1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000260637), received between Sep 13, 2023 and Jul 18, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 12, 2021Class IOpen, Classified

    Due to the presence of type IIIb endoleaks, stent fractures, and stent ring enlargement.

    Z-1201-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Leak/Splash2

Patient problems

  • Insufficient Information1
  • Vascular Dissection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 18, 2024Leak/Splash; Leak/Splash

    A VALIANT NAVION STENT GRAFT WAS IMPLANTED IN THE ENDOVASCULAR TREATMENT OF A IMH AND PAU. IT WAS REPORTED APPROXIMATELY 3 YEARS POST THE INDEX PROCEDURE, FOLLOW UP CT SHOWED THAT AN INCREASED ENDOLEAK AT THE LEVEL OF THE DISTAL ENDOGRAFT WAS NOTED. SPECIFICALLY, "BROAD-BASED CONTRAST OPACIFICATION EXTERNAL TO THE MEDIAL ASPECT OF THE DISTAL GRAFT PERSISTS AND IS SIMILAR IN APPEARANCE". PER THE PHYSICIAN THE CT "DEMONSTRATED EVIDENCE OF DSINE VERSUS TYPE IB ENDOLEAK". THIS ENDOLEAK WAS THOUGHT TO BE A TYPE II ENDOLEAK ON CT A YEAR PREVIOUSLY AND IS NOW DEEMED TO BE IB ENDOLEAK. NO CAUSE WAS REPORTED. NO ADDITIONAL CLINICAL SEQUELAE WERE PROVIDED AND THE PATIENT WILL BE MONITORED.

    Patient problems: Vascular Dissection

    MAUDE report 19771203 ↗
  • InjurySep 13, 2023Leak/Splash; Leak/Splash

    A VALIANT NAVION STENT GRAFT (B)(6) WAS IMPLANTED IN A ENDOVASCULAR TREATMENT OF A TYPE A AORTIC DISSECTION ON THE (B)(6) 2020 .A HEMI ARCH REPAIR WAS ALSO PERFORMED. FOUR MORE VALIANT NAVION STENT GRAFTS (B)(6) WERE IMPLANTED 2 MONTHS LATER FOR THE TREATMENT OF A TYPE A AORTIC DISSECTION WITH DISTAL EXTENSION AND A SYMPTOMATIC DISTAL THORACIC AORTIC ANEURYSM. DURING THE SECOND PROCEDURE, AN AORTOGRAPHY AND INTRAOPERATIVE VASCULAR ULTRASOUND REVEALED A WELL POSITIONED PREVIOUS TVAR IN THE TRUE LUMEN COVERING THE ORIFICE OF THE LEFT SUBCLAVIAN ARTERY, WHICH WAS COILED PREVIOUSLY. A TYPE I ENDOLEAK WITH FALSE LUMEN MEASURING 31 X 34MM WAS ALSO NOTED. USING A NON MDT SHEATH, THE 37 X 31 X 207MM VALIANT NAVION WAS DEPLOYED FIRSTLY IN THE TRUE LUMEN FOLLOWED BY THE 31 X 25 X 173MM STENT GRAFT WHICH WAS EXTENDED DOWN TO THE CELIAC ARTERY. SUBSEQUENTLY, INTRAVASCULAR ULTRASOUND SHOWED EXCELLENT PLACEMENT OF THE STENT GRAFTS WITH WIDELY OPENED CELIAC ARTERY. COMPLETION ANGIOGRAM REVEALED GOOD FLOW IN THE TRUE LUMEN. THERE WAS STILL FLOW AND PROXIMAL LEAK OBSERVED AT THE LEVEL OF THE PROXIMAL DESCENDING THORACIC AORTA. THEREFORE, THE FALSE LUMEN WAS STENTED WITH A 34 X 28 X 207MM STENT GRAF

    Patient problems: Insufficient Information

    MAUDE report 17736089 ↗