Strata™ 92866 (UDI 00763000027322)Recalls & reported problems
Medtronic Ps Medical, Inc. · VALVE 92866 FP STRATA 2 REG BG · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00763000027322), received between Mar 18, 2021 and Nov 14, 2025 — classified as 7 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (9 reports).
2 other recalls cover devices cleared under the same FDA submission (K060681); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak9
- Break4
- Fracture4
- Complete Blockage3
- Infusion or Flow Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Insufficient Information4
- Unspecified Tissue Injury2
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 14, 2025Break; Fluid/Blood Leak; Fracture; Break; Fluid/Blood Leak; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE VALVE. IT WAS REPORTED THAT THE DEVICE WAS LEAKING AND WAS SUSPECTED TO BE BROKEN. THE DEVICE WAS REPLACED WITH A NEW ONE. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23554439 ↗MalfunctionNov 14, 2025Break; Fluid/Blood Leak; Fracture; Break; Fluid/Blood Leak; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE VALVE. IT WAS REPORTED THAT THE DEVICE WAS LEAKING LIQUID AND WAS SUSPECTED TO BE DAMAGED. THE DEVICE WAS REPLACED WITH A NEW ONE. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23554301 ↗MalfunctionAug 6, 2025Break; Fluid/Blood Leak; Fracture; Break; Fluid/Blood Leak; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE VALVE. IT WAS REPORTED THAT DURING THE PRE-INFUSION TEST BEFORE IMPLANTATION; IT WAS FOUND THAT THE VALVE WAS DAMAGED AND LEAKING. THE DEVICE WAS REPLACED WITH A NEW ONE. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22725336 ↗InjuryDec 10, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE VALVE. IT WAS REPORTED THAT ON (B)(6) 2024, THE CLINICIANS PERFORMED A V ENTRICULOPERITONEAL SHUNT SURGERY. AFTER THE OPERATION WAS COMPLETED NORMALLY, THE PATIENT RETURNED TO THE WARD AND FOUND THAT THE WOUND WAS LEAKING. THE CLINICIANS URGENTLY REMOVED THE LEAKING VALVE AND REPLACED IT WITH A NEW ONE. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20891783 ↗InjuryNov 7, 2024Complete Blockage; Complete Blockage
MEDTRONIC RECEIVED INFORMATION REGARDING A SHUNT. IT WAS REPORTED THAT AFTER THE SHUNT WAS INSERTED, THE PATIENT EXPERIENCED POOR DRAINAGE, AND THE SYMPTOMS OF HYDROCEPHALUS DID NOT IMPROVE. AFTER INVESTIGATING, IT WAS FOUND THAT THE SHUNT WAS BLOCKED. A SECOND SURGERY WAS PERFORMED TO REPLACE WITH A NEW SHUNT. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.
Patient problems: Insufficient Information
MAUDE report 20629841 ↗InjuryOct 14, 2024Complete Blockage; Complete Blockage
MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE SHUNT. IT WAS REPORTED THAT THE PATIENT'S HYDROCEPHALUS DID NOT IMPROVE SIGNIFICANTLY AFTER THE SHUNT SURGERY. IT WAS LATER DISCOVERED THAT THE SHUNT WAS BLOCKED. A SECOND SURGERY WAS PERFORMED TO REPLACE THE SHUNT. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.
Patient problems: Insufficient Information
MAUDE report 20438605 ↗MalfunctionAug 22, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE VALVE. IT WAS REPORTED THAT DURING THE OPERATION, THE BASE OF THE VALVE NEAR THE ABDOMINAL CAVITY WAS LEAKING LIQUID. THE VALVE WAS REPLACED WITH A NEW ONE TO COMPLETE THE OPERATION. THERE WAS A PROCEDURE DELAY OF 10 MINUTES. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. THE PATIENT'S MEDICAL HISTORY WAS HYDROCEPHALUS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20032245 ↗InjuryAug 12, 2024Infusion or Flow Problem; Infusion or Flow Problem
MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE SHUNT. IT WAS REPORTED THAT THE SHUNT WAS IMPLANTED IN 2022, INITIALLY AT 1.5. THERE WAS A CASE OF SHUNT FAILURE DURING THE INITIAL IMPLANTATION, BUT THE CONDITION STABILIZED AFTER REPEATED PRESSURE ADJUSTMENTS. IN (B)(6) 2024 OF THIS YEAR, THE PATIENT CAME TO THE HOSPITAL WITH THE GEAR REDUCED TO 0.5, HOWEVER, THE SHUNT WAS STILL INSUFFICIENT. AFTER REPLACING WITH A NEW SHUNT, THE SYMPTOM OF HYDROCEPHALUS WAS RELIEVED. THE PATIENT'S STATUS AT THE TIME OF REPORT WAS ALIVE-NO INJURY.
