Fusion® RAD® 1884006EMRecalls & reported problems

Medtronic Xomed, Inc. · BLADE 1884006EM RAD40 4MM M4 ROTATE ROHS · Class 2

All models of Fusion RAD (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00763000043254), received between Oct 19, 2020 and Nov 8, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 31, 2021Open, Classified

    There was an increase in complaints related to tip breaks and wobble/vibration.

    Z-0157-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break4
  • Material Fragmentation1
  • Material Separation1
  • Vibration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 8, 2024Break; Break

    IT WAS REPORTED THAT THREE SHAVERS NEEDED TO BE OPEN BECAUSE THE DISTAL TIPS BROKE DURING THE CASE. NO FRAGMENTS DETACHED FROM THE BROKEN DEVICE. THE BROKEN PARTS WERE CONTAINED BY THE HAND PIECE. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20638781 ↗
  • MalfunctionNov 8, 2024Break; Break

    IT WAS REPORTED THAT THREE SHAVERS NEEDED TO BE OPEN BECAUSE THE DISTAL TIPS BROKE DURING THE CASE. NO FRAGMENTS DETACHED FROM THE BROKEN DEVICE. THE BROKEN PARTS WERE CONTAINED BY THE HAND PIECE. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20638775 ↗
  • MalfunctionDec 20, 2023Break; Break

    A HEALTHCARE PROFESSIONAL (HCP) REPORTED THAT DURING THE FUNCTIONAL ENDOSCOPIC SINUS SURGERY THREE QUADCUT BLADES HAD CHATTERING (WOBBLING) ISSUES, REMOVED BLADES, REPLACED, TRIED AGAIN BUT STILL IT WAS CHATTERING, CHANGED HADPIECE BUT STILL IT WAS CHATTERING. USER CHANGED TO A NEW BLADE RAD40 BLADE IT WAS NOT CHATTERING, BUT IT BROKE AT THE DISTAL TIP WHILE BEING USED ON THE PATIENT. THE TIP SPRUNG OUT AS SOON AS THE USER PUT IT IN THE PATIENT AND INITIATED THE BLADE TO BEGIN SHAVING. USER USED BALDE FROM DIFFIRENT LOT IT WAS NOT CHATTERING. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18361638 ↗
  • MalfunctionAug 5, 2021Material Fragmentation; Material Fragmentation

    A HEALTHCARE PROFESSIONAL (HCP) REPORTED THAT THE BLADE TIP BROKE INTO MULTIPLE PIECES DURING THE FESS (FUNCTIONAL ENDOSCOPIC SINUS SURGERY) PROCEDURE. THE TIP OF THE BLADE DID NOT BREAK OFF IN THE PATIENT. THERE WAS NO UNPLANNED PROCEDURES OR UNINTENDED EQUIPMENT REQUIRED TO LOCATE AND REMOVE THE FRAGMENTS. THERE WAS NO PIECES REMAIN IN THE PATIENT. THEY BELIEVE THAT THE IMPACT ON THE PLEDGET CAUSE THE TEETH OF THE BLADE TO JAM LEADING TO THE BREAK. THE BLADE WAS RUNNING AT THE SPEED OF 5000. THERE WAS A DELAY WITH THE PROCEDURE. THERE WAS A BACK-UP DEVICE USED. THEY WERE ABLE TO PULL OUT THE PIECES FROM THE PATIENT PRIOR TO CLOSING. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12281381 ↗
  • MalfunctionOct 20, 2020Material Separation; Material Separation

    A HEALTHCARE PROVIDER (HCP) REPORTED THAT DURING OPERATION WHEN THE DOCTOR INSERTED BLADE INTO THE PATIENT AND PRESSED THE FOOT PEDAL, THE BLADE BLEW UP IN HER HAND. PART OF THE SHAVER, THE TIP, SEPARATED FROM THE REST OF THE SHAVER BLADE. THERE WERE NO FRAGMENTS DETACHED. BOX, PACKAGE, NOR PRODUCT WAS NOT TORN BEFORE OPENING AND USING. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10708828 ↗
  • MalfunctionOct 19, 2020Break; Vibration; Break; Vibration

    A HEALTHCARE PROVIDER (HCP) REPORTED THAT DURING A FESS PROCEDURE THE BLADE BROKE. IT SEEMED TO BE SHAKING TOWARD THE END OF USE DURING THE CASE AND HAD NOT BEEN IN USE FOR VERY LONG. WHEN THEY WENT TO SWAP TO A DIFFERENT BLADE THEY FOUND THE TIP HAD BEEN BROKEN. NOTHING BROKE OFF IN THE PATIENT AND NO PATIENT HARM OR DELAY TO THE CLINICAL CASE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10701778 ↗

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →