Skip to content

Valiant Navion™ VNMC3428C207TU (UDI 00763000261023)Recalls & reported problems

Medtronic, Inc. · STENT GRAFT VNMC3428C207TU VAL NAV CS TP · Class 3 · Implant

All models of Valiant Navion (137)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000261023), received between Nov 3, 2020 and Oct 11, 2023 — classified as 2 injuries.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash1
  • Malposition of Device1

Patient problems

  • Insufficient Information2
  • Injury1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 11, 2023Leak/Splash; Leak/Splash

    A VALIANT NAVION STENT GRAFT WAS IMPLANTED IN THE TREATMENT ON A TAA ON THE (B)(6) 2020. 4 VALIANT NAVION STENT GRAFTS IMPLANTED IN TOTAL, FROM PROXIMAL TO DISTAL VNMC3434C90TU, VNMC3131C223TU,VNMC2525C94TU WERE IMPLANTED ON THE (B)(6) 2020. THE FINAL NAVION VNMC3428C207TU WAS IMPLANTED ONE MONTH LATER FOR THE ABDOMINAL AORTA FOR THE TREATMENT OF A AAA. IT WAS REPORTED THAT A FOLLOW-UP CT TAKEN ON THE (B)(6) 2023 SHOWED A TYPE IIIB ENDOLEAK IN THE NAVION STENT GRAFTVNMC3131C223TU V29818675. CORE LAB REVIEW OF CT IMAGING FROM (B)(6) 2023 IDENTIFIED A NEW TYPE IIIB ENDOLEAK IN VNMC3131C223TU (V29818675) AND VNMC3428C207TU (V29961271). A NEW STENT FRACTURE WAS IDENTIFIED ON RING 4 OF VNMC3131C223TU (V29818675). NEW SRE WAS REPORTED OF +4.3MM ON RING 4 OF VNMC3131C223TU (V29818675) AND + 2.0MM TO RING 5 OF VNMC3131C223TU (V29818675). NO STENT RING MIGRATION WAS OBSERVED. THE MAX AORTIC DIAMETER WAS NOTED AS 70.9MM. THE IMAGING WAS NOTED AS NA FOR AORTIC ENLARGEMENT AND NON EVALUABLE FOR SOME GRAFTS FOR STENT FRACTURES AND STENT RING MIGRATION DUE TO DEVICE OVERLAPS. THE STENT GRAFT WAS RELINED USING A VALIANT CAPTIVIA STENT GRAFT AND THE ENDOLEAK WAS CONFIRNED AS RESOLVED. PER THE PHYS

    Patient problems: Insufficient Information

    MAUDE report 17910928 ↗
  • InjuryNov 3, 2020Malposition of Device; Malposition of Device

    A VALIANT NAVION STENT GRAFT WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 245 MM THORACIC AORTIC DISSECTION. IT WAS REPORTED THAT DURING THE INDEX PROCEDURE, THE DEVICE WAS DEPLOYED UNDER FLUOROSCOPIC VISUALIZATION IN THE DESCENDING THORACIC AORTA FOR A TYPE B AORTIC DISSECTION. DELIVERY SYSTEM WAS REMOVED AND A 16 FRENCH SHEATH WAS PLACED OVER THE WIRE. IT WAS STATED THAT A NON-MEDTRONIC CATHETER WAS INTRODUCED AND IT WAS DETERMINED THAT THE GRAFT HAD MIGRATED. THE VALIANT NAVION STENT GRAFT WAS THEN EXPLANTED. AS PER THE PHYSICIAN, CAUSE OF THE EVENT CANNOT BE DETERMINED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE

    Patient problems: Injury; Insufficient Information

    MAUDE report 10777258 ↗

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page