Skip to content

Telescope™ TELE7FRecalls & reported problems

Medtronic, Inc. · GEC TELE 7F TELESCOPE US · Class 2

All models of Telescope (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000156978), received between Dec 8, 2021 and Nov 22, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Material Deformation3
  • Peeled/Delaminated2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 22, 2024Peeled/Delaminated; Detachment of Device or Device Component; Material Deformation; Peeled/Delaminated; Detachment of Device or Device Component; Material Deformation

    IT WAS REPORTED THAT DURING A PROCEDURE AN ATTEMPT WAS MADE TO USE ONE TELESCOPE GUIDE EXTENSION CATHETER TO TREAT A MODERATELY TORTUOUS, NON-CALCIFIED LESION LOCATED IN THE MID SAPHENOUS VEIN GRAFT (SVG) AND A DEVICE DETACHMENT OCCURRED. IT WAS STATED THAT DURING ADVANCEMENT THROUGH THE GUIDE THE TELESCOPE BROKE. THE DEVICE WAS REMOVED FROM THE GUIDE CATHETER AS IT WAS STILL CONNECTED. THE DEVICE WAS REMOVED FROM PACKAGING AND PREPPED PER IFU WITH NO ISSUES NOTED. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE AND EXCESSIVE FORCE WAS NOT USED DURING INSERTION/DELIVERY. THE PATIENT IS ALIVE WITHNO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20757474 ↗
  • InjuryApr 4, 2024Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation

    AN ATTEMPT WAS MADE TO USE ONE 7F TELESCOPE GUIDE EXTENSION CATHETER DURING A PROCEDURE. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES NOTED. IT WAS REPORTED THAT THE SHAFT DETACHED AT THE ENTRY PORT DURING DELIVERY TO/AT THE LESION. THE DETACHED FRAGMENT WAS REMOVED. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19040757 ↗
  • MalfunctionDec 8, 2021Peeled/Delaminated; Detachment of Device or Device Component; Material Deformation; Peeled/Delaminated; Detachment of Device or Device Component; Material Deformation

    AN ATTEMPT WAS MADE TO USE A TELESCOPE GUIDE EXTENSION CATHETER TO TREAT A MODERATELY TORTUOUS, NON CALCIFIED LESION WITH 90% STENOSIS IN THE MID CIRCUMFLEX (CX) ARTERY. THE DEVICE WAS INSPECTED WITH NO ISSUES. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT NOTED USED DURING INSERTION. IT WAS REPORTED THAT WHEN ATTEMPTING TO ADVANCE THE STENT THROUGH THE TELESCOPE GUIDE EXTENSION CATHETER, RESISTANCE WAS FELT. BOTH THE TELESCOPE AND STENT WERE REMOVED FROM THE PATIENT AND IT WAS NOTED THAT THE PROXIMAL PORTION OF THE TELESCOPE GUIDE EXTENSION CATHETER WAS SEPARATED FROM THE WIRE. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12960291 ↗

Other Catheter, percutaneous devices

All Catheter, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page