Skip to content

DeRoyal 53-1831Recalls & reported problems

Deroyal Industries, Inc. · ORTHO TOTAL JOINT TRACECART® · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00749756598088), received between Oct 3, 2024 and Oct 3, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 13, 2024Class IOpen, Classified

    DeRoyal manufactured and distributed Surgical Tracecarts that contain a Nurse Assist 16FR Silver Urine Meter Foley which were subsequently recalled.

    Z-1382-2024 · Root cause: Other · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 3, 2024Break; Break

    A YANKAUER SUCTION BROKE DURING A SURGERY IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20358628 ↗

FDA clearance and classification

The FDA regulates product code LRO (general surgery tray) under 21 CFR 878.4370 as a Class II device. 21 CFR 878.4370 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the DeRoyal 53-1831

Has the DeRoyal 53-1831 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1382-2024 (Class I), began on Feb 13, 2024: DeRoyal manufactured and distributed Surgical Tracecarts that contain a Nurse Assist 16FR Silver Urine Meter Foley which were subsequently recalled. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the DeRoyal 53-1831?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Oct 3, 2024. That report lists device problem “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the DeRoyal 53-1831?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the DeRoyal 53-1831, and is it still sold?

The FDA's device database lists Deroyal Industries, Inc. as the company responsible for it (brand DeRoyal, model 53-1831). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other General surgery tray devices

All General surgery tray devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page