DeRoyal 89-9338Recalls & reported problems
Deroyal Industries, Inc. · C-SECTION TRACEPACK · Class 1
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00749756927864), received between Nov 13, 2018 and Jan 11, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Arcing” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Oct 3, 2024Open, Classified
DeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual input when plugged into a power source or remains active after buttons are released.
Z-0306-2025 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Arcing2
- Electrical /Electronic Property Problem2
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 11, 2019Electrical /Electronic Property Problem; Arcing; Electrical /Electronic Property Problem; Arcing
A CONVENIENCE KIT CONTAINED A CAUTERY PENCIL THAT MALFUNCTIONED DURING USE AND REPORTEDLY HAD AN ELECTRICAL ARC.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8239717 ↗MalfunctionNov 13, 2018Electrical /Electronic Property Problem; Arcing; Electrical /Electronic Property Problem; Arcing
A CONVENIENCE KIT CONTAINED A CAUTERY PENCIL THAT ARCED DURING USE. IT WAS CONNECTED TO A BOVIE GENERATOR. NO INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8063734 ↗
Other Kit, surgical instrument, disposable devices
- Medline Industries, Inc. DYNJ83185AMedline Industries, Inc. · Kit, surgical instrument, disposable2 recalls0 reports
- CustomEyes® 58001713Beaver-Visitec International, Inc. · Kit, surgical instrument, disposable2 recalls0 reports
- Medline Industries, Inc. DYNJ69761Medline Industries, Inc. · Kit, surgical instrument, disposable2 recalls0 reports
- Synergetics 18069WSynergetics USA, Inc. · Kit, Surgical Instrument, Disposable2 recalls0 reports
- Synergetics 18060Synergetics USA, Inc. · Kit, Surgical Instrument, Disposable2 recalls0 reports
- Medline Industries, Inc. DYNJ68217AMedline Industries, Inc. · Kit, surgical instrument, disposable2 recalls0 reports
- Medline Industries, Inc. DYNJ80200BMedline Industries, Inc. · Kit, surgical instrument, disposable2 recalls0 reports
- Medline Industries, Inc. DYNJ39461JMedline Industries, Inc. · Kit, surgical instrument, disposable2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.