEL RUPTURE (CODED: 1400-E:-RUPTURE EVENTS - OTHER/UNKNOWN). A SECOND SIDE BRANCH (TSB080806/26265891) WAS INSERTED TO THE LEVEL OF THE VERTEBRAL ARTERY. THE DEPLOYMENT WAS AS EXPECTED. ANOTHER ANGIO SCAN WAS PERFORMED, AND THE RUPTURE WAS STILL PRESENT. A THIRD SIDE BRANCH (TSB080806/26265890) WAS DEPLOYED COVERING THE VERT TO THE LEVEL OF THE INTERNAL MAMMARY ARTERY (IMA). THE POST DEPLOYMENT ANGIOGRAPHY SHOWED NO RUPTURE WITH GOOD OUTFLOW. POST OP PATIENT HAD PALPABLE PULSE IN THE ARM AND NO SIGNS OF NEURO DEFICITS. REPORTEDLY, 14 HRS POST OP ( ~4 AM),
Patient problems: Aneurysm; Reduced Blood Flow; Vascular Dissection
MAUDE report 17094806 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P210032. One submission often covers many models, so check the product and lot details in each record.
- Z-2672-2026 · May 19, 2026 · Open, ClassifiedGORE TAG Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Region) United States (includes Puerto Rico) and Canada: TSB080806A, TSB081006A, TSB081206A, TSB08
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.