Delta® 42813 (UDI 00763000028367)Recalls & reported problems
Medtronic Ps Medical, Inc. · VALVE 42813 DELTA SMALL LEVEL 1.5 · Class 2 · Implant
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00763000028367), received between Jan 28, 2025 and May 15, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Complete Blockage” (3 reports).
5 other recalls cover devices cleared under the same FDA submission (K902783); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Complete Blockage3
- Break1
- Fracture1
- Infusion or Flow Problem1
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Dizziness1
- Headache1
- Insufficient Information1
- Intracranial Hemorrhage1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 15, 2026Complete Blockage
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE DOCTOR WAS CONFUSED, AND THE FAULTY PRODUCT WAS NOT A MEDTRONIC PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25182622 ↗MalfunctionJan 22, 2026Complete Blockage
MEDTRONIC RECEIVED INFORMATION REGARDING A VALVE. IT WAS REPORTED THAT THE VALVE WAS OCCLUDED AT THE TIME OF VALVE OCCLUSION TESTING. WHEN INJECTING TO CHECK WHETHER THE VALVE SYSTEM WAS PATENT, THE PHYSICIAN FOUND THAT INJECTION WAS VERY DIFFICULT. WHEN THE CATHETER WAS CONNECTED, NO FLUID FLOWED DESPITE VERY HIGH PRESSURE. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24147974 ↗MalfunctionJan 16, 2026Complete Blockage
MEDTRONIC RECEIVED INFORMATION REGARDING A VALVE. IT WAS REPORTED THE VALVE WAS OBSTRUCTED AT THE TIME OF THE PATENCY TEST. WHEN INJECTING TO CHECK WHETHER THE VALVE SYSTEM WAS PATENT, THE PHYSICIAN NOTED VERY HIGH RESISTANCE. WHEN CONNECTED TO THE CATHETER, NO FLUID FLOWED DESPITE VERY HIGH PRESSURE. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24092493 ↗InjuryAug 22, 2025Infusion or Flow Problem; Insufficient Device Problem Information; Infusion or Flow Problem; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION REGARDING A VALVE. IT WAS REPORTED THAT AFTER THE VALVE IMPLANTATION IN (B)(6), THE PATIENT DID NOT IMPROVE AS THE VALVE WAS NOT FUNCTIONAL. ANOTHER EMERGENCY PROCEDURE WAS PERFORMED TO REPLACE THE VALVE WITH AN ADJUSTABLE VALVE.
Patient problems: Dizziness; Headache; Insufficient Information; Intracranial Hemorrhage
MAUDE report 22878082 ↗MalfunctionJan 28, 2025Break; Fracture; Break; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING A VALVE. IT WAS REPORTED DURING THE SHUNT PLACEMENT OPERATION, A HOLE WAS DETECTED IN THE PRODUCT. THE PROCEDURE WAS COMPLETED WITH BACK UP PRODUCTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21252026 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K902783. One submission often covers many models, so check the product and lot details in each record.
- Z-1115-2021 · Jan 11, 2021 · Open, ClassifiedMedtronic Delta CSF (Cerebro Spinal Fluid) -Flow Control Valves for the following CFNs/REF numbers: 24048 - VALVE 24048 DELTA RGL PL 0.5, 24053 - CHAMBER 24053 DELTA, 24147 -
- Z-0872-2021 · Dec 16, 2020 · Open, ClassifiedREF 27219-2 X1, Medtronic, Delta Valve, Neonatal, Performance Level 2.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Contr
- Z-0873-2021 · Dec 16, 2020 · Open, ClassifiedREF 27219-5 X1, Medtronic, Delta Valve, Neonatal, Performance Level 1.5, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Contr
- Z-0871-2021 · Dec 16, 2020 · Open, ClassifiedREF 27219-1 X1, Medtronic, Delta Valve, Neonatal, Performance Level 1.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Contr
- Z-0398-2014 · Nov 18, 2013 · TerminatedMedtronic Delta Shunt Assembly Kit with Bioglide Small, Performance Level 1.0, Catalog No. 276201. Delta Shunt Assemblies are designed for use in shunting CSF from the lateral ve
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.