Medtronic Xomed, Inc. · PROBE 8225275 BALL TIP 1MM ROHS · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000036096), received between Jun 23, 2022 and Dec 18, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Device Sensing Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Sensing Problem1
- Failure to Sense1
- False Negative Result1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 18, 2025False Negative Result; False Negative Result
IT WAS REPORTED THAT INTRA OP, AFTER CONNECTING THE EQUIPMENT, IT WAS DISCOVERED THAT THE PROBE WAS NOT RELEASING CURRENT AND HAD NO EFFECT WHEN PLACED ON THE NERVE. REPLACING THE PROBE WITH A NEW ONE RESOLVED THE ISSUE. THERE IS NO PATIENT IMPACT. THERE IS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23836686 ↗MalfunctionNov 9, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING THE THYROIDECTOMY PROCEDURE, WHEN THE DOCTOR USED THE PROBE TO DETECT THE NERVE, HE FOUND THAT THE PROBE WOULD REACTION DELAY. THE DOCTOR TRIED TO RECONNECT THE PROBE BUT USELESS. THE STIM RETURNED SHOWS 3MA. THERE WERE NO VISUAL AND/OR AUDIBLE INDICATORS THAT ALERTED THE USER OF THE ISSUE. NO ERROR CODES ARE DISPLAYED. THE ISSUE WAS NOT RESOLVED BY REPOSITIONING OR TROUBLESHOOTING. FINALLY, THE DOCTOR CHANGED A NEW PROBE TO COMPLETE THE SURGERY. THERE WAS NO INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18108245 ↗MalfunctionJul 14, 2023Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED DURING THYROIDECTOMY PROCEDURE, THAT WHEN THE DOCTOR USED THE PROBE FOR ABOUT 30 MINUTES, DOCTOR FOUND THAT THE NIM WOULD DELAY THE RESPONSE WHEN HE DETECT THE NERVE. AFTER USED THE PROBE FOR ABOUT 1.5 HOURS, THE DOCTOR FOUND THAT THE NIM HAD NO RESPONSE WHEN HE DETECT THE NERVE. SO, THE DOCTOR CHANGED A NEW PROBE AND THE NEW PROBE COULD FUNCTION NORMALLY. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT. NO PATIENT IMPACT. NO PROCEDURE DELAY. ADDITIONAL INFORMATION RECEIVED, THE STIM RETURN WAS NOT SHOWING 0. THE ISSUE DISCOVERED AFTER USED ON A PATIENT. NO VISUAL OR AUDIBLE ALERTED THE USER OF THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17323568 ↗MalfunctionJun 23, 2022Failure to Sense; Failure to Sense
IT WAS REPORTED THAT DURING PROCEDURE AFTER BEING CONNECTED, THE PROBE DID NOT REACT. THE PROBE WAS SWITCHED OUT WITH BACKUP PRODUCT AND THE PROCEDURE WAS CARRIED OUT NORMALLY. DEVICE HAD CONTACT WITH PATIENT. THERE WAS NO SURGICAL DELAY. THERE WAS NO PATIENT IMPACT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14792196 ↗
Other STIMULATOR, NERVE devices
- NIM Vital NIM4CM01Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls260 reports
- NIM TriVantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls176 reports
- NIM® 8229306Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall1 death167 reports
- NIM Vital™ NIM4CPB1Medtronic Xomed, Inc. · STIMULATOR, NERVE3 recalls45 reports
- NIM Trivantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall142 reports
- NIM® 8229307Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall138 reports
- NIM TriVantage® 8229706Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls78 reports
- NIM-Response® 8253002Medtronic Xomed, Inc. · STIMULATOR, NERVE171 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page