Skip to content

Valiant Navion™ VNMC3125C207TU (UDI 00763000101220)Recalls & reported problems

Medtronic, Inc. · STENT GRAFT VNMC3125C207TU VAL NAV CS TP · Class 3 · Implant

All models of Valiant Navion (137)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000101220), received between Oct 17, 2022 and Jul 31, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 12, 2021Class IOpen, Classified

    Due to the presence of type IIIb endoleaks, stent fractures, and stent ring enlargement.

    Z-1201-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Leak/Splash1

Patient problems

  • Dyspnea1
  • Hypoxia1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    TWO VALIANT NAVION STENT GRAFTS WERE IMPLANTED DURING THE ENDOVASCULAR TREATMENT OF A DISSECTION ON (B)(6) 2020. IT WAS REPORTED VIA CORE LAB REVIEW OF FILMS APPROXIMATELY 4 YEARS AND 3 MONTHS POST IMPLANT THAT STENT RING ENLARGEMENT WAS IDENTIFIED IN (B)(6). NO CAUSE WAS PROVIDED. NO ADDITIONAL CLINICAL SEQUALAE WERE PROVIDED AND THE PATIENT WILL BE MONITORED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22677311 ↗
  • MalfunctionOct 17, 2022Leak/Splash; Leak/Splash

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

    Patient problems: Dyspnea; Hypoxia; Pain

    MAUDE report 15614620 ↗

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page