Strata™ 92866 (UDI 00763000027322)Recalls & reported problems

Medtronic Ps Medical, Inc. · VALVE 92866 FP STRATA 2 REG BG · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00763000027322), received between Mar 18, 2021 and Nov 14, 2025 — classified as 7 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (9 reports).

2 other recalls cover devices cleared under the same FDA submission (K060681); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak9
  • Break4
  • Fracture4
  • Complete Blockage3
  • Infusion or Flow Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Insufficient Information4
  • Unspecified Tissue Injury2

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2025Break; Fluid/Blood Leak; Fracture; Break; Fluid/Blood Leak; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE VALVE. IT WAS REPORTED THAT THE DEVICE WAS LEAKING AND WAS SUSPECTED TO BE BROKEN. THE DEVICE WAS REPLACED WITH A NEW ONE. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23554439 ↗
  • MalfunctionNov 14, 2025Break; Fluid/Blood Leak; Fracture; Break; Fluid/Blood Leak; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE VALVE. IT WAS REPORTED THAT THE DEVICE WAS LEAKING LIQUID AND WAS SUSPECTED TO BE DAMAGED. THE DEVICE WAS REPLACED WITH A NEW ONE. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23554301 ↗
  • MalfunctionAug 6, 2025Break; Fluid/Blood Leak; Fracture; Break; Fluid/Blood Leak; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE VALVE. IT WAS REPORTED THAT DURING THE PRE-INFUSION TEST BEFORE IMPLANTATION; IT WAS FOUND THAT THE VALVE WAS DAMAGED AND LEAKING. THE DEVICE WAS REPLACED WITH A NEW ONE. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22725336 ↗
  • InjuryDec 10, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE VALVE. IT WAS REPORTED THAT ON (B)(6) 2024, THE CLINICIANS PERFORMED A V ENTRICULOPERITONEAL SHUNT SURGERY. AFTER THE OPERATION WAS COMPLETED NORMALLY, THE PATIENT RETURNED TO THE WARD AND FOUND THAT THE WOUND WAS LEAKING. THE CLINICIANS URGENTLY REMOVED THE LEAKING VALVE AND REPLACED IT WITH A NEW ONE. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20891783 ↗
  • InjuryNov 7, 2024Complete Blockage; Complete Blockage

    MEDTRONIC RECEIVED INFORMATION REGARDING A SHUNT. IT WAS REPORTED THAT AFTER THE SHUNT WAS INSERTED, THE PATIENT EXPERIENCED POOR DRAINAGE, AND THE SYMPTOMS OF HYDROCEPHALUS DID NOT IMPROVE. AFTER INVESTIGATING, IT WAS FOUND THAT THE SHUNT WAS BLOCKED. A SECOND SURGERY WAS PERFORMED TO REPLACE WITH A NEW SHUNT. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: Insufficient Information

    MAUDE report 20629841 ↗
  • InjuryOct 14, 2024Complete Blockage; Complete Blockage

    MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE SHUNT. IT WAS REPORTED THAT THE PATIENT'S HYDROCEPHALUS DID NOT IMPROVE SIGNIFICANTLY AFTER THE SHUNT SURGERY. IT WAS LATER DISCOVERED THAT THE SHUNT WAS BLOCKED. A SECOND SURGERY WAS PERFORMED TO REPLACE THE SHUNT. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: Insufficient Information

