Skip to content

NIM Vital NIM4CPB1RFRecalls & reported problems

Medtronic Xomed, Inc. · PATIENT INTFC NIM4CPB1RF NIM 4.0 REFURB · Class 2

All models of NIM Vital (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000003005), received between Dec 26, 2025 and Dec 26, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

2 other recalls cover devices cleared under the same FDA submission (K200759); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 10, 2025Class IIOpen, Classified

    Nerve Monitoring System with certain software versions has potential for increased stimulus artifact, which may require troubleshooting, and may lead to procedure delay, extubation, and medical intervention.

    Z-1623-2025 · Root cause: Software change control · Recall details · FDA record ↗

  2. Jun 24, 2024Class IOpen, Classified

    Potential for false negative response on the NIM Vital Nerve Monitoring System

    Z-2409-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 26, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT INTRA-OP, THERE WAS PATIENT INTERFACE FAULT DETECTED MESSAGE AND WOULD NOT STOP OR GO AWAY. THE MESSAGE KEPT DI SPLAYING. THE CONSOLE WORKED FINE WITH ANOTHER PI BOX AND THE PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23894746 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K200759. One submission often covers many models, so check the product and lot details in each record.

  • Z-1223-2022 · May 3, 2022 · Open, Classified
    NIM Vital Console 4.0 (P/N: NIM4CM01) NIM with Software version 1.3.2
  • Z-1224-2022 · May 3, 2022 · Open, Classified
    NIM Vital Patient Interface 4.0 (P/N: NIM4CPB1) NIM with Software version 1.3.2

Other STIMULATOR, NERVE devices

All STIMULATOR, NERVE devices →

Sources and updates