N/A 960-556 (UDI 00763000223069)Recalls & reported problems
Medtronic Navigation, Inc. · PLANAR 960-556 PASSIVE BLUNT PROBE · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000223069), received between Jan 3, 2019 and Dec 13, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Device Sensing Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Sensing Problem2
- Incorrect, Inadequate or Imprecise Result or Readings1
- Material Integrity Problem1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Patient Involvement1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 13, 2024Mechanical Problem; Device Sensing Problem; Mechanical Problem; Device Sensing Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A CRANIAL RESECTION PROCEDURE. IT WAS REPORTED THAT THAT DURING A CASE, THE PASSIVE PLANAR PROBE WAS UNABLE TO BE TRACKED. A MEDTRONIC REPRESENTATIVE (REP) ARRIVED ON SITE AND REPORTED THAT THE TOP LEG OF THE PROBE WAS VISIBLY BENT TO THE LEFT SIDE. WHEN THE INSTRUMENT WAS HELD IN FRONT OF THE CAMERA, THE GEOMETRY ERROR OF THE PROBE WAS 0.8 MM. WHEN THE TOP SPHERE WAS COVERED, THE GEOMETRY ERROR DROPPED TO 0.4 MM AND THE INSTRUMENT COULD BE TRACKED. WHEN COVERING ANY OTHER SPHERE, THE GEOMETRY ERROR REMAINED AT 0.8 MM. THERE WAS A 30 SECOND DELAY. THERE WAS NO IMPACT ON THE PATIENT'S OUTCOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20926324 ↗MalfunctionSep 11, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ADDITIONAL INFORMATION WAS RECEIVED. THE INACCURACY WAS 2 MM LEFT OF THE EXPECTED TARGET. THE LIKELY CAUSE OF THE ISSUE WAS THERE MIGHT BE SOME ERROR WITH THE PRONGS THAT HOLD THE SPHERES ON THE HANDLE OF THE PROBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17720082 ↗MalfunctionNov 21, 2022Device Sensing Problem; Device Sensing Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A GUIDANCE SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE PASSIVE PLANAR PROBE WOULD NOT VERIFY. AFTER CLOSER INSPECTION THEY NOTICED THAT THE FRAME HAD BEEN BENT. NO PATIENT WAS PRESENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15837265 ↗MalfunctionJan 3, 2019Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING AN INSTRUMENT THAT WAS USED WITH A NAVIGATION SYSTEM. IT WAS REPORTED OUTSIDE OF A PROCEDURE THAT THE PASSIVE PLANAR BLUNT PROBE WAS BROKEN. THERE WAS NO PATIENT PRESENT.
Patient problems: No Patient Involvement
MAUDE report 8214270 ↗
Other Neurological stereotaxic Instrument devices
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument4 deaths4,668 reports
- Stealthstation® S7™ 9733856Medtronic Navigation, Inc. · Neurological stereotaxic Instrument6 deaths3,908 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death1,896 reports
- StealthStation S8 9735669Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1,831 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument2 deaths1,542 reports
- Fusion™ ENT Image Guidance System 9733560XOMMedtronic Navigation, Inc. · Neurological stereotaxic Instrument1,316 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument8 deaths984 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death695 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page