Valiant Navion™ VNMF3731C173TU (UDI 00763000260651)Recalls & reported problems
Medtronic, Inc. · STENT GRAFT VNMF3731C173TU VAL NAV FF TP · Class 3 · Implant
All models of Valiant Navion (137)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000260651), received between Sep 15, 2022 and Jun 18, 2024 — classified as 1 death, 2 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 12, 2021Class IOpen, Classified
Due to the presence of type IIIb endoleaks, stent fractures, and stent ring enlargement.
Z-1201-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Leak/Splash2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
DeathJun 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
VALIANT NAVION STENT GRAFTS WERE IMPLANTED IN THE ENDOVASCULAR TREATMENT OF A THORACIC AORTIC DISSECTION. (B)(6) WERE IMPLANTED INTO ZONES 3 <(>&<)> 4. TECHNICAL SUCCESS WAS ACHIEVED. IT WAS REPORTED THAT ABOUT 18 MONTHS LATER, THE PATIENT EXPIRED DUE TO AN UNKNOWN CAUSE. THE PATIENT HAD PREVIOUSLY NOT BEEN RESPONDING TO CONTACT ATTEMPTS. THE SITE ASSESSED THE DEATH AS POSSIBLY RELATED TO THE DEVICE, NOT RELATED TO THE STUDY PROCEDURE AS POSSIBLY RELATED TO THE DISSECTION UNDER STUDY. THE MEDICAL MONITOR ASSESSED THE DEATH AS NOT RELATED TO THE STUDY DEVICE, STUDY PROCEDURE OR DISSECTION UNDER STUDY. THE CEC ASSESSED THE DEATH AS NOT RELATED TO STUDY DEVICE OR PROCEDURE.
Patient problems: Insufficient Information
MAUDE report 19555722 ↗MalfunctionDec 5, 2022Leak/Splash; Leak/Splash
VALIANT NAVION STENT GRAFTS WERE IMPLANTED FOR THE ENDOVASCULAR TREATMENT OF A THORACIC AORTIC DISSECTION . IT WAS REPORTED THAT CT IMAGING TAKEN 6 MONTHS LATER NOTED EXPANSION IN THE OUTER DIMENSIONS OF THE THORACIC AORTA RELATED TO CONTINUED EXPANSION OF FALSE LUMEN 59MM X 57MM AT ITS WIDEST. THERE WAS MILD PROGRESSIVE ENLARGEMENT OF SUPRA CELIAC ABDOMINAL AORTA DISTAL TO ILIAC. THERE WAS NO SECONDARY PROCEDURE PERFORMED. IT IS UNKNOWN IF THE CONTINUED FALSE LUMEN EXPANSION WAS IN AN AREA THAT WAS COVERED BY A STENT GRAFT. IT WAS REPORTED THAT THE PATIENT EXPIRED A FURTHER 6 MONTHS LATER. THE CAUSE OF DEATH WAS REPORTED AS A CARDIAC ARREST. THE PHYSICIAN HAS ASSESSED THE DEATH AS NOT RELATED TO THE DEVICE OR PROCEDURE. NO ADDITIONAL SEQUELAE WAS REPORTED AND THE PATIENT IS EXPIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15914057 ↗MalfunctionSep 15, 2022Leak/Splash; Leak/Splash
A VALIANT NAVION STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN UNKNOWN SIZE THORACIC AORTIC ANEURYSM. IT WAS REPORTED THAT CORE LAB REVIEW OF CT IMAGING PERFORMED APPROXIMATELY 2 MONTHS POST THE INDEX PROCEDURE IDENTIFIED A TYPE IIIB ENDOLEAK. NO CAUSE OF THE ENDOLEAK WAS REPORTED. NO ADDITIONAL CLINICAL SEQUALAE WERE REPORTED AND THE PATIENT WILL BE MONITORED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15426236 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page