N/A 960-527 (UDI 00763000055363)Recalls & reported problems

Medtronic Navigation, Inc. · INST 960-527 CANCELLOUS BIT 3MM UC DRIL · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000055363), received between Feb 3, 2023 and Feb 13, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem1
  • Mechanical Problem1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2025Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A CERVICAL SPINE PROCEDURE. IT WAS REPORTED THAT THE SITE WAS PREPARING TO UTILIZE A DRILL BIT WITH THE UNIVERSAL DRILL GUIDE (UDG), WHEN IT WAS FOUND THAT THE TIP WAS BENT. THE SITE ATTEMPTED TO OPEN AN ALTERNATE KIT, HOWEVER, BOTH KITS HAD BENT TIPS. WHEN THE DRILL BIT WAS LOADED INTO THE UDG, THEY WENT TO SPIN IT AND IT WOBBLED. THE DRILL BIT COULD BE ROLLED ACROSS A TABLE, AND THERE WAS A BEND IN THE TIP THAT WENT INTO THE ACTUAL DRILL. WHAT THEY DETERMINED HAPPENED MOST LIKELY WAS POOR HANDLING IN THE STERILE PROCESSING DEPARTMENT (SPD) THAT CAUSED THE BENT TIPS. THE SURGERY WAS ABORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21374185 ↗
  • MalfunctionFeb 3, 2023Use of Device Problem; Material Integrity Problem; Use of Device Problem; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A PELVIC TRAUMA SURGERY. IT WAS REPORTED THAT THE SURGEON HAD TO USE MORE FORCE AND WEIRD ANGLED TRAJECTORIES CAUSING THE DRILL BIT TO BEND SLIGHTLY AND CAUSED A 3.0 AND 2.1 DRILL BIT TO BREAK. ONE BROKE FROM PULLING AND ONE BROKE FROM PUSHING. THIS CAUSED LESS THAN ONE HOUR OF DELAY AND HAD NO IMPACT ON PATIENT OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16296600 ↗

Other Neurological stereotaxic Instrument devices