    MAUDE report 20438605 ↗
  • MalfunctionAug 22, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE VALVE. IT WAS REPORTED THAT DURING THE OPERATION, THE BASE OF THE VALVE NEAR THE ABDOMINAL CAVITY WAS LEAKING LIQUID. THE VALVE WAS REPLACED WITH A NEW ONE TO COMPLETE THE OPERATION. THERE WAS A PROCEDURE DELAY OF 10 MINUTES. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. THE PATIENT'S MEDICAL HISTORY WAS HYDROCEPHALUS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20032245 ↗
  • InjuryAug 12, 2024Infusion or Flow Problem; Infusion or Flow Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE SHUNT. IT WAS REPORTED THAT THE SHUNT WAS IMPLANTED IN 2022, INITIALLY AT 1.5. THERE WAS A CASE OF SHUNT FAILURE DURING THE INITIAL IMPLANTATION, BUT THE CONDITION STABILIZED AFTER REPEATED PRESSURE ADJUSTMENTS. IN (B)(6) 2024 OF THIS YEAR, THE PATIENT CAME TO THE HOSPITAL WITH THE GEAR REDUCED TO 0.5, HOWEVER, THE SHUNT WAS STILL INSUFFICIENT. AFTER REPLACING WITH A NEW SHUNT, THE SYMPTOM OF HYDROCEPHALUS WAS RELIEVED. THE PATIENT'S STATUS AT THE TIME OF REPORT WAS ALIVE-NO INJURY.

    Patient problems: Insufficient Information

    MAUDE report 19959665 ↗
  • MalfunctionAug 5, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING A VALVE/SHUNT. IT WAS REPORTED THAT THE SURGEON FOUND THE VALVE WAS BROKEN WHEN CHECKING THE PATENCY BEFORE IMPLANTATION AND REPORTED A QUALITY PROBLEM. A DIFFERENT PRODUCT WAS USED. THERE WAS NO IMPACT TO PATIENT OUTCOME AS THE DEVICE WAS NEVER IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19906355 ↗
  • MalfunctionJan 26, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE VALVE. IT WAS REPORTED THAT DURING THE OPERATION, VALVE LEAKAGE WAS FOUND WHEN CHECKING THE PATENCY BEFORE THE VALVE WAS IMPLANTED. THE VALVE WAS REPLACED. THERE WAS A PROCEDURE DELAY OF 60 MINUTES. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE - NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18587487 ↗
  • InjuryJul 19, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING A SHUNT. IT WAS REPORTED THAT 4 MONTHS AFTER CEREBRAL HEMORRHAGE, THE VP SHUNT OPERATION WAS PERFORMED ON JULY-17, AND LIQUID LEAKAGE OF SHUNT VALVE WAS FOUND DURING THE OPERATION. THE VALVE WAS REPLACED AND RE-IMPLANTED. THERE WAS AN EXTENDED PROCEDURE TIME OF 120 MINUTES. THE PATIENT WAS NOTED TO HAVE INJURY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17350292 ↗
  • InjurySep 29, 2022Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING A SHUNT BEING IMPLANTED DURING A HYDROCEPHALUS CASE. IT WAS REPORTED THE VALVE LEAKED DURING IMPLANTATION. THE DEVICE WAS EXPLANTED AND REPLACED AND THE ISSUE WAS RESOLVED. THERE WAS NO INJURY TO THE PATIENT AND ABOUT A 20 MINUTE DELAY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 15501716 ↗
  • MalfunctionSep 1, 2021Fluid/Blood Leak; Fluid/Blood Leak

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THERE MAYBE WAS A MORE HARSH MECHANICAL FORCE TO THE VALVE BY THE PHYSICIAN DURING THE VALVE CHECK THAT LEAD TO THE LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12407992 ↗
  • InjuryMar 18, 2021Complete Blockage; Complete Blockage

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THERE WERE NO ENVIRONMENTAL/EXTERNAL/PATIENT FACTORS THAT MAY HAVE LED OR CONTRIBUTED TO THE ISSUE.

    Patient problems: Insufficient Information

    MAUDE report 11517318 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K060681. One submission often covers many models, so check the product and lot details in each record.

  • Z-1116-2021 · Jan 11, 2021 · Open, Classified
    Medtronic Cerebro Spinal Fluid Flow Control Valves and Shunts. STRATA II Valves and Shunts for the following CFNs: 27739 - SHUNT 27739 STRATA NSC SNAP SML 120CM, 27740 - SHU
  • Z-1121-2017 · Jan 17, 2017 · Terminated
    Medtronic Strata II / Strata NSC Valves (includes Burr Hole and Lumbar Peritoneal. All Catalog Numbers. Strata-type valves are used in the management of hydrocephalus. The Strata

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →