Skip to content

Valiant Navion™ VNMF4034C185TU (UDI 00763000100872)Recalls & reported problems

Medtronic, Inc. · STENT GRAFT VNMF4034C185TU VAL NAV FF TP · Class 3 · Implant

All models of Valiant Navion (137)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000100872), received between May 23, 2022 and Apr 19, 2024 — classified as 1 death, 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash4

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Hematoma1
  • Insufficient Information1
  • Obstruction/Occlusion1
  • Pneumonia1
  • Vascular Dissection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 19, 2024Leak/Splash; Leak/Splash

    A VALIANT NAVION STENT GRAFT (VNMF4034C185TU) WAS IMPLANTED DURING THE ENDOVASCULAR TREATMENT OF A THORACIC AORTIC DISSECTION. THE PATIENT HAD AN EXISTING VALIANT CAPTIVIA (VAMF4040C150TU) THORACIC STENT GRAFT IMPLANTED 6 YEARS PREVIOULSY. IT WAS REPORTED CORE LAB REVIEW OF CT IMAGING TAKEN APPROXIMATELY 4 YEARS POST THE INDEX PROCEDURE SHOWED A TYPE IA (NEW) ENDOLEAK. THE MAX AORTIC DIAMETER WAS NOTED AS 89.1MM. THE IMAGING WAS NA FOR AORTIC ENLARGEMENT. NO STENT FRACTURE, STENT RING ENLARGEMENT OR STENT RING MIGRATION WAS OBSERVED. A REINTERVENTION WAS PERFORMED. THE CAUSE OF THE TYPE IA ENDOLEAK IS UNDETERMINED.  NO ADDITIONAL CLINICAL SEQUALAE WERE PROVIDED AND THE PATIENT WILL BE MONITORED.

    Patient problems: Insufficient Information

    MAUDE report 19141811 ↗
  • DeathJul 17, 2023Leak/Splash; Leak/Splash

    VALIANT NAVION STENT GRAFTS WERE IMPLANTED IN THE ENDOVASCULAR TREATMENT OF A 70MM TAA.  IT WAS REPORTED ONE DAY POST THE INDEX PROCEDURE, CT SHOWED THE FOLLOWING OBSERVATIONS: INTERNAL DEVELOPMENT OF A SHORT SEGMENT TYPE B DISSECTION / TYPE I ENDOLEAK ALONG PROXIMAL ORIGIN OF STENT. A 2.4X2.9 CM HEMATOMA MEDIAL TO THE LEFT DISTAL EXTERNAL ILIAC ARTERY AND A OCCLUSION OF THE ANTERIOR DIVISION OF THE LEFT MAIN RENAL ARTERY. IT IS UNKNOWN WHICH STENT GRAFT THE ENDOLEAK OCCURRED.  THE PATIENT SUBSEQUENTLY EXPIRED APPROXIMATELY 1 MONTH POST THE INDEX PROCEDURE DUE TO PNEUMONIA. IT IS UNKNOWN WHEN THE PATIENT DEVELOPED THE PNEUMONIA. PER THE PHYSICIAN THE CAUSE OF THE EVENTS SEEN ON THE CT COULD NOT BE DETERMINED.  NO ADDITIONAL CLINICAL SEQUALAE WERE PROVIDED AND THE PATIENT EXPIRED.

    Patient problems: Hematoma; Obstruction/Occlusion; Pneumonia; Vascular Dissection

    MAUDE report 17331324 ↗
  • MalfunctionJun 30, 2023Leak/Splash; Leak/Splash

    VALIANT NAVION STENT GRAFTS WERE IMPLANTED IN THE ENDOVASCULAR TREATMENT OF A TAA AND DISSECTION .  IT WAS REPORTED CT IMAGING WAS PERFORMED ABOUT THREE AND HALF YEARS LATER . CORELAB HAVE REVIEWED THIS IMAGING AND IDENTIFIED AN UND ETERMINED TYPE ENDOLEAK.  A FRACTURE WAS NOTED ON THE STENT GRAFT WITH SN (B)(6). STENT RING ENLARGEMENT WAS ALSO NOTED ON STENT GRAFT WITH (B)(6) , STENT RING 4 (+ 4.2) , STENT RING 5 (+7.3), STENT RING 6 (+7.1). NO STENT RING MIGRATION WAS NOTED. THE MAXIMUM AORTIC DIAMETER MEASURED 71.9MM.  NO CAUSE WAS REPORTED.  NO ADDITIONAL CLINICAL SEQUALAE WERE PROVIDED AND THE PATIENT WILL BE MONITORED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17237617 ↗
  • MalfunctionMay 23, 2022Leak/Splash; Leak/Splash

    VALIANT NAVION STENT GRAFTS WERE IMPLANTED DURING AN UNKNOWN ENDOVASCULAR PROCEDURE. IT WAS REPORTED THE PATIENT RETURNED FOR A FOLLOW-UP CT. AN UNKNOWN ENDOLEAK WAS IDENTIFIED. NO CAUSE AS PER PHYSICIAN WAS PROVIDED. NO ADDITIONAL CLINICAL SEQUALAE WAS PROVIDED AND THE PATIENT WILL BE MONITORED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14469308 ↗

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page