Skip to content

Valiant Navion™ VNMF3731C173TU (UDI 00763000260651)Recalls & reported problems

Medtronic, Inc. · STENT GRAFT VNMF3731C173TU VAL NAV FF TP · Class 3 · Implant

All models of Valiant Navion (137)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
1
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000260651), received between Sep 15, 2022 and Jun 18, 2024 — classified as 1 death, 2 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 12, 2021Class IOpen, Classified

    Due to the presence of type IIIb endoleaks, stent fractures, and stent ring enlargement.

    Z-1201-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Leak/Splash2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • DeathJun 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    VALIANT NAVION STENT GRAFTS WERE IMPLANTED IN THE ENDOVASCULAR TREATMENT OF A THORACIC AORTIC DISSECTION. (B)(6) WERE IMPLANTED INTO ZONES 3 <(>&<)> 4. TECHNICAL SUCCESS WAS ACHIEVED.  IT WAS REPORTED THAT ABOUT 18 MONTHS LATER, THE PATIENT EXPIRED DUE TO AN UNKNOWN CAUSE. THE PATIENT HAD PREVIOUSLY NOT BEEN RESPONDING TO CONTACT ATTEMPTS.  THE SITE ASSESSED THE DEATH AS POSSIBLY RELATED TO THE DEVICE, NOT RELATED TO THE STUDY PROCEDURE AS POSSIBLY RELATED TO THE DISSECTION UNDER STUDY. THE MEDICAL MONITOR ASSESSED THE DEATH AS NOT RELATED TO THE STUDY DEVICE, STUDY PROCEDURE OR DISSECTION UNDER STUDY. THE CEC ASSESSED THE DEATH AS NOT RELATED TO STUDY DEVICE OR PROCEDURE.

    Patient problems: Insufficient Information

    MAUDE report 19555722 ↗
  • MalfunctionDec 5, 2022Leak/Splash; Leak/Splash

    VALIANT NAVION STENT GRAFTS WERE IMPLANTED FOR THE ENDOVASCULAR TREATMENT OF A THORACIC AORTIC DISSECTION .  IT WAS REPORTED THAT CT IMAGING TAKEN 6 MONTHS LATER NOTED EXPANSION IN THE OUTER DIMENSIONS OF THE THORACIC AORTA RELATED TO CONTINUED EXPANSION OF FALSE LUMEN 59MM X 57MM AT ITS WIDEST. THERE WAS MILD PROGRESSIVE ENLARGEMENT OF SUPRA CELIAC ABDOMINAL AORTA DISTAL TO ILIAC. THERE WAS NO SECONDARY PROCEDURE PERFORMED. IT IS UNKNOWN IF THE CONTINUED FALSE LUMEN EXPANSION WAS IN AN AREA THAT WAS COVERED BY A STENT GRAFT. IT WAS REPORTED THAT THE PATIENT EXPIRED A FURTHER 6 MONTHS LATER. THE CAUSE OF DEATH WAS REPORTED AS A CARDIAC ARREST. THE PHYSICIAN HAS ASSESSED THE DEATH AS NOT RELATED TO THE DEVICE OR PROCEDURE. NO ADDITIONAL SEQUELAE WAS REPORTED AND THE PATIENT IS EXPIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15914057 ↗
  • MalfunctionSep 15, 2022Leak/Splash; Leak/Splash

    A VALIANT NAVION STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN UNKNOWN SIZE THORACIC  AORTIC ANEURYSM. IT WAS REPORTED THAT CORE LAB REVIEW OF CT IMAGING PERFORMED APPROXIMATELY 2 MONTHS POST THE INDEX PROCEDURE IDENTIFIED A TYPE IIIB ENDOLEAK. NO CAUSE OF THE ENDOLEAK WAS REPORTED. NO ADDITIONAL CLINICAL SEQUALAE WERE REPORTED AND THE PATIENT WILL BE MONITORED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15426236 ↗

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page