Silver Bullet® 1882905HRERecalls & reported problems
Medtronic Xomed, Inc. · BLADE 1882905HRE SILVER BULLET M4 2.9MM · Class 1
All models of Silver Bullet (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000036782), received between Mar 31, 2020 and Jun 1, 2021 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Material Fragmentation” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Aug 31, 2021Class IIOpen, Classified
There was an increase in complaints related to tip breaks and wobble/vibration.
Z-0156-2022 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Fragmentation2
- Break1
Patient problems
- Unintended Radiation Exposure2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJun 1, 2021Material Fragmentation; Material Fragmentation
IT WAS REPORTED BY A HEALTH CARE PROFESSIONAL THAT A SILVER BULLET BLADE WAS USED TO TRIM THE SINUS TURBINATE AND THE PIECE INSIDE THE BLADE BROKE OFF. THEY REPLACED WITH ANOTHER BLADE WITH THE SAME LOT NUMBER AND IT BROKE AS WELL. AN X RAY WAS DONE AS PROTOCOL AT THE END OF THE CASE TO MAKE SURE NOTHING WAS IN THE NOSE; IT WAS CONFIRMED THAT NOTHING REMAINED IN THE NOSE. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT(S) WITH 30MINS DELAY. THERE WAS NO PATIENT IMPACT. UPON FOLLOW UP IT WAS REPORTED THAT X-RAY WAS DONE SPECIFICALLY TO CHECK FOR FRAGMENTS. THE PART OF THE BLADE THAT BROKE WAS DISTAL TIP. THERE WERE FRAGMENTS DETACHED FROM THE BROKEN BLADE THAT FELL INTO THE PATIENT. THE ANESTHESIA HAD TO DELAY WAKING THE PATIENT UP SO AN X-RAY COULD BE PERFORMED TO MAKE SURE NO FRAGMENTS REMAINED IN THE SINUSES.
Patient problems: Unintended Radiation Exposure
MAUDE report 11913137 ↗InjuryJun 1, 2021Material Fragmentation; Material Fragmentation
IT WAS REPORTED BY A HEALTH CARE PROFESSIONAL THAT A SILVER BULLET BLADE WAS USED TO TRIM THE SINUS TURBINATE AND THE PIECE INSIDE THE BLADE BROKE OFF. THEY REPLACED WITH ANOTHER BLADE WITH THE SAME LOT NUMBER AND IT BROKE AS WELL. AN X RAY WAS DONE AS PROTOCOL AT THE END OF THE CASE TO MAKE SURE NOTHING WAS IN THE NOSE; IT WAS CONFIRMED THAT NOTHING REMAINED IN THE NOSE. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT(S) WITH 30 MINS DELAY. THERE WAS NO PATIENT IMPACT. UPON FOLLOW UP IT WAS REPORTED THAT X-RAY WAS DONE SPECIFICALLY TO CHECK FOR FRAGMENTS. THE PART OF THE BLADE THAT BROKE WAS DISTAL TIP. THERE WERE FRAGMENTS DETACHED FROM THE BROKEN BLADE THAT FELL INTO THE PATIENT. THE ANESTHESIA HAD TO DELAY WAKING THE PATIENT UP SO AN X-RAY COULD BE PERFORMED TO MAKE SURE NO FRAGMENTS REMAINED IN THE SINUSES.
Patient problems: Unintended Radiation Exposure
MAUDE report 11913124 ↗MalfunctionMar 31, 2020Break; Break
A HEALTHCARE PROVIDER (HCP) REPORTED THAT THE BLADE BROKE INTRA OPERATIVELY. CUSTOMER USED BACK-UP PRODUCT TO COMPLETE THE PROCEDURE WITH NO ISSUES. NO PATIENT INJURY AND NO DELAY IN PROCEDURE. NOTHING BROKE OFF IN PATIENT OR WAS LEFT INSIDE PATIENT. ADDITIONAL INFORMATION RECEIVED FROM REP STATING THAT THE BLADE DID NOT HAVE FRAGMENTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9902079 ↗
Other BUR, EAR, NOSE AND THROAT devices
- Quadcut® 1884380HRMedtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall7 reports
- Diego Elite MBUR6060FRCVGyrus ACMI, LLC · BUR, EAR, NOSE AND THROAT1 recall2 reports
- Quadcut® 1884380HREMedtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall1 report
- RAD® 1882916Medtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall1 report
- N/A 1882936EMedtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall0 reports
- N/A 1884033Medtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall0 reports
- Tricut® 1884033HREMedtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall0 reports
- N/A 1883502HREMedtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page