Skip to content

CD HORIZON® Fenestrated Screw Set 55750027555Recalls & reported problems

Medtronic Sofamor Danek, Inc. · FNS MAS 55750027555 STER CCM 7.5 X 55 · Class 2 · Implant

All models of CD HORIZON Fenestrated Screw Set (27)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000149192), received between Feb 7, 2022 and Feb 7, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Leak/Splash2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 7, 2022Leak/Splash; Leak/Splash

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT IMPLANTED WITH CEMENT IN AN AUGMENTATION PROCEDURE FOR AN UNKNOWN SPINAL INDICATION. IT WAS REPORTED THAT THE CEMENT LEAKED INTO SCREW TULIP DUE TO CEMENTATION WITH VOYAGER SCREWS. SCREWS HAVE BEEN USED WITH AUGMENTATION PROCEDURE BUT THE CEMENT LEAKED OUT BETWEEN SCREW CRONE AND VOYAGER DISPOSABLE. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. THERE WAS A DELAY OF 30 MINUTES IN THE SURGERY. NO FURTHER COMPLICATIONS WERE REPORTED/ ANTICIPATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13469687 ↗
  • MalfunctionFeb 7, 2022Leak/Splash; Leak/Splash

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT IMPLANTED WITH CEMENT IN AN AUGMENTATION PROCEDURE FOR AN UNKNOWN SPINAL INDICATION. IT WAS REPORTED THAT THE CEMENT LEAKED INTO SCREW TULIP DUE TO CEMENTATION WITH VOYAGER SCREWS. SCREWS HAVE BEEN USED WITH AUGMENTATION PROCEDURE BUT THE CEMENT LEAKED OUT BETWEEN SCREW CRONE AND VOYAGER DISPOSABLE. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. THERE WAS A DELAY OF 30 MINUTES IN THE SURGERY. NO FURTHER COMPLICATIONS WERE REPORTED/ ANTICIPATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13469684 ↗

Other CEMENT, BONE, VERTEBROPLASTY devices

All CEMENT, BONE, VERTEBROPLASTY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page