Valiant Navion™ VNMF3131C97TU (UDI 00763000100599)Recalls & reported problems
Medtronic, Inc. · STENT GRAFT VNMF3131C97TU VAL NAV FF · Class 3 · Implant
All models of Valiant Navion (137)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 1
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00763000100599), received between May 1, 2019 and May 16, 2022 — classified as 1 death, 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Leak/Splash” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Leak/Splash3
- Adverse Event Without Identified Device or Use Problem1
- Malposition of Device1
Patient problems
- Insufficient Information2
- Aneurysm1
- Injury1
- Intimal Dissection1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMay 16, 2022Leak/Splash; Leak/Splash
A VALIANT NAVION STENT GRAFT WAS IMPLANTED DURING AN ENDOVASCULAR PROCEDURE FOR THE TREATMENT OF A TYPE IA ENDOLEAK. IT WAS REPORTED APPROXIMATELY 2 YEARS POST THE INDEX PROCEDURE THE PATIENT RETURNED FOR A FOLLOW-UP CT. A TYPE III ENDOLEAK AND AORTIC ENLARGEMENT OF 6.5->6.6 WAS NOTED. AS PER THE PHYSICIAN THE TYPE III ENDOLEAK MAY BE ON A NON-MDT DEVICE. NO ADDITIONAL CLINICAL SEQUELAE WAS PROVIDED AND THE PATIENT WILL BE MONITORED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 14406936 ↗MalfunctionJan 4, 2022Leak/Splash; Leak/Splash
A VALIANT NAVION WAS IMPLANTED DURING AN ENDOVASCULAR PROCEDURE FOR THE TREATMENT OF A THORACIC AORTIC ANEURYSM WITH A PENETRATING THORACIC ULCER. IT WAS REPORTED THE PATIENT RETURNED FOR A FOLLOW-UP CT APPROXIMATELY 19 MONTHS LATER WHERE IT WAS DISCOVERED THAT THERE WAS ANEURYSMAL DILATION/PENETRATING ULCERATION NEAR THE LEVEL OF THE DESCENDING AORTIC GRAFT WHICH HAS SIGNIFICANTLY INCREASED MEASURING UP TO 56MM AT THE PROXIMAL MARGIN OF THE GRAFT WITH ASSOCIATED ENDOLEAK AROUND THE GRAFT. THE AORTA MEASURES UP TO 51MM AT THE LEVEL OF DISTAL MARGIN OF THE GRAFT WITH ASSOCIATED ENDOLEAK AROUND THE DISTAL STENT. THE ANEURYSM MAY HAVE AN INFLAMMATORY COMPONENT GIVEN SOME EROSION ALONG THE ANTERIOR MARGIN OF THE T10 VERTEBRAL BODY. THE AORTA AT THE LEVEL OF THE HIATUS MEASURES UP TO 53MM , PREVIOUSLY 34MM. NO CAUSE OF EVENT AS PER THE PHYSICIAN HAS BEEN PROVIDED. NO ADDITIONAL CLINICAL SEQUALAE HAS BEEN PROVIDED AND THE PATIENT IS FINE.
Patient problems: Aneurysm
MAUDE report 13151047 ↗DeathNov 10, 2021Leak/Splash; Leak/Splash
VALIANT NAVION STENT GRAFTS WERE IMPLANTED DURING AN UNKNOWN ENDOVASCULAR PROCEDURE. IT WAS REPORTED THAT CORE LAB IDENTIFIED A TYPE II ENDOLEAK FROM AN IMAGE THAT WAS DATED 4 MONTHS LATER, CORE LAB REPORTED NO STENT RING ENLARGEMENT, NO FRACTURES AND NO OTHER ENDOLEAKS. DURING FOLLOWING UP FROM THE CORE LAB'S OBSERVATIONS OF A TYPE II ENDOLEAK, IT WAS REPORTED THE PATIENT PASSED AWAY JUST OVER A YEAR LATER WITH NO OTHER SPECIFIC DETAILS. THE CAUSE OF DEATH IS UNDETERMINED. NO ADDITIONAL CLINICAL SEQUELAE WAS REPORTED AND THE PATIENT HAS EXPIRED.
Patient problems: Insufficient Information
MAUDE report 12785352 ↗InjuryNov 12, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A VALIANT NAVION STENT GRAFT WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF PAU (PENETRATING ATHEROSCLEROTIC ULCER) AND IMH (INTRAMURAL HEMATOMA) OF THE THORACIC AORTA. IT WAS REPORTED THAT AFTER THE PROCEDURE WAS COMPETED, COMPLETION ANGIOGRAM SHOWED EXCLUSION OF THE PAU. IT WAS REPORTED THAT CTA CARRIED OUT 5 DAYS POST INDEX PROCEDURE, PRIOR TO DISCHARGE, SHOWED A NEW DISSECTION PROXIMAL TO THE THORACIC STENT GRAFT AND THE IMH APPEARED TO HAVE INCREASED WITH MORE ADDITIONAL INVOLVEMENT OF THE TRANSVERSE ARCH. AS PER THE PHYSICIAN, THE CAUSE OF THE EVENT CANNOT BE DETERMINED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS BEING MONITORED.
Patient problems: Intimal Dissection
MAUDE report 9309219 ↗InjuryMay 1, 2019Malposition of Device; Malposition of Device
A VALIANT NAVION STENT GRAFT WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A THORACIC AORTIC ULCER. IT WAS REPORTED THAT DURING THE INDEX PROCEDURE, THE DEVICE WAS NOT IMPLANTED AT THE DESIRED LANDING POSITION, DUE TO POSITIONING DIFFICULTIES. AS PER THE PHYSICIAN, THE CAUSE OF THE EVENT WAS RELATED TO THE CALCIFIED PATIENT ANATOMY. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Patient problems: Injury; No Consequences Or Impact To Patient
MAUDE report 8568940 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page