Skip to content

CD HORIZON® Fenestrated Screw Set 55750024535Recalls & reported problems

Medtronic Sofamor Danek, Inc. · FNS MAS 55750024535 STER CCM 4.5 X 35 · Class 2 · Implant

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000149321), received between Dec 4, 2025 and Dec 4, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 4, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR FIXAT ION USING 2 ABOVE 2 BELOW PROCEDURE WAS PERFORMED FOR A VERTEBRAL FRACTURE AT TH8 DUE TO DIFFUSE IDIOPATHIC SKELETAL HYPEROSTOSIS (DISH). IT WAS REPORTED THAT DURING PROCEDURE, BONE CEMENT LEAKAGE WAS DETECTED INTRAOPERATIVELY INTO THE BLOOD VESSEL AT THE RIGHT TH7 AND RIGHT TH6 LEVELS, AND POSSIBLE LEAKAGE TOWARD THE POSTERIOR SIDE AT LEFT TH6 WHILE FILLING EACH VERTEBRAL BODY AND THE FILLING WAS STOPPED IMMEDIATELY UPON DETECTION, MINIMIZING THE EXTENT OF LEAKAGE. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION WAS RECEIVED THAT THE OPINION AT THE END OF THE OPERATION WAS THAT THE TH6 LEFT SCREW WAS INSERTED FROM THE OUTSIDE AND THE CEMENT FLOWED LATERALLY, BUT AFTER THE OPERATION, THE CT REVEALED THAT THE SCREW WAS OUT OF BOUNDS IN THE INTERIOR AND THAT THE CEMENT HAD LEAKED INTO THE SPINAL CANAL. THERE IS NO OTHER PROBLEM WITH THE PATIENT AT THIS STAGE. ADDITIONAL INFORMATION WAS RECEIVED THAT THE PHYSICIAN¿S POSTOPERATIVE ASSESSMENT WAS INCORRECT. AT THE TIME OF THE A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23714446 ↗

Other CEMENT, BONE, VERTEBROPLASTY devices

All CEMENT, BONE, VERTEBROPLASTY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.