n/a 6550041Recalls & reported problems
Medtronic Sofamor Danek, Inc. · VOYAGER 6550041 CEMENT DELIVERY GUIDE · Class 1
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00763000216740), received between Nov 13, 2023 and Nov 10, 2025 — classified as 16 malfunctions.
The most frequently reported specific device problem is “Mechanics Altered” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanics Altered10
- Mechanical Jam4
- Mechanical Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions16
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered
INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23509855 ↗MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered
INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23509837 ↗MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered
INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23509825 ↗MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered
INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23509800 ↗MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered
INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23509667 ↗MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered
INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23509480 ↗MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered
INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23509408 ↗MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered
INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23509363 ↗MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered
INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23509310 ↗MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered
INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23509236 ↗MalfunctionAug 5, 2025Mechanical Problem; Mechanical Problem
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING TH12 WAS STABILIZED USING A 22-FENESTRATED SYSTEM SPINAL THERAPY FOR VERTEBRAL FRACTURE. IT WAS REPORTED THAT WHILE THE DELIVERY GUIDE WAS ATTACHED TO THE SCREW, AN ATTEMPT WAS MADE TO SET THE OUTER SHEATH OF THE DELIVERY DEVICE ONTO THE GUIDE. HOWEVER, THE CLAWS OF THE OUTER SHEATH FAILED TO REACH THE NECK OF THE DELIVERY GUIDE, PREVENTING PROPER ATTACHMENT. THERE WAS NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION RECEIVED FROM THE MANUFACTURER REPRESENTATIVE THAT ALTHOUGH THE GREEN LINE OF THE GUIDES EXCEEDED THE BLACK LINE OF THE SCREW TAB, THE DOCTOR SAID THAT THERE WAS NO FEELING THAT THE AXES WERE ALIGNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22711533 ↗MalfunctionAug 20, 2024Mechanical Problem; Mechanical Problem
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A FNS TIP AND CEMENT DELIVERY GU IDE USED ON PATIENT HAVING LAMINECTOMY THERAPY, STABILIZATION WITH VOYAGER 5.5 IMPLANTS FOR D12 FRACTURE ON OSTEOPOROTIC PATIENT. IT WAS REPORTED THAT THE CEMENT LEAKED BETWEEN TIPS AND SCREWS. THE SURGEON WAS FORCED TO REMOVED A SCREW BECAUSE A TIP WAS IMPOSSIBLE TO REMOVE FROM THE TULIP. GUIDE DIDN'T SECURELY HOLD THE TIP AND OFTEN THE TIP REMAINED ENGAGED IN THE SCREW TULIP. SO IT MADE IMPOSSIBLE TO COMPLETE BONE AUGMENTATION. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WAS A DELAY OF 1 HOUR IN OVERALL PROCEDURE. THE PROCEDURE WAS PARTIALLY COMPLETED, IT WAS IMPOSSIBLE TO INJECT CEMENT DURING BONE AUGMENTATION. LEVEL IMPLANTED D10 ¿ D11¿ L1 ¿ L2. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20006886 ↗MalfunctionNov 13, 2023Mechanical Jam; Mechanical Jam
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING DORSAL STABILIS ATION, FENESTRATED FROM L3 TO L5 FOR AN INDICATION OF L4 OSTEOPOROTIC FRACTURE. IT WAS REPORTED THAT AFTER CEMENTING, THE FNS TIP REMAINED STUCK IN THE TULIP DESPITE IMMEDIATE REMOVAL FROM THE SCREW HEAD. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18127929 ↗MalfunctionNov 13, 2023Mechanical Jam; Mechanical Jam
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING DORSAL STABILIZATION, FENESTRATED FROM L3 TO L5 FOR AN INDICATION OF L4 OSTEOPOROTIC FRACTURE. IT WAS REPORTED THAT AFTER CEMENTING, THE FNS TIP REMAINED STUCK IN THE TULIP DESPITE IMMEDIATE REMOVAL FROM THE SCREW HEAD. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18127924 ↗MalfunctionNov 13, 2023Mechanical Jam; Mechanical Jam
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING DORSAL STABILIS ATION, FENESTRATED FROM L3 TO L5 FOR AN INDICATION OF L4 OSTEOPOROTIC FRACTURE. IT WAS REPORTED THAT AFTER CEMENTING, THE FNS TIP REMAINED STUCK IN THE TULIP DESPITE IMMEDIATE REMOVAL FROM THE SCREW HEAD. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18127919 ↗MalfunctionNov 13, 2023Mechanical Jam; Mechanical Jam
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING DORSAL STABILIS ATION, FENESTRATED FROM L3 TO L5 FOR AN INDICATION OF L4 OSTEOPOROTIC FRACTURE. IT WAS REPORTED THAT AFTER CEMENTING, THE FNS TIP REMAINED STUCK IN THE TULIP DESPITE IMMEDIATE REMOVAL FROM THE SCREW HEAD. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18127918 ↗
Other DISPENSER, CEMENT devices
- SMARTSET 3102-080Depuy (Ireland) · DISPENSER, CEMENT75 reports
- Medtronic Reusable Instrument 6550202Medtronic Sofamor Danek, Inc. · DISPENSER, CEMENT2 deaths59 reports
- KYPHON® Express™ Directional Bone Filler Device F06BMedtronic Sofamor Danek, Inc. · DISPENSER, CEMENT1 recall1 report
- EXPEDIUM 279726500Medos International Sàrl · DISPENSER, CEMENT44 reports
- Revolution 0606600000Stryker Corporation · DISPENSER, CEMENT33 reports
- NA 0606100000Stryker Corporation · DISPENSER, CEMENT15 reports
- NA 0206600000Stryker Corporation · DISPENSER, CEMENT14 reports
- N/A 6550044Medtronic Sofamor Danek, Inc. · DISPENSER, CEMENT11 reports