Delta® 25132-5 (UDI 00763000027940)Recalls & reported problems

Medtronic Ps Medical, Inc. · SHUNT 25132-5 INTL ONLY DELTA REG 1.5 · Class 2 · Implant

All models of Delta (31)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00763000027940), received between Mar 20, 2024 and Sep 4, 2025 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Complete Blockage” (2 reports).

5 other recalls cover devices cleared under the same FDA submission (K902783); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Complete Blockage2
  • Break1
  • Disconnection1
  • Fluid/Blood Leak1
  • Fracture1
  • Infusion or Flow Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Apnea2
  • Convulsion/Seizure2
  • Hydrocephalus2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjurySep 4, 2025Disconnection; Disconnection

    MEDTRONIC RECEIVED INFORMATION REGARDING A SHUNT. IT WAS REPORTED THAT AFTER SURGERY, THE ABDOMINAL END TUBE WAS FOUND TO BE FALLEN OFF. THE TUBE WAS REPLACED WITH A NEW ONE. THERE WAS A PROCEDURE DELAY OF 30 MINUTES. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22963053 ↗
  • InjurySep 3, 2025Complete Blockage; Complete Blockage

    MEDTRONIC RECEIVED INFORMATION REGARDING A VALVE/SHUNT. IT WAS REPORTED THAT DESPITE THE PRESENCE OF PRESSURED CEREBROSPINAL FLUID (CSF), WHEN THE PROXIMAL SIDE OF THE SHUNT WAS IMPLANTED IN THE PATIENT, NO CSF FLOW WAS OBSERVED FROM THE DISTAL END WHEN THE PUMP AND DISTAL END WERE CONNECTED. EVEN WHEN THE TIP WAS LOWERED BELOW THE VENTRICLE LEVEL, NO CSF FLOW WAS OBSERVED DESPITE THE FLOW OF CSF BY PUMPING. THE PATIENT WAS PLACED ON DRAINAGE AND AN ADJUSTABLE SHUNT WAS IMPLANTED. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE PATIENT DEVELOPED HYDROCEPHALUS BECAUSE THE SHUNT DID NOT FUNCTION AFTER THE PROCEDURE. THE PATIENT EXPERIENCED EPILEPTIC SEIZURES AND SPONTANEOUS RESPIRATORY LOSS AND WERE CONNECTED TO MECHANICAL VENTILATORS, THAT IN PATIENT UNDERGOING REVISION, AN EVD HAD TO BE INSERTED, ON GROUNDS THE SHUNT WAS SEEN TO BE FUNCTIONING BY PUMPING.

    Patient problems: Apnea; Convulsion/Seizure; Hydrocephalus

    MAUDE report 22955589 ↗
  • InjuryAug 19, 2025Complete Blockage; Complete Blockage

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE PATIENT DEVELOPED HYDROCEPHALUS BECAUSE THE SHUNT DID NOT FUNCTION AFTER THE PROCEDURE. THE PATIENT EXPERIENCED EPILEPTIC SEIZURES AND SPONTANEOUS RESPIRATORY LOSS AND WERE CONNECTED TO MECHANICAL VENTILATORS, THAT IN PATIENT UNDERGOING REVISION, AN EVD HAD TO BE INSERTED, ON GROUNDS THE SHUNT WAS SEEN TO BE FUNCTIONING BY PUMPING.

    Patient problems: Apnea; Convulsion/Seizure; Hydrocephalus; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22839506 ↗
  • MalfunctionJan 21, 2025Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING A VALVE. IT WAS REPORTED THAT THE VALVE PRESENTED FAILURES AS THE RESERVOIR HAD A HOLE. THE PROCEDURE WAS COMPLETED WITH BACK UP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21196507 ↗
  • MalfunctionMay 6, 2024Infusion or Flow Problem; Infusion or Flow Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A VALVE. IT WAS REPORTED THAT BEFORE THE PRODUCT WAS IMPLANTED, THE SURGEON CONDUCT A SHUNT PATENCY TEST AND FOUND THAT THE SHUNT COULD NOT FLOW EVEN UNDER A PRESSURE OF 200 MM OF WATER AND COULD ONLY FLOW WHEN THE FLUID RESERVOIR BAG WAS PRESSED. THE SURGEON THOUGHT THE ACTUAL PRESSURE WAS MUCH GREATER THAN THE DESIGN. THE SURGERY WAS SUCCESSFULLY COMPLETED AFTER REPLACING WITH ANOTHER SET OF THE SAME PRODUCT. THERE WAS A 30 MINUTE DELAY IN THE PROCEDURE. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19247679 ↗
  • MalfunctionMar 20, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING A VALVE. IT WAS REPORTED THAT WHEN UNPACKING FOR PRE-PERFUSION, IT WAS FOUND THAT THE SHUNT BODY WAS LEAKING. THE DEVICE WAS REPLACED WITH A BACKUP DEVICE OF THE SAME MODEL. THE PATIENT'S STATUS WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18943387 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K902783. One submission often covers many models, so check the product and lot details in each record.

  • Z-1115-2021 · Jan 11, 2021 · Open, Classified
    Medtronic Delta CSF (Cerebro Spinal Fluid) -Flow Control Valves for the following CFNs/REF numbers: 24048 - VALVE 24048 DELTA RGL PL 0.5, 24053 - CHAMBER 24053 DELTA, 24147 -
  • Z-0871-2021 · Dec 16, 2020 · Open, Classified
    REF 27219-1 X1, Medtronic, Delta Valve, Neonatal, Performance Level 1.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Contr
  • Z-0872-2021 · Dec 16, 2020 · Open, Classified
    REF 27219-2 X1, Medtronic, Delta Valve, Neonatal, Performance Level 2.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Contr
  • Z-0873-2021 · Dec 16, 2020 · Open, Classified
    REF 27219-5 X1, Medtronic, Delta Valve, Neonatal, Performance Level 1.5, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Contr
  • Z-0398-2014 · Nov 18, 2013 · Terminated
    Medtronic Delta Shunt Assembly Kit with Bioglide Small, Performance Level 1.0, Catalog No. 276201. Delta Shunt Assemblies are designed for use in shunting CSF from the lateral ve

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →