GORE TAG Conformable Thoracic Stent Graft TGMR403415Recalls & reported problems

W. L. Gore & Associates, Inc. · CTAG CMDS 40MMX34MMX15CM 22Fr · Class 3 · Implant

All models of GORE TAG Conformable Thoracic Stent Graft (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00733132661695), received between Nov 20, 2025 and May 15, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

6 other recalls cover devices cleared under the same FDA submission (P040043); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Obstruction/Occlusion1
  • Vascular Dissection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 15, 2026Obstruction of Flow

    THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2026, THE PATIENT UNDERWENT AN ENDOVASCULAR TAMBE PROCEDURE UTILIZING A GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM (CTAGAC), GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) AND A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX) WERE USED. REPORTEDLY, A CTAGAC WAS FIRST PLACED PROXIMAL TO THE CELIAC ARTERY TO ACHIEVE SEAL PRIOR TO DEPLOYMENT OF THE TAMBE DEVICE. FOLLOWING PLACEMENT OF THE TAMBE, AN INITIAL ANGIOGRAM DEMONSTRATED THAT THE CTAGAC APPEARED TO BE POSITIONED AT THE ORIGIN OF THE CELIAC ARTERY, WITH CONCERN FOR PARTIAL OCCLUSION. AS THE CELIAC PORTAL WAS ON THE SUPERIOR MESENTERIC ARTERY (SMA), A VBX WAS THEN PUT IN FROM THE CTAGAC AND COULD NO LONGER VISUALIZE THE CELIAC ARTERY. THE CELIAC ARTERY BECAME FULLY OCCLUDED WITH NO BLOOD FLOW. NO ADDITIONAL CLINICAL ISSUES RELATED TO THE OCCLUSION WERE REPORTED, AND THERE WERE NO PLANS TO REINTERVENE ON THE OCCLUDED CELIAC ARTERY. THE PHYSICIAN ATTRIBUTED THE CELIAC ARTERY OCCLUSION TO A COMBINATION OF THE CTAGAC AND VBX DEVICES.

    Patient problems: Obstruction/Occlusion

    MAUDE report 25183509 ↗
  • MalfunctionDec 12, 2025Break; Break

    THE FOLLOWING WAS SENT TO GORE: ON (B)(6) 2025, PATIENT UNDERWENT AN THORACIC ENDOVASCULAR AORTIC REPAIR (TEVAR) PROCEDURE TO TREAT A PENETRATING ATHEROSCLEROTIC ULCER UTILIZING A GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM. REPORTEDLY, DURING THE PROCEDURE, A TAPERED CTAG NOSE CONE BROKE OFF INSIDE THE PATIENT'S BODY AND THE GRAFT LANDED 4CM DISTAL TO LEFT SUBCLAVIAN ARTERY AND THEN DISTALLY EXTENDED WITH ANOTHER CTAG. PATIENT HAD A PREVIOUS MEDTRONIC GRAFT IN THE INFRARENAL THAT HAD SLIPPED DOWN WITH VERY TORTUOUS ANATOMY AND CALCIFIED ILIAC ARTERY. THE CTAG TRACKED FINE AND DEPLOYED FINE. ONCE THE CATHETER WAS BROUGHT OUT, IT WAS NOTED THAT THE NOSE CONE WAS MISSING OF THE DELIVERY CATHETER AND UPON CHECKING, IT WAS SITTING JUST OFF THE INFRARENAL ON THE GUIDEWIRE. PHYSICIAN PUT A LUNDERQUIST WITH SNARE OVER IT AND TRACK WHERE IT WOULD OPEN UP ABOVE IT, BUT THE SNARE WOULD NOT GO PAST DUE TO TORTUOSITY OF THE INFRANRENAL AORTA, SO A 2ND GUIDEWIRE WAS PUT UP AND SNARE CATHETER THROUGH AND THE NOSE CONE WOULD THEN MOVE OVER THE LUNDERQUIST WITH SNARE THRU THE CTAG THAT THEY DEPLOYED AND SNARED THE TOP OF THE LUNDERQUIST WIRE AND PULL IT OUT THRU THE SHEA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23794393 ↗
  • InjuryNov 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2025, THIS PATIENT UNDERWENT PLANNED/STAGED ENDOVASCULAR TREATMENT OF A ZONE 4 DESCENDING THORACIC AORTIC DISSECTION AND WAS IMPLANTED WITH TWO GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM DEVICES (MOST PROXIMAL DEVICE TGMR403415/30110178. THE PATIENT TOLERATED THE PROCEDURE. ON (B)(6) 2025, THIS PATIENT UNDERWENT A PLANNED/STAGED ENDOVASCULAR TREATMENT OF A ZONE 4 DESCENDING THORACIC AORTIC ANEURYSM USING A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) DEVICE, GORE® EXCLUDER® AAA ENDOPROSTHESIS DEVICES AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS DEVICES. THE PATIENT TOLERATED THE PROCEDURE. (CAPTURED AS A NON-COMPLAINT ON TAMBE IN (B)(4) . ON (B)(6) 2025, THE PATIENT PRESENTED TO THE HOSPITAL WITH BACK PAIN AND CHEST PAIN. IT WAS REPORTED THAT A CT SCAN WAS PERFORMED AND THE PATIENT IS BEING WORKED UP FOR A POSSIBLE RE-INTERVENTION. A CAUSED WAS NOT PROVIDED. IT WAS REPORTED THAT THE PATIENT¿S BLOOD PRESSURE WAS CONTROLLED AND THE ¿PAIN¿ HAS SUBSIDED. ON (B)(6) 2025, THE PATIENT WAS SCHEDULED FOR ENDOVASCULAR TREATMENT OF AN ACUTE/SUBACUTE TEVAR PROCEDURE FOR THE DISSECTION WITH GORE® TAG® THORACIC BRANCH ENDOPROSTH

    Patient problems: Vascular Dissection

    MAUDE report 23607092 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P040043. One submission often covers many models, so check the product and lot details in each record.

  • Z-0291-2021 · Sep 9, 2020 · Open, Classified
    GORE TAG Conformable Thoracic Stent Graft with ACTIVE CONTROL System
  • Z-0290-2021 · Sep 9, 2020 · Open, Classified
    GORE TAG CONFORMABLE Thoracic Stent Graft with ACTIVE CONTROL System
  • Z-2052-2020 · Dec 17, 2019 · Terminated
    Conformable Thoracic Stent Graft, Rx only, Sterile EO - Product Usage: This device is intended for endovascular repair of lesions of the descending thoracic aorta.
  • Z-0048-2018 · Sep 25, 2017 · Terminated
    GORE TAG, THORACIC ENDOPROSTHESIS, CONFORMABLE. The GORE TAG Thoracic Endoprosthesis is intended for endovascular repair of all lesions of the descending thoracic aorta.
  • Z-1236-06 · Apr 28, 2006 · Terminated
    GORE TAG THORACIC ENDOPROSTHESIS System, Endovascular Graft, Aortic Aneurysm Treatment, Catalogue Number TG3410, 34mm x 10cm
  • Z-1237-06 · Apr 28, 2006 · Terminated
    GORE TAG THORACIC ENDOPROSTHESIS System, Endovascular Graft, Aortic Aneurysm Treatment, Catalogue Number TG3415, 34mm x 15cm

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page