Delta® 46832Recalls & reported problems
Medtronic Ps Medical, Inc. · SHUNT 46832 NEONATAL DELTA SNAP L 1 · Class 2 · Implant
All models of Delta (31)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000029524), received between Apr 8, 2026 and Apr 8, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “Disconnection” (1 report).
5 other recalls cover devices cleared under the same FDA submission (K902783); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Disconnection1
Patient problems
- Unspecified Infection1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 8, 2026Disconnection
MEDTRONIC RECEIVED INFORMATION REGARDING A SHUNT. IT WAS REPORTED THE SHUNT BECAME DISCONNECTED FROM THE RESERVOIR TO THE PERITONEAL CATHETER. THE DOCTOR WENT IN ON (B)(6) 2026 AND RECONNECTED IT. ON (B)(6) 2026, THE DOCTORS FOUND THAT THE SAME SHUNT WAS AGAIN DISCONNECTED AND INFECTED.
Patient problems: Unspecified Infection
MAUDE report 24820955 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K902783. One submission often covers many models, so check the product and lot details in each record.
- Z-1115-2021 · Jan 11, 2021 · Open, ClassifiedMedtronic Delta CSF (Cerebro Spinal Fluid) -Flow Control Valves for the following CFNs/REF numbers: 24048 - VALVE 24048 DELTA RGL PL 0.5, 24053 - CHAMBER 24053 DELTA, 24147 -
- Z-0872-2021 · Dec 16, 2020 · Open, ClassifiedREF 27219-2 X1, Medtronic, Delta Valve, Neonatal, Performance Level 2.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Contr
- Z-0873-2021 · Dec 16, 2020 · Open, ClassifiedREF 27219-5 X1, Medtronic, Delta Valve, Neonatal, Performance Level 1.5, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Contr
- Z-0871-2021 · Dec 16, 2020 · Open, ClassifiedREF 27219-1 X1, Medtronic, Delta Valve, Neonatal, Performance Level 1.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Contr
- Z-0398-2014 · Nov 18, 2013 · TerminatedMedtronic Delta Shunt Assembly Kit with Bioglide Small, Performance Level 1.0, Catalog No. 276201. Delta Shunt Assemblies are designed for use in shunting CSF from the lateral ve
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page