Patient problems: Insufficient Information
MAUDE report 19959665 ↗MalfunctionAug 5, 2024Break; Fracture; Break; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING A VALVE/SHUNT. IT WAS REPORTED THAT THE SURGEON FOUND THE VALVE WAS BROKEN WHEN CHECKING THE PATENCY BEFORE IMPLANTATION AND REPORTED A QUALITY PROBLEM. A DIFFERENT PRODUCT WAS USED. THERE WAS NO IMPACT TO PATIENT OUTCOME AS THE DEVICE WAS NEVER IMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19906355 ↗MalfunctionJan 26, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE VALVE. IT WAS REPORTED THAT DURING THE OPERATION, VALVE LEAKAGE WAS FOUND WHEN CHECKING THE PATENCY BEFORE THE VALVE WAS IMPLANTED. THE VALVE WAS REPLACED. THERE WAS A PROCEDURE DELAY OF 60 MINUTES. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE - NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18587487 ↗InjuryJul 19, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION REGARDING A SHUNT. IT WAS REPORTED THAT 4 MONTHS AFTER CEREBRAL HEMORRHAGE, THE VP SHUNT OPERATION WAS PERFORMED ON JULY-17, AND LIQUID LEAKAGE OF SHUNT VALVE WAS FOUND DURING THE OPERATION. THE VALVE WAS REPLACED AND RE-IMPLANTED. THERE WAS AN EXTENDED PROCEDURE TIME OF 120 MINUTES. THE PATIENT WAS NOTED TO HAVE INJURY.
Patient problems: Unspecified Tissue Injury
MAUDE report 17350292 ↗InjurySep 29, 2022Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION REGARDING A SHUNT BEING IMPLANTED DURING A HYDROCEPHALUS CASE. IT WAS REPORTED THE VALVE LEAKED DURING IMPLANTATION. THE DEVICE WAS EXPLANTED AND REPLACED AND THE ISSUE WAS RESOLVED. THERE WAS NO INJURY TO THE PATIENT AND ABOUT A 20 MINUTE DELAY.
Patient problems: Unspecified Tissue Injury
MAUDE report 15501716 ↗MalfunctionSep 1, 2021Fluid/Blood Leak; Fluid/Blood Leak
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THERE MAYBE WAS A MORE HARSH MECHANICAL FORCE TO THE VALVE BY THE PHYSICIAN DURING THE VALVE CHECK THAT LEAD TO THE LEAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12407992 ↗InjuryMar 18, 2021Complete Blockage; Complete Blockage
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THERE WERE NO ENVIRONMENTAL/EXTERNAL/PATIENT FACTORS THAT MAY HAVE LED OR CONTRIBUTED TO THE ISSUE.
Patient problems: Insufficient Information
MAUDE report 11517318 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K060681. One submission often covers many models, so check the product and lot details in each record.
- Z-1116-2021 · Jan 11, 2021 · Open, ClassifiedMedtronic Cerebro Spinal Fluid Flow Control Valves and Shunts. STRATA II Valves and Shunts for the following CFNs: 27739 - SHUNT 27739 STRATA NSC SNAP SML 120CM, 27740 - SHU
- Z-1121-2017 · Jan 17, 2017 · TerminatedMedtronic Strata II / Strata NSC Valves (includes Burr Hole and Lumbar Peritoneal. All Catalog Numbers. Strata-type valves are used in the management of hydrocephalus. The Strata
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page