Arctic Front Advance Pro™ AFAPRO28 (UDI 00763000088637)Recalls & reported problems
Medtronic, Inc. · CATH AFAPRO28 AFA PRO 28MM US PB · Class 3
All models of Arctic Front Advance Pro (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Arctic Front Advance Pro cryoablation balloon catheter is used to treat atrial fibrillation. The 32 reports mostly describe the balloon twisting, bending, or breaking, plus air getting into the balloon, often noticed when the catheter was tested after the procedure and found out of specification. Six injuries were reported, including nerve damage, fluid around the heart, low blood pressure, and cardiac tamponade; no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 32
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 00763000088637), received between May 13, 2019 and Aug 31, 2020 — classified as 6 injuries, 26 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (17 reports).
1 other recall cover devices cleared under the same FDA submission (P100010); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Twisted/Bent17
- Material Integrity Problem13
- Adverse Event Without Identified Device or Use Problem6
- Gas/Air Leak3
- Break2
- Protective Measures Problem1
Patient problems
- No Known Impact Or Consequence To Patient23
- Nerve Damage3
- No Consequences Or Impact To Patient3
- Low Blood Pressure/ Hypotension2
- Pericardial Effusion2
- Cardiac Tamponade1
- No Code Available1
Recent adverse event reports
The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 31, 2020Material Integrity Problem; Material Twisted/Bent; Material Integrity Problem; Material Twisted/Bent
AFTER A COMPLETED CASE, THE BALLOON CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10473329 ↗MalfunctionJul 22, 2020Material Integrity Problem; Material Twisted/Bent; Material Integrity Problem; Material Twisted/Bent
AFTER A COMPLETED CASE, THE BALLOON CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10308164 ↗MalfunctionJul 2, 2020Material Twisted/Bent; Material Twisted/Bent
AFTER A COMPLETED CASE, THE BALLOON CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10224677 ↗MalfunctionMay 7, 2020Material Twisted/Bent; Material Twisted/Bent
AFTER A SUCCESSFUL CRYO ABLATION PROCEDURE, THE BALLOON CATHETER AND SHEATH SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10036906 ↗MalfunctionMay 6, 2020Material Integrity Problem; Material Integrity Problem
AFTER A COMPLETED CASE, THE BALLOON CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10030537 ↗MalfunctionMay 6, 2020Material Integrity Problem; Material Twisted/Bent; Material Integrity Problem; Material Twisted/Bent
THE BALLOON CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10029476 ↗MalfunctionApr 28, 2020Material Integrity Problem; Protective Measures Problem; Material Integrity Problem; Protective Measures Problem
AFTER A COMPLETED CASE, THE BALLOON CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10006291 ↗MalfunctionApr 22, 2020Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, AIR INGRESS OCCURRED WHEN THE BALLOON CATHETER WAS INSERTED INTO THE SHEATH. THE CASE WAS COMPLETED WITH CRYO. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9988315 ↗MalfunctionApr 17, 2020Material Twisted/Bent; Material Twisted/Bent
AFTER A COMPLETED CASE, THE BALLOON CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9974304 ↗MalfunctionMar 25, 2020Material Integrity Problem; Material Integrity Problem
AFTER A COMPLETED CASE, THE BALLOON CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9880105 ↗MalfunctionMar 23, 2020Material Integrity Problem; Material Twisted/Bent; Material Integrity Problem; Material Twisted/Bent
AFTER A COMPLETED CASE, THE BALLOON CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9869761 ↗MalfunctionMar 13, 2020Material Integrity Problem; Material Twisted/Bent; Material Integrity Problem; Material Twisted/Bent
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, THE MAPPING CATHETER WAS UNABLE TO PASS THROUGH THE BALLOON CATHETER AND RESISTANCE WAS FELT. A KINK WAS OBSERVED IN THE BALLOON CATHETER. THE BALLOON CATHETER WAS REPLACED. THE CASE WAS COMPLETED WITH CRYO. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9830058 ↗MalfunctionMar 4, 2020Material Integrity Problem; Material Integrity Problem
AFTER A COMPLETED CASE, THE BALLOON CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9788796 ↗MalfunctionFeb 11, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, THE GUIDEWIRE LUMEN KINKED WHEN THE BALLOON CATHETER WAS INFLATED. THE BALLOON CATHETER WAS REPLACED. THE CASE WAS COMPLETED WITH CRYO. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9694516 ↗InjuryFeb 6, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, SEVERAL ATTEMPTS WERE MADE TO CROSS THE SEPTUM DURING THE TRANSSEPTAL PUNCTURE. THE CASE WAS CONTINUED WITH CRYO. ON THE LAST FREEZE, THE PATIENT'S BLOOD PRESSURE DROPPED AND AN EFFUSION WAS OBSERVED ON ICE WITH A TAMPONADE. A PERICARDIOCENTESIS WAS PERFORMED. THE PATIENT'S HOSPITALIZATION STAY WAS EXTENDED. THE CASE WAS COMPLETED WITH CRYO. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Cardiac Tamponade; Low Blood Pressure/ Hypotension; Pericardial Effusion
