Skip to content

CD HORIZON® Fenestrated Screw Set 55851025550Recalls & reported problems

Medtronic Sofamor Danek, Inc. · FNS SAS 55851025550 CCM 5.5 X 50 · Class 2 · Implant

All models of CD HORIZON Fenestrated Screw Set (27)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000143763), received between Aug 10, 2022 and Aug 10, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 10, 2022Break; Break

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING PERCUTANEOUS TRAUMA STABILIZATION FOR T12 BURST FRACTURE. LEVELS IMPLANTED: T11-L1 IT WAS REPORTED THAT THE SCREWS SHANKS BROKE UNDER TULIP. 1-MONTH POST-OPERATIVE, THE SURGEON REPORTED THAT 1 SCREW HAD BROKEN, AT 3 MONTHS FOLLOW UP 2 SCREWS HAD BROKEN. REMOVAL OF SCREWS AND FUSION IS SCHEDULED TO HAPPEN IN THE NEXT 2 WEEKS DUE TO BROKEN SCREWS AND LOSS OF CORRECTION. PATIENT SYMPTOMS ARE UNKNOWN. NO FURTHER COMPLICATIONS WERE REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION WAS RECEIVED THAT ONLY 2 SCREWS BROKE AND IT IS UNKNOWN WHICH SCREWS BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15201655 ↗

Other CEMENT, BONE, VERTEBROPLASTY devices

All CEMENT, BONE, VERTEBROPLASTY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page