Catalyft PL Expandable Interbody System 6069116Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SPACER 6069116 CATALYFT PL40 LONG 11MM · Class 2 · Implant

All models of Catalyft PL Expandable Interbody System (16)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000246853), received between Jun 6, 2024 and Jun 23, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 16, 2025Open, Classified

    Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.

    Z-0580-2026 · Root cause: Labeling design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Mechanical Problem1
  • Mechanics Altered1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 23, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT SIX MONTHS AFTER A SPINAL PROCEDURE THE CAGE WAS FOUND TO HAVE FOLDED BACK INTO ITS ORIGINAL POSITION, THERE WAS A LOSS OF LORDOSIS, AND THE CAGE HAD CLOSED BY ITSELF. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22306118 ↗
  • MalfunctionJun 6, 2024Mechanical Problem; Mechanical Problem

    ADDITIONAL INFORMATION WAS RECEIVED THAT THERE WAS NO DELAY IN OVERALL PROCEDURE TIME. THE PROCEDURE INVOLVED WAS A MINIMALLY INVASIVE SURGICAL TRANSFORAMINAL LUMBAR INTERBODY FUSION MIS L4-5 TLIF ON (B)(6) 2024. THE PATIENT WAS DOING WELL POST-OP WITH THE RESOLUTION OF PRE-OP SYMPTOMS AND WAS CURRENTLY DOING PHYSICAL THERAPY AND WAS PROGRESSING WITH REHAB PROGRAM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19474785 ↗

Other Intervertebral fusion device with bone graft, lumbar devices

All Intervertebral fusion device with bone graft, lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page