Valiant Navion™ VNMC4646C218TU (UDI 00763000101190)Recalls & reported problems
Medtronic, Inc. · STENT GRAFT VNMC4646C218TU VAL NAV CS · Class 3 · Implant
All models of Valiant Navion (137)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00763000101190), received between Feb 22, 2019 and Apr 1, 2025 — classified as 1 death, 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash3
- Activation, Positioning or Separation Problem1
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
- Fracture1
- Material Deformation1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Aneurysm expansion1
- Cardiac Arrest1
- Death1
- Insufficient Information1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- Vascular Dissection1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryApr 1, 2025Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component
VALIANT NAVION STENT GRAFTS WERE IMPLANTED DURING THE ENDOVASCULAR TREATMENT OF A THORACIC AORTIC ANEURYSM. IT WAS REPORTED CORE LAB REVIEW OF CT IMAGING FROM APPROX. 48 MONTHS POST INDEX PROCEDURE, NOTED A NEW STENT RING ENLARGEMENT OF + 1.5MM WAS NOTED ON RING 4 OF (B)(6) AND +1.6MM ON RING 6 OF(B)(6). A NEW TYPE IIIB ENDOLEAK OBSERVED ON STENT GRAFT (B)(6). NO FRACTURE OR STENT RING MIGRATION WERE NOTED. THE MAXIMUM AORTIC DIAMETER MEASURED 58.5MM. NO AORTIC ENLARGEMENT WAS REPORTED. NO STENT RING MIGRATION OR STENT FRACTURES WERE OBSERVED. NO ADDITIONAL CLINICAL SEQUALAE WERE PROVIDED AND THE PATIENT WILL BE MONITORED.
Patient problems: Aneurysm expansion; No Clinical Signs, Symptoms or Conditions; Vascular Dissection
MAUDE report 21738599 ↗MalfunctionMar 8, 2022Leak/Splash; Leak/Splash
A VALIANT NAVION STENT GRAFT WAS IMPLANTED DURING AN ENDOVASCULAR PROCEDURE FOR THE TREATMENT OF A RUPTURED THORACIC AORTIC ANEURYSM . IT WAS REPORTED A CT SCAN REVEALED A POSSIBLE TYPE I ENDOLEAK. AS PER THE PHYSICIAN THE CAUSE CANNOT BE DETERMINED. NO ADDITIONAL CLINICAL SEQUALAE WAS PROVIDED AND THE PATIENT WILL BE MONITORED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13688639 ↗InjuryApr 23, 2021Fracture; Leak/Splash; Material Deformation; Fracture; Leak/Splash; Material Deformation
VALIANT CAPTIVIA AND NAVION STENT GRAFTS WERE IMPLANTED IN THE PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 67.8MM X 78.5MM THORACIC ANEURYSM. APPROX. 1 YEAR 8 MONTHS POST INDEX, AN ADDITIONAL 2 VALIANT NAVION DEVICES WERE IMPLANTED FOR AN UNKNOWN PROCEDURE IT WAS REPORTED THE CORE LAB OBSERVED THE FOLLOWING AFTER ANALYSIS OF THE 12M, INTERIM <(>&<)> 24M CTS: TYPE IIIB ENDOLEAK OF V07404460 VALIANT CAPTIVIA DEVICE <(>&<)> TYPE IIIA ENDOLEAK BETWEEN V08146953 AND V08097833 NAVION DEVICES. MOREOVER, 7 STENT FRACTURES (6 FRACTURES IN MOST DISTAL STENT AND 1 FRACTURE IN SECOND MOST DISTAL STENT) <(>&<)> STENT ENLARGEMENT IN SECOND STENT RING OF V08097833 NAVION DEVICE. NO CAUSE OF THE EVENT WAS PROVIDED. NO ADDITIONAL CLINICAL SEQUALAE WERE REPORTED AND THE PATIENT WILL BE MONITORED.
Patient problems: Insufficient Information
MAUDE report 11714961 ↗MalfunctionSep 11, 2019Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
A VALIANT NAVION STENT GRAFT WAS PLANNED TO BE IMPLANTED IN A PATIENT FOR A THORACIC AORTIC ENDOVASCULAR REPAIR. IT WAS REPORTED THAT DURING THE INDEX PROCEDURE THE DEVICE WAS ADVANCED INTO THE PROXIMAL DESCENDING THORACIC AORTA SUCCESSFULLY. HOWEVER, WITH RELEASE OF THE PROXIMAL CONSTRAINING MECHANISM OF THE COVERED SEAL PORTION OF THE DEVICE THE DEVICE WAS NOT SUCCESSFULLY RELEASED. THEREFORE, THE BACK PORTION OF THE DELIVERY SYSTEM WAS MANUALLY EXPOSED AND WITH MULTIPLE MOVEMENTS FORWARD AND BACKWARD OF THE INNER CANNULA OF THE DELIVERY SYSTEM ULTIMATELY THE PROXIMAL COVERED SEAL PORTION OF THE DEVICE WAS ABLE TO BE SUCCESSFULLY DEPLOYED. THE ENTIRE DELIVERY SYSTEM WAS THEN SUCCESSFULLY REMOVED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8988811 ↗MalfunctionAug 27, 2019Mechanical Problem
DURING TEVAR PROCEDURE MEDTRONIC VALIANT NAVION DEVICE MALFUNCTIONED DURING DEPLOYMENT. PORTION OF HANDLE DISASSEMBLED, DEPLOYMENT WAS SUCCESSFUL. FROM THE OPERATIVE NOTE: NEXT, THE PROXIMAL DEVICE A MEDTRONIC NEVI ON A 46 X 46 X 218 MM, WAS ADVANCED FROM THE LEFT FEMORAL ACCESS OVER THE DOUBLE CURVE LUNDERQUIST WIRE INTO THE PROXIMAL DESCENDING THORACIC AORTA AND ON SHEATH SUCCESSFULLY. HOWEVER, WITH RELEASE OF THE PROXIMAL CONSTRAINING MECHANISM OF THE COVERED SEAL PORTION OF THE DEVICE THE DEVICE WAS NOT SUCCESSFULLY RELEASED. THEREFORE, THE BACK PORTION OF THE DELIVERY SYSTEM WAS MANUALLY EXPOSED AND WITH MULTIPLE MOVEMENTS FORWARD AND BACKWARD OF THE INNER CANNULA OF THE DELIVERY SYSTEM ULTIMATELY THE PROXIMAL COVERED SEAL PORTION OF THE DEVICE WAS ABLE TO BE SUCCESSFULLY DEPLOYED. THE ENTIRE DELIVERY SYSTEM WAS THEN SUCCESSFULLY REMOVED OVER THE LUNDERQUIST WIRE AND A 20 FRENCH SHEATH WAS ADVANCED OVER THE LUNDERQUIST WIRE INTO THE LEFT COMMON FEMORAL ARTERY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8935021 ↗DeathFeb 22, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A VALIANT NAVION STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 370 MM THORACIC AORTIC DISSECTION. IT WAS REPORTED THAT TWO DAYS POST THE INDEX PROCEDURE THE PATIENT HAD A SUDDEN CARDIAC ARREST AND DIED. THE CAUSE OF THE CARDIAC ARREST IS UNKNOWN. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT HAS EXPIRED.
Patient problems: Cardiac Arrest; Death
MAUDE report 8361601 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page