Medtronic Sofamor Danek, Inc. · VOYAGER 6550041 CEMENT DELIVERY GUIDE · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00763000216740), received between Nov 13, 2023 and Nov 10, 2025 — classified as 16 malfunctions.

The most frequently reported specific device problem is “Mechanics Altered” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 10 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanics Altered10
  • Mechanical Jam4
  • Mechanical Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions16

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23509855 ↗
  • MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23509837 ↗
  • MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23509825 ↗
  • MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23509800 ↗
  • MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23509667 ↗
  • MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23509480 ↗
  • MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23509408 ↗
  • MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23509363 ↗
  • MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23509310 ↗
  • MalfunctionNov 10, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. IT WAS REPORTED THAT THE GUIDE SLEEVES HELD THE CEMENT DELIVERY TIPS VERY LOOSELY AND WHEN CEMENT WAS APPLIED CEMENT CAME OUT OF THE TULIPS OF THE GUIDES THAT DID NOT HAVE A TIGHT FIT TO THE TIPS. NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23509236 ↗
  • MalfunctionAug 5, 2025Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING TH12 WAS STABILIZED USING A 22-FENESTRATED SYSTEM SPINAL THERAPY FOR VERTEBRAL FRACTURE. IT WAS REPORTED THAT WHILE THE DELIVERY GUIDE WAS ATTACHED TO THE SCREW, AN ATTEMPT WAS MADE TO SET THE OUTER SHEATH OF THE DELIVERY DEVICE ONTO THE GUIDE. HOWEVER, THE CLAWS OF THE OUTER SHEATH FAILED TO REACH THE NECK OF THE DELIVERY GUIDE, PREVENTING PROPER ATTACHMENT. THERE WAS NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION RECEIVED FROM THE MANUFACTURER REPRESENTATIVE THAT ALTHOUGH THE GREEN LINE OF THE GUIDES EXCEEDED THE BLACK LINE OF THE SCREW TAB, THE DOCTOR SAID THAT THERE WAS NO FEELING THAT THE AXES WERE ALIGNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22711533 ↗
  • MalfunctionAug 20, 2024Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A FNS TIP AND CEMENT DELIVERY GU IDE USED ON PATIENT HAVING LAMINECTOMY THERAPY, STABILIZATION WITH VOYAGER 5.5 IMPLANTS FOR D12 FRACTURE ON OSTEOPOROTIC PATIENT. IT WAS REPORTED THAT THE CEMENT LEAKED BETWEEN TIPS AND SCREWS. THE SURGEON WAS FORCED TO REMOVED A SCREW BECAUSE A TIP WAS IMPOSSIBLE TO REMOVE FROM THE TULIP. GUIDE DIDN'T SECURELY HOLD THE TIP AND OFTEN THE TIP REMAINED ENGAGED IN THE SCREW TULIP. SO IT MADE IMPOSSIBLE TO COMPLETE BONE AUGMENTATION. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WAS A DELAY OF 1 HOUR IN OVERALL PROCEDURE. THE PROCEDURE WAS PARTIALLY COMPLETED, IT WAS IMPOSSIBLE TO INJECT CEMENT DURING BONE AUGMENTATION. LEVEL IMPLANTED D10 ¿ D11¿ L1 ¿ L2. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20006886 ↗
  • MalfunctionNov 13, 2023Mechanical Jam; Mechanical Jam

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING DORSAL STABILIS ATION, FENESTRATED FROM L3 TO L5 FOR AN INDICATION OF L4 OSTEOPOROTIC FRACTURE. IT WAS REPORTED THAT AFTER CEMENTING, THE FNS TIP REMAINED STUCK IN THE TULIP DESPITE IMMEDIATE REMOVAL FROM THE SCREW HEAD. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18127929 ↗
  • MalfunctionNov 13, 2023Mechanical Jam; Mechanical Jam

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING DORSAL STABILIZATION, FENESTRATED FROM L3 TO L5 FOR AN INDICATION OF L4 OSTEOPOROTIC FRACTURE. IT WAS REPORTED THAT AFTER CEMENTING, THE FNS TIP REMAINED STUCK IN THE TULIP DESPITE IMMEDIATE REMOVAL FROM THE SCREW HEAD. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18127924 ↗
  • MalfunctionNov 13, 2023Mechanical Jam; Mechanical Jam

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING DORSAL STABILIS ATION, FENESTRATED FROM L3 TO L5 FOR AN INDICATION OF L4 OSTEOPOROTIC FRACTURE. IT WAS REPORTED THAT AFTER CEMENTING, THE FNS TIP REMAINED STUCK IN THE TULIP DESPITE IMMEDIATE REMOVAL FROM THE SCREW HEAD. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18127919 ↗
  • MalfunctionNov 13, 2023Mechanical Jam; Mechanical Jam

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING DORSAL STABILIS ATION, FENESTRATED FROM L3 TO L5 FOR AN INDICATION OF L4 OSTEOPOROTIC FRACTURE. IT WAS REPORTED THAT AFTER CEMENTING, THE FNS TIP REMAINED STUCK IN THE TULIP DESPITE IMMEDIATE REMOVAL FROM THE SCREW HEAD. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18127918 ↗

FDA clearance and classification

The FDA regulates product code KIH (dispenser, cement) under 21 CFR 888.4200 as a Class I device. 21 CFR 888.4200 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the n/a 6550041

Has the n/a 6550041 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00763000216740), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the n/a 6550041?

The FDA's MAUDE database lists 16 adverse event reports for this device, received between Nov 13, 2023 and Nov 10, 2025. In those reports, the most common specific device problems are “Mechanics Altered” (10), “Mechanical Jam” (4) and “Mechanical Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the n/a 6550041?

None of the reports are classified as deaths, 0 as injuries and 16 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the n/a 6550041 FDA approved?

Devices of this type (product code KIH) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Medtronic Sofamor Danek, Inc.

Who makes the n/a 6550041, and is it still sold?

The FDA's device database lists Medtronic Sofamor Danek, Inc. as the company responsible for it (brand n/a, model 6550041). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other DISPENSER, CEMENT devices

All DISPENSER, CEMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page