Valiant Navion™ VNMF4343C183TU (UDI 00763000100766)Recalls & reported problems
Medtronic, Inc. · STENT GRAFT VNMF4343C183TU VAL NAV FF · Class 3 · Implant
All models of Valiant Navion (137)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000100766), received between May 22, 2023 and Jul 17, 2023 — classified as 1 death, 1 injury.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Feb 12, 2021Class IOpen, Classified
Due to the presence of type IIIb endoleaks, stent fractures, and stent ring enlargement.
Z-1201-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- Aneurysm1
- Swelling/ Edema1
- Thrombosis/Thrombus1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
DeathJul 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A VALIANT NAVION STENT GRAFT WAS IMPLANTED DURING THE ENDOVASCULAR TREATMENT OF AN ULCER. IT WAS REPORTED THAT A FOLLOW-UP CT TAKEN APPROXIMATELY 5 MONTHS POST THE INDEX PROCEDURE SHOWED THE SACCULAR ANEURYSM OF THE DISTAL THORACIC AORTA MEASURING UP TO 40MM. IT WAS REPORTED THE PATIENT EXPIRED APPROXIMATELY 15 MONTHS POST THE INDEX PROCEDURE. THE CAUSE OF THE ANEURYSM ENLARGEMENT AND DEATH ARE UNDETERMINED. NO ADDITIONAL CLINICAL SEQUELAE WERE PROVIDED AND THE PATIENT IS EXPIRED.
Patient problems: Aneurysm
MAUDE report 17331315 ↗InjuryMay 22, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
VALIANT NAVION STENT GRAFTS WERE IMPLANTED DURING THE TREATMENT OF A TYPE IB ENDOLEAK IN AN EXISTING VALIANT NAVION STENT GRAFT ON (B)(6)2020. THE ANEURYSM SAC DIAMETER WAS NOTED TO MEASURE 95MM AND THE PATIENT WAS ASYMPTOMATIC. AN EXISTING VALIANT NAVION STENT GRAFT WAS IMPLANTED DURING THE ENDOVASCULAR TREATMENT OF A THORACIC AORTIC ANEURYSM ON (B)(6)2019. ENDOANCHORS WERE ALSO IMPLANTED ON AN UNKNOWN DATE. IT WAS REPORTED ON AN UNKNOW DATE IN 2021, A REVISION OF THE THORACOABDOMINAL ENDOGRAFT REPAIR WAS PERFORMED IN 2021 WITH REVISION OF THE ANEURYSM SAC, DRAINAGE OF FLUID AND REMOVAL OF THROMBUS. IT WAS REPORTED THAT THE PATIENT HAD A LEFT THORACOABDOMINAL WOUND POST THE 2021 ANEURYSM SAC REVISION. NO CAUSE WAS PROVIDED. NO ADDITIONAL SEQUALAE WAS REPORTED AND THE PATIENT WILL BE MONITORED.
Patient problems: Swelling/ Edema; Thrombosis/Thrombus
MAUDE report 16972603 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record