N/A 9733457 (UDI 00763000121198)Recalls & reported problems

Medtronic Navigation, Inc. · PROBE 9733457 PEDICLE 2.25mm · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00763000121198), received between Apr 9, 2019 and Mar 6, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 4 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem8
  • Material Fragmentation1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • No Patient Involvement1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 6, 2026Material Fragmentation

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT PROBE PEDICEL BROKE ON THE FIRST USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24533185 ↗
  • MalfunctionSep 18, 2025Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE INSTRUMENT WAS RETURNED AFTER THE LOANER PERIOD AND IT WAS FOUND TO BE DEFORMED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23085706 ↗
  • MalfunctionAug 25, 2025Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT BENDING OF THE TIP OF THE INSTRUMENT WAS CONFIRMED DURING PRE-USAGE PREPARATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22890711 ↗
  • MalfunctionAug 25, 2025Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED INTRA-OPERATIVELY DURING A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT THE TIP OF THE INSTRUMENT DEFORMED DURING USE IN A SCOLIOSIS OPERATION. IT WAS BELIEVED THAT THE INSTRUMENT WAS USED IN A HARD PART OF THE BONE. THE OPERATION WAS COMPLETED BY USING ANOTHER INSTRUMENT. NO KNOWN IMPACT TO PATIENT OUTCOME. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22890090 ↗
  • MalfunctionAug 6, 2025Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM. IT WAS REPORTED OUTSIDE OF A PROCEDURE THAT THE PROBE, WHICH WAS RETURNED DUE TO THE END OF THE LOAN PERIOD, WAS FOUND TO HAVE A DEFORMED TIP DURING THE ACCEPTANCE INSPECTION. NO PATIENT WAS PRESENT AT THE TIME OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22730181 ↗
  • MalfunctionAug 5, 2025Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM. IT WAS REPORTED OUTSIDE OF A PROCEDURE THAT THREE TRACKERS HAD POOR ATTACHMENT TO THE PASSIVE MARKERS. THE DRIVER HAD A SCREW INSTALLATION FAILURE. THE PROBE HAD A TIP DEFORMATION. NO PATIENT WAS PRESENT AT THE TIME OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22718481 ↗
  • MalfunctionJul 29, 2025Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT THERE WAS DEFORMATION OF THE TIP OF THE CERVICAL PROBE. IT WAS NOTED THAT THE PROBABLE CAUSE OF THE ISSUE WAS THAT THE DEVICE WAS USED IN A HARD PART OF THE BONE AND BENT. THE OPERATION WAS COMPLETED BY USING A DIFFERENT INSTRUMENT. THIS ISSUE CAUSED NO SURGICAL DELAY. THERE WAS NO REPORTED IMPACT ON PATIENT OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22648060 ↗
  • MalfunctionJul 17, 2025Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED FOR A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT THERE WAS A DEFORMATION OF CERVICAL PROBE TIP. IT IS UNKNOWN HOW THIS OCCURRED. THERE WAS NO REPORTED IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22534795 ↗
  • MalfunctionFeb 3, 2025Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE 2.25 MILLIMETER (MM) PEDICLE PROBE WAS BENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21291537 ↗
  • MalfunctionApr 9, 2019Material Integrity Problem; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE A PROCEDURE. IT WAS REPORTED THAT THE CERVICAL PROBE TIP WAS BENT. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.

    Patient problems: No Patient Involvement

    MAUDE report 8497001 ↗

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →