Skip to content

N/A 960-523Recalls & reported problems

Medtronic Navigation, Inc. · BIT GUIDE 960-523, 3MM-UC DRILL GUIDE · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000055349), received between May 2, 2018 and Mar 10, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation2
  • Mechanical Jam1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Patient Involvement1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 10, 2026Mechanical Jam

    IT WAS REPORTED THAT THE GUIDE PIN WAS ACTIVELY BEING USED/IN ROTATION WHILE IT GOT STUCK. THE ISSUE WAS RESOLVED BY HAVING ANOTHER GUIDE PIN BROUGHT OUT TO COMPLETE THE PROCEDURE UNDER FLUOROSCOPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24560276 ↗
  • MalfunctionOct 31, 2025Material Fragmentation; Material Fragmentation

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT THE WELDED KNOB CAME OFF THE INNER TROCAR OF THE UNIVERSAL DRILL GUIDE (UDG) PRE-CASE DURING SETUP, BEFORE INCISION AND AFTER ANESTHESIA HAD BEEN ADMINISTERED. NO ISSUES WERE NOTED BY CENTRAL STERILE SERVICES DEPARTMENT (CSSD) DURING CLEANING/STERILIZATION. A DIFFERENT TROCAR WAS USED AND THE SURGERY PROCEEDED AS USUAL WITHOUT DELAY OR DISCONTINUATION OF NAVIGATION. THERE WAS NO DELAY, AND NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23438876 ↗
  • MalfunctionAug 15, 2024Material Fragmentation; Material Fragmentation

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE DRILL BIT FROM THE DRILL GUIDE BROKE. THEY WIPED THE DRILL BIT AT THE FRONT END AND THEN HAD TWO PARTS IN THEIR HAND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19990940 ↗
  • MalfunctionMay 2, 2018Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION THAT A SITE NURSE DISCOVERED A DAMAGED DRILL BIT WHILE WASHING AFTER A PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Patient Involvement

    MAUDE report 7481730 ↗

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →

Sources and updates