Skip to content

Attain Command™ + SureValve™ 6250V-MB2XRecalls & reported problems

Medtronic, Inc. · CATHETER 6250V-MB2X ATTAIN COMMAND · Class 2

All models of Attain Command + SureValve (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000160319), received between Jun 4, 2025 and Jun 4, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Accessory Incompatible” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Accessory Incompatible2
  • Break2
  • Material Integrity Problem2
  • Mechanical Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2025Accessory Incompatible; Break; Mechanical Problem; Material Integrity Problem; Accessory Incompatible; Break; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, A FEW DIFFERENT CATHETERS WERE USED AS THERE WAS SLITTING DIFFICULTY THAT CAUSED THE LEFT VENTRICULAR (LV) LEAD TO DISLODGE. IT WAS NOTED THAT THE CATHETERS KEPT BREAKING AND THAT THE VALVES WERE VERY DIFFICULT TO PASS THROUGH. ONE OF THE CATHETERS BECAME STUCK IN THE VALVE DURING THE LEAD IMPLANT. A DIFFERENT LV LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22144206 ↗
  • MalfunctionJun 4, 2025Accessory Incompatible; Break; Mechanical Problem; Material Integrity Problem; Accessory Incompatible; Break; Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, A FEW DIFFERENT CATHETERS WERE USED AS THERE WAS SLITTING DIFFICULTY THAT CAUSED THE LEFT VENTRICULAR (LV) LEAD TO DISLODGE. IT WAS NOTED THAT THE CATHETERS KEPT BREAKING AND THAT THE VALVES WERE VERY DIFFICULT TO PASS THROUGH. ONE OF THE CATHETERS BECAME STUCK IN THE VALVE DURING THE LEAD IMPLANT. A DIFFERENT LV LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22144041 ↗