Skip to content

CD HORIZON® Fenestrated Screw Set 55750025550Recalls & reported problems

Medtronic Sofamor Danek, Inc. · FNS MAS 55750025550 STER CCM 5.5 X 50 · Class 2 · Implant

All models of CD HORIZON Fenestrated Screw Set (27)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000149598), received between Jan 30, 2025 and Mar 12, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Backflow” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Backflow2
  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 12, 2025Backflow; Product Quality Problem; Backflow; Product Quality Problem

    INFORMATION RECEIVED FROM HEALTHCARE PROVIDER VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. PRE-OP DIAGNOSIS WAS SPINAL CANAL STENOSIS. IT WAS REPORTED THAT, ONE OF THE REPORTED SCREWS WAS USED AT L2. THE DELIVERY DEVICE ON THE LEFT SIDE OF L2 COULD NOT BE REMOVED BY TURNING IT BY HAND, AND IT CAME OFF WHEN THE DEVICE WAS GRASPED WITH PLIERS. THE CEMENT WAS FILLED ACCORDING TO THE IMAGE CHECK, BUT THE TIP CAME OFF THE TORX DUE TO SHAFT SHAKE, ETC. DURING THE PROCESS. THE CEMENT LEAKED INTO THE SCREW HEAD. THE TAB WAS REMOVED, AND THE TIP WAS COLLECTED USING A CHISEL, A LUER, AND PLIERS. THE ROD WAS EVEN BROUGHT TO A STATE WHERE IT CAN BE PUT ON. THE OTHER SCREW WAS USED AT L4, IT WAS OBSERVED THAT AN ABNORMAL CHANGE OCCURRED DURING CEMENT MANIPULATION, THE PUSHING ROD COULD NOT BE ADVANCED DUE TO THE TIGHTNESS DURING INJECTION OF CEMENT AT THE RIGHT L4 (PERCUTANEOUS PORTION). AT THAT STAGE, IT WAS MENTIONED TO WAIT AND CHECK, BUT THE PUSH STICK MOVED FORWARD WHILE CALLING OUT AND THE PUSHING STICK REMAINED IN PLACE. CEMENT LEAKAGE IN THE SCREW HEAD AT THE RIGHT SIDE OF L4 WHERE AN ABNORMALITY WAS FELT. IT WAS IDENTIFIED THAT THE PRODUCT WAS NOT FILLED W

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21582542 ↗
  • MalfunctionJan 30, 2025Backflow; Backflow

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING AN EVENT WHICH OCCURRED DURING AN UNKNOWN SPINAL PROCEDURE. IT WAS REPORTED THAT THE CEMENT COULD NOT BE INSERTED PROPERLY AND LEAKED OUT AGAIN ABOVE THE SCREW HEAD. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21272332 ↗

FDA clearance and classification

Cleared through 510(k) K201362 on Aug 19, 2020 (decision: Substantially Equivalent). Submitted by Medtronic as “CD Horizon™ Fenestrated Screw Set, CD Horizon™ Spinal System, Kyphon™ HV-R™ Bone Cement”. The FDA review took 89 days. FDA record ↗

The FDA regulates product code NDN (cement, bone, vertebroplasty) under 21 CFR 888.3027 as a Class II device. 21 CFR 888.3027 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the CD HORIZON® Fenestrated Screw Set 55750025550

Has the CD HORIZON® Fenestrated Screw Set 55750025550 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00763000149598), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the CD HORIZON® Fenestrated Screw Set 55750025550?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jan 30, 2025 and Mar 12, 2025. In those reports, the most common specific device problems are “Backflow” (2) and “Product Quality Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CD HORIZON® Fenestrated Screw Set 55750025550?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CD HORIZON® Fenestrated Screw Set 55750025550 FDA approved?

It was cleared through 510(k) K201362 on Aug 19, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the CD HORIZON® Fenestrated Screw Set 55750025550 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the CD HORIZON® Fenestrated Screw Set 55750025550, and is it still sold?

The FDA's device database lists Medtronic Sofamor Danek, Inc. as the company responsible for it (brand CD HORIZON® Fenestrated Screw Set, model 55750025550). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CEMENT, BONE, VERTEBROPLASTY devices

All CEMENT, BONE, VERTEBROPLASTY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page