N/A 8225275Recalls & reported problems

Medtronic Xomed, Inc. · PROBE 8225275 BALL TIP 1MM ROHS · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000036096), received between Jun 23, 2022 and Dec 18, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Sensing Problem1
  • Failure to Sense1
  • False Negative Result1
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 18, 2025False Negative Result; False Negative Result

    IT WAS REPORTED THAT INTRA OP, AFTER CONNECTING THE EQUIPMENT, IT WAS DISCOVERED THAT THE PROBE WAS NOT RELEASING CURRENT AND HAD NO EFFECT WHEN PLACED ON THE NERVE. REPLACING THE PROBE WITH A NEW ONE RESOLVED THE ISSUE. THERE IS NO PATIENT IMPACT. THERE IS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23836686 ↗
  • MalfunctionNov 9, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING THE THYROIDECTOMY PROCEDURE, WHEN THE DOCTOR USED THE PROBE TO DETECT THE NERVE, HE FOUND THAT THE PROBE WOULD REACTION DELAY. THE DOCTOR TRIED TO RECONNECT THE PROBE BUT USELESS. THE STIM RETURNED SHOWS 3MA. THERE WERE NO VISUAL AND/OR AUDIBLE INDICATORS THAT ALERTED THE USER OF THE ISSUE. NO ERROR CODES ARE DISPLAYED. THE ISSUE WAS NOT RESOLVED BY REPOSITIONING OR TROUBLESHOOTING. FINALLY, THE DOCTOR CHANGED A NEW PROBE TO COMPLETE THE SURGERY. THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18108245 ↗
  • MalfunctionJul 14, 2023Device Sensing Problem; Device Sensing Problem

    IT WAS REPORTED DURING THYROIDECTOMY PROCEDURE, THAT WHEN THE DOCTOR USED THE PROBE FOR ABOUT 30 MINUTES, DOCTOR FOUND THAT THE NIM WOULD DELAY THE RESPONSE WHEN HE DETECT THE NERVE. AFTER USED THE PROBE FOR ABOUT 1.5 HOURS, THE DOCTOR FOUND THAT THE NIM HAD NO RESPONSE WHEN HE DETECT THE NERVE. SO, THE DOCTOR CHANGED A NEW PROBE AND THE NEW PROBE COULD FUNCTION NORMALLY. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT. NO PATIENT IMPACT. NO PROCEDURE DELAY. ADDITIONAL INFORMATION RECEIVED, THE STIM RETURN WAS NOT SHOWING 0. THE ISSUE DISCOVERED AFTER USED ON A PATIENT. NO VISUAL OR AUDIBLE ALERTED THE USER OF THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17323568 ↗
  • MalfunctionJun 23, 2022Failure to Sense; Failure to Sense

    IT WAS REPORTED THAT DURING PROCEDURE AFTER BEING CONNECTED, THE PROBE DID NOT REACT. THE PROBE WAS SWITCHED OUT WITH BACKUP PRODUCT AND THE PROCEDURE WAS CARRIED OUT NORMALLY. DEVICE HAD CONTACT WITH PATIENT. THERE WAS NO SURGICAL DELAY. THERE WAS NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14792196 ↗

Other STIMULATOR, NERVE devices

All STIMULATOR, NERVE devices →