KYPHON Express II Inflatable Bone Tamp KE102Recalls & reported problems
Medtronic Sofamor Danek, Inc. · BONE TAMP KE102 10/2 EXPRESS II IBT · Class 1
All models of KYPHON Express II Inflatable Bone Tamp (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000025632), received between Jan 15, 2019 and Sep 20, 2019 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Rupture2
- Inflation Problem1
- Leak/Splash1
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 20, 2019Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash
IT WAS REPORTED THAT THE PATIENT UNDERWENT A SURGERY DUE TO ACUTE VERTEBRAL COMPRESSION FRACTURE. INTRA-OP, IT WAS FOUND THAT THE BALLOON WHICH WAS INSERTED IN THE VERTEBRAL BODY DID NOT INFLATE. DISTAL TIP OF BALLOON LEAKED CONTRAST SLOWLY. PRESSURE WAS NOT VERY HIGH AND APPEARED ON IMAGING THAT IT WOULD NOT HOLD INFLATION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THE ORIGINAL PRODUCT. NO PATIENT COMPLICATION WAS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9096372 ↗MalfunctionMay 29, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PATIENT UNDERWENT KYPHOPLASTY AT T9 DUE TO FALL CAUSES FRACTURE. INTRA-OP, AFTER RE-DRILLING IBT WAS RE-INSERTED, AND ON INFLATION IT GOT BURST. PRESSURE PRIOR TO RUPTURE WAS 550 PSI. VOLUME PRIOR TO RUPTURE WAS 1 CC. SURGEON REMOVED CONTRAST FROM VERTEBRAL BODY BY USING CONTRAST SYRINGE. SURGEON WAS UNABLE TO REMOVE IBT THROUGH CANNULA THEN TRIED TO FIX THE STYE LEAD. FINALLY MOVED THE HANDLE OF THE CANNULA AROUND IN DIFFERENT ANGLES REMOVED THE BALLOON AND CANNULA TOGETHER. NO FRAGMENT OF THE PRODUCT REMAINING IN THE PATIENT. THERE WAS A DELAY OF APPROXIMATELY 10-15 MIN IN PROCEDURE TIME AS A RESULT OF EVENT. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8649007 ↗MalfunctionJan 15, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PATIENT WAS PRE-OPERATIVELY DIAGNOSED WITH COMPRESSION FRACTURE AND UNDERWENT BALLOON KYPHOPLASTY (BKP) SURGERY. DURING SURGERY, THE INFLATED BALLOON TAMP (IBT) RUPTURED WHILE INFLATING AT 400PSI. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8251637 ↗
Other TAMP devices
- KYPHON® Express™ Inflatable Bone Tamp KE152Medtronic Sofamor Danek, Inc. · TAMP35 reports
- KYPHON Express II Inflatable Bone Tamp KE152Medtronic Sofamor Danek, Inc. · TAMP13 reports
- KYPHON Xpander™ II Inflatable Bone Tamp KX203Medtronic Sofamor Danek, Inc. · TAMP11 reports
- KYPHON® Express™ Inflatable Bone Tamp KE102Medtronic Sofamor Danek, Inc. · TAMP10 reports
- KYPHON Xpander™ II Inflatable Bone Tamp KX153Medtronic Sofamor Danek, Inc. · TAMP9 reports
- NA 356.301Synthes GmbH · TAMP6 reports
- KYPHON Xpander II Inflatable Bone Tamp KX203Medtronic Sofamor Danek, Inc. · TAMP6 reports
- RECLAIM 2975-00-675Depuy (Ireland) · TAMP5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page