Skip to content

Attain Command™ + SureValve™ 6250VC (UDI 00763000160357)Recalls & reported problems

Medtronic, Inc. · CATHETER 6250VC ATTAIN COMMAND · Class 2

All models of Attain Command + SureValve (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000160357), received between Jul 24, 2024 and Jul 24, 2024 — classified as 1 death.

The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation, Positioning or Separation Problem1

Patient problems

  • Bradycardia1
  • Cardiac Arrest1
  • Cardiac Tamponade1
  • Hypoxia1
  • Low Blood Pressure/ Hypotension1
  • Pericardial Effusion1
  • Syncope/Fainting1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathJul 24, 2024Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem

    IT WAS FURTHER CONFIRMED IN FOLLOW UP THAT THE CAUSE OF DEATH WAS NOT RELATED TO ANY LEAD PERFORMANCE OR PATIENT HISTORY OF DISEASE OR PATIENT CONDITION.

    Patient problems: Bradycardia; Cardiac Arrest; Cardiac Tamponade; Hypoxia; Low Blood Pressure/ Hypotension; Pericardial Effusion; Syncope/Fainting

    MAUDE report 19816770 ↗

Other Catheter, percutaneous devices

All Catheter, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page