N/A 960-556 (UDI 00763000223069)Recalls & reported problems

Medtronic Navigation, Inc. · PLANAR 960-556 PASSIVE BLUNT PROBE · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000223069), received between Jan 3, 2019 and Dec 13, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Sensing Problem2
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Material Integrity Problem1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Patient Involvement1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 13, 2024Mechanical Problem; Device Sensing Problem; Mechanical Problem; Device Sensing Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A CRANIAL RESECTION PROCEDURE. IT WAS REPORTED THAT THAT DURING A CASE, THE PASSIVE PLANAR PROBE WAS UNABLE TO BE TRACKED. A MEDTRONIC REPRESENTATIVE (REP) ARRIVED ON SITE AND REPORTED THAT THE TOP LEG OF THE PROBE WAS VISIBLY BENT TO THE LEFT SIDE. WHEN THE INSTRUMENT WAS HELD IN FRONT OF THE CAMERA, THE GEOMETRY ERROR OF THE PROBE WAS 0.8 MM. WHEN THE TOP SPHERE WAS COVERED, THE GEOMETRY ERROR DROPPED TO 0.4 MM AND THE INSTRUMENT COULD BE TRACKED. WHEN COVERING ANY OTHER SPHERE, THE GEOMETRY ERROR REMAINED AT 0.8 MM. THERE WAS A 30 SECOND DELAY. THERE WAS NO IMPACT ON THE PATIENT'S OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20926324 ↗
  • MalfunctionSep 11, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    ADDITIONAL INFORMATION WAS RECEIVED. THE INACCURACY WAS 2 MM LEFT OF THE EXPECTED TARGET. THE LIKELY CAUSE OF THE ISSUE WAS THERE MIGHT BE SOME ERROR WITH THE PRONGS THAT HOLD THE SPHERES ON THE HANDLE OF THE PROBE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17720082 ↗
  • MalfunctionNov 21, 2022Device Sensing Problem; Device Sensing Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A GUIDANCE SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE PASSIVE PLANAR PROBE WOULD NOT VERIFY. AFTER CLOSER INSPECTION THEY NOTICED THAT THE FRAME HAD BEEN BENT. NO PATIENT WAS PRESENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15837265 ↗
  • MalfunctionJan 3, 2019Material Integrity Problem; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING AN INSTRUMENT THAT WAS USED WITH A NAVIGATION SYSTEM. IT WAS REPORTED OUTSIDE OF A PROCEDURE THAT THE PASSIVE PLANAR BLUNT PROBE WAS BROKEN. THERE WAS NO PATIENT PRESENT.

    Patient problems: No Patient Involvement

    MAUDE report 8214270 ↗

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →