Skip to content

GORE® VIABAHN® FORTEGRA Venous Stent VNS181002ARecalls & reported problems

W. L. Gore & Associates, Inc. · FORTEGRA VENOUS STENT 18mm x 100mm · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00733132663828), received between Jul 28, 2026 and Jul 28, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Partial Blockage” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Partial Blockage1

Patient problems

  • Thrombosis/Thrombus1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 28, 2026Partial Blockage

    The following information was reported to Gore: On (b)(6) 2026 the patient underwent endovascular treatment of thrombosis in the inferior vena cava with a GORE® VIABAHN® FORTEGRA Venous Stent (VNS device). On (b)(6) 2026, thrombosis was identified within previously implanted VNS devices, and the physician performed a thrombectomy using a RevCore¿ Thrombectomy Catheter. During the thrombectomy procedure, one of the implanted VNS devices became entangled with the RevCore¿ Thrombectomy Catheter. To resolve the situation, the physician obtained alternate access from brachial artery, cut the stent wire, and removed the affected device. The previously implanted VNS devices were subsequently realigned with new VNS devices. The procedure was completed successfully, and no adverse patient consequences were reported.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 50062525 ↗

Other Stent, Vena Cava devices

All Stent, Vena Cava devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page