Delta® 27219-1 (UDI 00763000030964)Recalls & reported problems

Medtronic Ps Medical, Inc. · VALVE 27219-1 NEONATE DELTA · Class 2 · Implant

Quick summary

This Medtronic Delta shunt valve is used to drain excess fluid from the brain in newborns. The 24 reports mostly describe valves arriving with mismatched or shortened connectors that were noticed before use, with no deaths and no injuries mentioned. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00763000030964), received between Sep 14, 2020 and Aug 27, 2024 — classified as 24 malfunctions.

The most frequently reported specific device problem is “Device Misassembled During Manufacturing /Shipping” (20 reports).

5 other recalls cover devices cleared under the same FDA submission (K902783); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Misassembled During Manufacturing /Shipping20
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Infusion or Flow Problem1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions24
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 27, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION REGARDING A VALVE. IT WAS REPORTED THAT THERE WAS AN OUT OF BOX FAILURE. NO PATIENT INJURY OCCURRED, AND THE ONLY DELAY WAS OPENING ANOTHER VALVE FROM A DIFFERENT COMPANY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20076679 ↗
  • MalfunctionSep 3, 2021Infusion or Flow Problem; Infusion or Flow Problem

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THERE WERE NO ENVIRONMENTAL/EXTERNAL/PATIENT FACTORS THAT MAY HAVE LED OR CONTRIBUTED TO THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12422238 ↗
  • MalfunctionJan 28, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11243816 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856930 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856873 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856830 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856820 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856753 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856742 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856696 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856692 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856653 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856623 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856545 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856539 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856525 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856506 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856486 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856352 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856345 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856309 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10856307 ↗
  • MalfunctionNov 17, 2020Device Misassembled During Manufacturing /Shipping; Device Misassembled During Manufacturing /Shipping

    IT WAS REPORTED THAT IT SEEMED THAT THE VENTRICULAR CONNECTION SITE WAS MUCH SHORTER THAN IT WAS SUPPOSED TO BE. THE FACILITY WAS CURRENTLY CEASING ALL USE OF THE VALVES. ADDITIONAL INFORMATION RECEIVED REPORTED THAT JUST BY LOOKING AT THE VALVE IN THE STERILE PACKAGING, THERE WAS A DISTINCT DIFFERENCE BETWEEN THE TWO CONNECTORS. THE DISTAL CONNECTOR WAS MUCH LONGER THAN THE PROXIMAL CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10855611 ↗
  • MalfunctionSep 14, 2020Break; Break

    IT WAS REPORTED THAT THE VALVE HAD A DAMAGED DISTAL CATHETER CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10530957 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K902783. One submission often covers many models, so check the product and lot details in each record.

  • Z-1115-2021 · Jan 11, 2021 · Open, Classified
    Medtronic Delta CSF (Cerebro Spinal Fluid) -Flow Control Valves for the following CFNs/REF numbers: 24048 - VALVE 24048 DELTA RGL PL 0.5, 24053 - CHAMBER 24053 DELTA, 24147 -
  • Z-0872-2021 · Dec 16, 2020 · Open, Classified
    REF 27219-2 X1, Medtronic, Delta Valve, Neonatal, Performance Level 2.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Contr
  • Z-0873-2021 · Dec 16, 2020 · Open, Classified
    REF 27219-5 X1, Medtronic, Delta Valve, Neonatal, Performance Level 1.5, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Contr
  • Z-0871-2021 · Dec 16, 2020 · Open, Classified
    REF 27219-1 X1, Medtronic, Delta Valve, Neonatal, Performance Level 1.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Contr
  • Z-0398-2014 · Nov 18, 2013 · Terminated
    Medtronic Delta Shunt Assembly Kit with Bioglide Small, Performance Level 1.0, Catalog No. 276201. Delta Shunt Assemblies are designed for use in shunting CSF from the lateral ve

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.