MAUDE report 9679699 ↗InjuryJan 14, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, THE PATIENT EXPERIENCE PHRENIC NERVE PALSY (PNP). AT THE END OF THE PROCEDURE, THE PNP HAD NOT RECOVERED. THE CASE WAS COMPLETED WITH CRYO. THE PATIENT WAS WITHOUT SYMPTOMS PRIOR TO DISCHARGE, HOWEVER IT WAS NOT CLEAR IF THE PNP HAD RECOVERED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Nerve Damage
MAUDE report 9586384 ↗MalfunctionDec 26, 2019Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, THE BALLOON CATHETER SHAFT APPEARED TO BE DISPLACED FROM ITS 'NORMAL' MID-LINE POSITION IN THE BALLOON WHILE THE BALLOON WAS INFLATED. THE BALLOON CATHETER WAS REPLACED WITH RESOLVE. THE CASE WAS COMPLETED WITH CRYO. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9520852 ↗InjuryDec 11, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, BLOOD WAS OBSERVED IN THE ENDOTRACHEAL TUBE. THE LUNGS WERE SCOPED AND NO ACTIVE BLEEDING WAS OBSERVED. THE BLOOD WAS SUCTIONED OUT. ON THE NEXT OCCLUSION, BLOOD WAS AGAIN OBSERVED IN THE ET TUBE. THE PROCEDURE WAS STOPPED AND ABORTED WHILE THE PATIENT WAS UNDER GENERAL ANESTHESIA. THE PATIENT WAS TRANSFERRED TO THE INTENSIVE CARE UNIT FOR HOSPITALIZATION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Code Available
MAUDE report 9453583 ↗MalfunctionNov 25, 2019Material Integrity Problem; Material Twisted/Bent; Material Integrity Problem; Material Twisted/Bent
AFTER A COMPLETED CASE, THE BALLOON CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9371364 ↗MalfunctionNov 20, 2019Gas/Air Leak; Material Integrity Problem; Gas/Air Leak; Material Integrity Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, A SYSTEM NOTICE WAS RECEIVED INDICATING THAT THE SAFETY SYSTEM DETECTED A COMPROMISED OUTER VACUUM. THE BALLOON CATHETER WAS REPLACED WITH RESOLVE. THE CASE WAS COMPLETED WITH CRYO. THEN, WHEN PERFORMING TEST ABLATIONS, AIR BUBBLES WERE NOTED COMING FROM THE SURFACE OF THE BALLOON OF THE SECOND BALLOON CATHETER. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9348938 ↗InjuryNov 14, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, THE PATIENT EXPERIENCED PHRENIC NERVE PALSY (PNP). THE CASE WAS COMPLETED WITH CRYO. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Nerve Damage
MAUDE report 9321579 ↗MalfunctionNov 13, 2019Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, A KINK WAS NOTED TO THE BALLOON OF THE BALLOON CATHETER. THE CASE WAS COMPLETED WITH RADIOFREQUENCY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9315659 ↗MalfunctionNov 7, 2019Material Twisted/Bent; Material Twisted/Bent
AFTER A COMPLETED CASE, THE BALLOON CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9293452 ↗MalfunctionNov 5, 2019Material Integrity Problem; Material Twisted/Bent; Material Integrity Problem; Material Twisted/Bent
AFTER A COMPLETED CASE, THE BALLOON CATHETER AND MAPPING CATHETER SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9279227 ↗InjuryOct 21, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, THE PATIENT HAD A DROP IN BLOOD PRESSURE DESPITE MEDICATION SUPPORT. AN ECHOCARDIOGRAM NOTED A PERICARDIAL EFFUSION AND A CARDIAC TAMPONADE WAS DIAGNOSED. A PERICARDIOCENTESIS WAS PERFORMED. THE CASE WAS COMPLETED WITH CRYO. THE PATIENT'S HOSPITAL STAY WAS EXTENDED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Low Blood Pressure/ Hypotension; Pericardial Effusion
MAUDE report 9215296 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100010. One submission often covers many models, so check the product and lot details in each record.
- Z-0149-2012 · Aug 29, 2011 · TerminatedMedtronic Arctic Front Catheter, Sterile EO, Medtronic CryoCath LP, Kirkland, Quebec, H9H 5 H3, Canada, Model 2AF282 and 2AF231 ( not distributed in US) Product Usage: The Medtr
Other Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation devices
- TactiCath™ A-TCSE-DFSt. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 recall23 deaths923 reports
- TactiCath™ PN-004075St. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation7 deaths338 reports
- TactiFlex™ Sensor Enabled™ A-TFSE-DFSt. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 recall6 deaths241 reports
- TactiCath™ A-TCSE-FJSt. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 recall3 deaths174 reports
- Sphere-9™ Catheter AFR-00001Medtronic, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 recall3 deaths147 reports
- TactiCath™ A-TCSE-DSt. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation2 deaths175 reports
- TactiSys™ PN-004400St. Jude Medical, Inc. · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation1 recall125 reports
- QDOT MICRO D139402Biosense Webster Inc · Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation3 deaths171 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page